Advertisement

Quick Links

Confidence and control.
INSTRUCTION MANUAL
MEDIFRIDGE DISPLAY FRIDGE
MEDO220GD-NR
MEDO374GD-NR
08/06/2022

Advertisement

Table of Contents
loading

Summary of Contents for Bromic Heating MEDIFRIDGE MEDO220GD-NR

  • Page 1 Confidence and control. INSTRUCTION MANUAL MEDIFRIDGE DISPLAY FRIDGE MEDO220GD-NR MEDO374GD-NR 08/06/2022...
  • Page 3: Table Of Contents

    CONTENTS 1. PREFIX 2. IMPORTANT NOTES & WARNINGS 3. PRODUCT OVERVIEW 4. INSTALLATION 5. OPERATION 6. PRODUCT LOADING AND STORAGE 7. MAINTENANCE 8. TROUBLESHOOTING 9. ENVIRONMENTAL & DISPOSAL NOTES 10. WARRANTY Note: Bromic Refrigeration reserves the right to make changes to specifications, parts, components and equipment without prior notification. This Installation, operation and service manual may not be reproduced in any form with prior written consent from Bromic Refrigeration.
  • Page 4: Prefix

    1. PREFIX This manual contains important information about the installation, operation, and maintenance of Bromic MediFridge . Please pay close attention to the important safety information shown throughout this instruction manual. Any safety information will be accompanied by the following safety alert symbols: DANGER, WARNING, IMPORTANT...
  • Page 5: Important Notes & Warnings

    2. IMPORTANT NOTES AND WARNINGS • Read all instructions before installing or using this appliance. • Retain this manual for future reference. • Use this appliance only as described in this manual. Any other use not recommended by the manufacturer may cause fire, electric shock or injury.
  • Page 6 Warning: Risk of fire / Flammable materials • WARNING: The refrigerant (R600a Iso-Butane) used in this appliance is flammable, use caution during handling, servicing and disposal of the appliance. • WARNING: Do not store explosive substances such as aerosol cans with a flammable propellant in this appliance.
  • Page 7: Product Overview

    3. PRODUCT OVERVIEW MED0220GD-NR MED0374GD-NR 3736217-NR 3736218-NR Feature Unit, Brand, Description Data Data Height, mm 1310 1840 External Dimension Width, mm Depth, mm Height, mm 1025 1555 Internal Dimension Width, mm Depth, mm Capacity Litres Weight Air Circulation Fan-Forced Fan-Forced Climate Class Range +2°C to 8°C...
  • Page 8: Installation

    4. INSTALLATION NOTE: The maximum ambient temperature for the refrigerator is 30°C. Conditions outside this temperature will result in cabinet temperatures outside the required 2 to 8°C range. Bromic recommends that the refrigerator is installed at an ambient temperature of 20 to 25°C and relative humidity of 55%. Locating the Cabinet Ensure the following instructions are observed as this is vital for proper and trouble-free operation.
  • Page 9 Power Supply • The appliance must not be plugged into a multiple outlet power board. • The appliance is supplied with a flexible power cord fitted with a 3-pin plug. • The socket used must be properly earthed and fuse protected. •...
  • Page 10: Operation

    5. OPERATION Start-up • Connect the appliance to the mains power supply and check that the compressor, electronic controller, lighting and all fans are functioning correctly. • The compressor and condenser fan should operate should operate 3 minutes after the unit has been connected to the power supply.
  • Page 11 MAX/MIN temperature recording IMPORTANT: Record the maximum and minimum temperatures two (2) times a day. Recordings should be taken before the first door opening of the day. MIN Temperature Press the arrow key. The ‘Lo’ message will be displayed followed by the minimum temperature recorded. iii.
  • Page 12: Product Loading And Storage

    6. PRODUCT LOADING AND STORAGE WARNING: Failure to follow the guidelines below will result in cabinet temperatures below 2°C and above 8°C IMPORTANT: Run the refrigerator for at least 2 hours before loading. Start-up • Clean the interior surface with a soft cloth and warm water before loading. •...
  • Page 13: Maintenance

    7. MAINTENANCE WARNING: Disconnect the appliance from the power supply before performing any maintenance. Never unplug the appliance by pulling on the cord. Always grasp the plug firmly and pull straight out of the outlet. WARNING: Exposure to significant water ingress will void warranty and cause potential safety issues.
  • Page 14: Troubleshooting

    8. TROUBLESHOOTING IMPORTANT: Repairs on the unit must be carried out by an authorised and licensed person only. Contact the manufacturer or an approved service agent for any repairs needed. Symptom Possible Cause(s) Unit does not work A. No power supply A.
  • Page 15: Environmental & Disposal Notes

    9. ENVIRONMENTAL AND DISPOSAL NOTES These refrigerated cabinets are designed to high standards of environmental protection and recycling convenience. Cyclopentane, an environmentally friendly foaming agent, has been used. Cyclopentane is distinguished by zero ozone depletion potential (ODP) and zero global warming potential (GWP).
  • Page 16: Warranty

    10. WARRANTY Read all relevant documentations before operating the refrigerator and storage of any stock. Failure to follow instructions may result in loss of or damage to stock. BROMIC CANNOT BE RESPONSIBLE FOR THE LOSS OF OR DAMAGE TO STOCK Please refer to the ‘National Vaccine Storage Guidelines: Strive for 5’...
  • Page 17 HOW TO PLACE A SERVICE REQUEST • Please call 1300 276 642 to place a warranty service request and let the operator know whether you need Priority A or B service. More information can be found below. • Ensure you have the serial number of your product ready before contacting us. •...
  • Page 18 NOTES...
  • Page 20 BROMIC REFRIGERATION 10 Phiney Place, Ingleburn (Sydney) NSW 2565 Australia T. 1300 276 642 For further product information bromicrefrigeration.com.au BROMIC REFRIGERATION 10 Phiney Place, Ingleburn (Sydney) NSW 2565 Australia T. 1300 276 642 SYDNEY HEAD OFFICE & SHOWROOM For further product information 10 Phiney Place, Ingleburn bromicrefrigeration.com.au NSW 2565 Australia...

This manual is also suitable for:

Medifridge medo374gd-nr

Table of Contents