Care and cleaning
The probe tip is the most delicate part of the thermometer. It must be clean and intact to ensure accurate
readings. If the thermometer is ever accidentally used without a Hygiene cap, clean the probe tip as follows:
Very gently wipe the surface with a cotton swab or soft cloth moistened with alcohol. After the
alcohol has completely dried out, you can put a new Hygiene cap on and take a temperature
measurement.
If the probe tip is damaged, please contact Consumer Relations.
Use a soft, dry cloth to clean the thermometer display and exterior. Do not use abrasive cleaners.
Never submerge this thermometer in water or any other liquid. Store thermometer and Hygiene
caps in a dry location free from dust and contamination and away from direct sunlight.
Additional Hygiene caps (LF 40) are available at most stores carrying Braun ThermoScan.
Replacing the batteries
The thermometer is supplied with two 1.5 V type AA (LR 06) batteries. For best performance, we recommend Duracell®
alkaline batteries.
1. Insert new batteries when the battery symbol appears on the display.
2. Open the battery compartment. Remove the batteries and replace with new batteries, making
sure the poles are in the right direction.
3. Snap battery door into place.

Errors and troubleshooting

Error Message
Situation
No Hygiene cap is attached.
The thermometer probe was not positioned
securely in the ear.
An accurate measurement
was not possible.
POS = position error
Ambient temperature is not within
the allowed operating range
(10 – 40 °C or 50 – 104 °F).
Temperature taken is not within
typical human temperature range
(34 – 42.2 °C or 93.2 – 108 °F).
HI = too high
LO = too low
System error – self-check display flashes
continuously and will not be followed by
the ready beep and the ready symbol.
If error persists,
If error still persists,
6
Solution
Attach new, clean Hygiene cap.
Take care that the positioning of the probe is correct
and remains stable.
Replace Hygiene cap and reposition.
Press Start button to begin a new measurement.
Allow the thermometer to remain for
30 minutes in a room where
the temperature is between 10 and 40 °C
or 50 and 104 °F .
Make sure the probe tip and lens are clean and a
new, clean Hygiene cap is attached. Make sure the
thermometer is properly inserted. Then, take a new
temperature.
Wait 1 minute until the thermometer turns off
automatically, then turn on again.
... reset the thermometer by removing the batteries
and putting them back in.
... please contact the Service Centre.
Error Message
Situation
Battery is low, but thermometer will still
operate correctly.
Battery is too low to take correct
temperature measurement.

Product specifications

Displayed temperature range:
Operating ambient temperature range:
Storage temperature range:
Operating and storage relative humidity:
Display resolution:
Accuracy for displayed temperature range
35 – 42 °C (95 – 107.6 °F):
Outside this range:
Clinical repeatability:
Battery life:
Service life:
This thermometer is specified to operate at 1 atmospheric pressure or at altitudes with an atmospheric pressure
up to 1 atmospheric pressure (700 - 1060 hPa).
40 °C
10 °C
Equipment with
See Instruction for use
Operating temperature
type BF applied parts
Subject to change without notice.
This product conforms to the provisions of the EC directive 93/42/EEC. This product carries the CE mark and is
manufactured in conformity with the RoHS Directive 2011/65/EU.
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed description of EMC
requirements please contact your authorized local Service Centre (See guarantee card).
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT.
This product contains batteries and recyclable electronic waste. To protect the environment, do not dispose of
it in the household waste, but take it to appropriate local collection points.
Calibration
This product has been calibrated at the time of manufacture. If used according to the use instructions, periodic
calibration is not required. If at any time, you question the measuring accuracy, please contact our service
representative (see last page for contact information) or visit www.hot-europe.com/support.
Manufacturing date is given by the LOT number located in the battery compartment.
The first three (3) digits after LOT represent the Julian date that the product was manufactured and the next two (2)
digits represent the last two numbers of the calendar year the product was manufactured. The last identifiers are
the letters that represent the manufacturer.
An example: LOT 11614k, this product has been manufactured on the 116th day of the year 2014.
Guarantee
EMC Tables and Consumer Card are available on our website at www.hot-europe.com/support.
For support requests, please visit www.hot-europe.com/support or find service contact information on the last page
of this manual.
UK only: This guarantee in no way affects your rights under statutory law.
Solution
Insert new batteries.
Insert new batteries.
34 – 42.2 °C
(93.2 – 108 °F)
10 – 40 °C
(50 – 104 °F)
-25 – 55 °C
(-13 – 131 °F)
10-95% RH (non condensing)
0.1 °C or °F
Maximum Laboratory Error
±0.2 °C
(±0.4 °F)
±0.3 °C
(±0.5 °F)
±0.14 °C
(±0.26 °F)
2 years / 350 measurements
5 years
Storage temperature
Keep it dry
English
7