Advertisement

Quick Links

EO-150
Patient
Pocket guide
Series Ventilators

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Air Liquide EO Series and is the answer not in the manual?

Questions and answers

Summary of Contents for eove Air Liquide EO Series

  • Page 1 EO-150 Patient Pocket guide Series Ventilators...
  • Page 3: Table Of Contents

    SUMMARY Indications for use Preferences menu Views Calibration Start / Stop ventilation Using Supplemental Oxygen Inlet The home screen Export ventilation data Navigating the patient screen and menu Cleaning...
  • Page 4 Quick Start: Press on the front panel keyboard to turn on the device. The home screen displays. In case of emergency, the ventilation module can be directly turned on simply by pressing button on the module keyboard. To start ventilation Click on on the keyboard or on the fan icon found on screen display.
  • Page 5: Indications For Use

    User’s manual. Non-specialist operators or institutions cautions encountering problems with set-up, operation or maintenance should immediately contact their EOVE representative. Clinician must read and understand the entire manual before using the An alternative mean of ventilation should always be EO-150 ventilator available for ventilator-dependent patients.
  • Page 6: Views

    Display screen EO device housing unit Ventilation module Inspiratory / Circuit Port Proximal pressure, valve, and Menu bar / keyboard proximal flow connectors Power source indicator High priority alarm indicator Technical alarm indicator Circuit alarm indicator Ventilation start / stop ON / OFF switch Physiological alarm indicator Medium priority alarm indicator...
  • Page 7: Start / Stop Ventilation

    Turning ON To start the device ventilation using the Touch Screen Ensure that the device has been charged prior to use or 1. Press on the touch screen connect to AC power or DC connector inlet. If the Home 2. Ventilation starts Screen is turned off, press the STANDBY Button until it turns on before turning on the device.
  • Page 8: The Home Screen

    On the home screen, the user can find important information regarding the alarms (2, 3, 7), the ventilation pressure (11), the preset modes set up by your clinician (9), the button for choosing your preferences and options and calibrating the ventilator in case of circuit change (14).
  • Page 9: Preferences Menu

    Preset mode menu (1-4): Presets installed by the clinician Patient, circuit and ventilation mode indications. and accessible by the patient when needed. Alarm signal: Lights up when there is an active alarm. Start / Stop button: Starts or stops ventilation. Alarm indicators: Indicates the type of alarm Airway pressure indicator: Indicates the pressure (technical, physiological, interface).
  • Page 10: Navigating The Patient Screen And Menu

    From this screen you can change preferences calibrate the ventilator in case of circuit change. From Home Screen, choose to access the Preferences, Calibration, Events Log or Trends menus.
  • Page 11 From this screen the patient can adjust Allows the screen to be rotated 180°. Rotation of the following settings for the device. Press the small circle to rotate the the screen screen. Allows brightness levels from low to high. Brightness/ Simply swipe the blue circle from left to Contrast right to choose the preferred brightness.
  • Page 12: Calibration

    The circuit calibration checks the caracteristics of the circuit configuration and must be executed after each circuit configuration change. Follow the 2 steps calibration processes when prompted and push on “Result” to go back to the home page NOTE: If a “caution”...
  • Page 13: Using Supplemental Oxygen Inlet

    Using Supplemental Oxygen inlet WARNING • Use only medical grade oxygen. • Ensure that the device is ventilating before the oxygen supply is turned on. • The oxygen flow must be turned off when the device is not ventilating so that oxygen does not accumulate within the device.
  • Page 14: Export Ventilation Data

    Export ventilation data with a USB key Insert a USB key in the USB port at the rear of the device In the menu tab, select “export data” Select “export on USB key” to copy files on the USB key...
  • Page 15 Export ventilation data with a USB key NOTE: The device saves a data file each day. It’s possible to manually save a new file by pushing on the “new file” button. Each files contains the last 24 hours of detailed waveforms data and up to one year of trends.
  • Page 16: Cleaning

    CLEANING AND DISINFECTION OF THE DEVICE Should be performed regularly. For that, wipe the exterior of the device with a damp cloth using a mild cleaning solution. WARNING Do not spray any solution inside the docking station when the ventilation module is removed. Use only a damp cloth with a mild cleaning solution.
  • Page 17 NOTE ..................................................................................................................................................................................................................................................................................................................................................................................................................................................................................
  • Page 18 NOTE ............................................................................................................................................................................................................................................................................................................................................................................................................................................................................
  • Page 19 NOTE ..................................................................................................................................................................................................................................................................................................................................................................................................................................................................................
  • Page 20 4 Boulevard Lucien Favre Immeuble Poincaré 64000 PAU www.eove.fr Contact Changing care. With you. Air Liquide Healthcare | VitalAire Canada 6990 Creditview Road, Unit 6 Mississauga, Ontario L5N 8R9 www.airliquidehealthcare.ca Tel: 1-888-629-0202 Fax: 1-888-828-0202 E-Mail: cs.vitalaire@airliquide.com Air Liquide Healthcare is a world leader in medical gases, home healthcare, hygiene products and healthcare specialty ingredients.

This manual is also suitable for:

Air liquide eo-150

Table of Contents