In order to ensure safe operation of the diagnostic rack rotor, the following general safety regulations
must be followed:
• Do not remove the magnet at the rotor bottom.
• Do not use rotors which show any signs of corrosion and/or cracks.
• Use only with rotors which have been loaded properly.
• Never overload the rotor.
• Operate the rotor always with the lid closed.
• Use only accessories which have been approved by Thermo Fisher Scientific. Exceptions to this
rule are commercially available glass or plastic centrifuge tubes, provided they have been approved
for the speed or the RCF value of the rotor.
• Please observe the safety instructions.
Please pay particular attention to the following aspects:
• Rotor installation: Check that the rotor is locked properly into place before operating the
centrifuge.
• Always balance the samples.
Maximum sample density at maximum speed: 1.2