Technical Specifications - M&C PMA 05 Manual

Oxygen transmitter
Table of Contents

Advertisement

6

TECHNICAL SPECIFICATIONS

Part No.
Measuring range
Measurement output
Time for 90% value
Influence of barometric
pressure
Heating according to each
selected transmitter
Temperature status contact
on transmitters with heating
Accuracy after calibration
Sample gas outlet pressure
Influence of sample gas flow
Flow rate of sample gas
Influence of ambient
temperature
Ambient/measurement gas
temperature
Storage temperature
Ambient humidity
Electrical connection
Power supply
Materials in contact with
medium
Gas connections
Dimensions
Weight
Options
PT100 for thermostated version for temperature display or monitoring
Other measuring range than standard range
Zero suppression for spreading of measuring range
For more technical data please refer to the data sheets of your chosen transmitter.
9-3.20-MD
PMA05 MB10
01A0510 (lowest measuring range
0-10 vol.% O
)
2
0-100 vol% O
2
measuring ranges according to
client requirements
0-10V DC non-isolated and 4-20mA for the chosen range, non-isolated,
max. load 300Ω
<3 seconds at 60 l/h
the O
display varies proportionally with barometric pressure.
2
55 ° C
Contact load max. 48V 1A AC/DC
Contact closes at Temp. > 50 ° C
deviation: analogue signal output = ±1% of span at range 3-100% / digital
indicator = ±0,1 vol.% O
analogue signal output = ±2% of span at range 1%
the sample gas must flow without pressure out to atmosphere.
variation in gas flow between 0-60 Nl/hr air will cause a difference of < 0,1
vol.% O
2
max. 60Nl/hr
no influence up to +45°C
-10 ° C to +45 ° C
-20 ° C to +60 ° C.
0-95% RH.
12-pin connector
24V DC max. 1.5A
Platinum, glass, stainless steel 1.4571, viton, polypropylene, epoxy
Hose connection 4/6, other possible
115 x 115 x 112 mm (HxWxL)
1kg
Gas sampling and gas conditioning technology
(standard), other
2
PMA05 MB01
01A0520 (lowest measuring range
0-1 vol.% O
)
2
0-25 vol% O
(standard), other
2
measuring ranges according to
client requirements
7

Advertisement

Table of Contents
loading

This manual is also suitable for:

Pma05 mb10Pma05 mb01