Rotor Loading

Before a Run

Before a Run
1. Please read and observe the safety instructions contained in this operating instruction and the
instructions for use.
2. Check the rotor and all accessory parts for damages such as cracks, scratches or traces of corrosion.
3. Check the rotor chamber, centrifuge spindle and the fixation of the rotor for damages.
4. Check the rotor's suitability using the chemical compatibility chart on on
5. Make sure the tubes or bottles do not touch the bucket lid.
6. Check the rotor bolts and apply grease (75003786) before using for the first time.
7. Check that each bucket or micro-plate carrier can swing freely by moving it carefully with your
hand. Weigh the bucket content (adapter and tube). Make sure you do not exceed the maximum
compartment load.

Proper Loading

To ensure safe operation of the centrifuge, the rotor must be evenly loaded at all times.
The buckets and micro plate carriers must always be fully loaded with racks. If there are racks missing,
the running racks might fall over and cause a rotor failure due to imbalance.
TM
4-2
75006480 Diagnostik
CAUTION Always be sure to use buckets of the same weight class. Operate the rotor always
with the lid closed.