Download Print this page

Contec CMS50F User Manual page 2

Hide thumbs Also See for CMS50F:

Advertisement

After setting , short press the button to select
"
Exit "
then long press the button to exit the system
menu
menu
.
return to
main
.
5.3
5 Bluetooth
setting
Under
main menu
,
short press
the button
to
select "
Bluetooth
"
then
long
press the button
to
selection interface
as
shown in Figure 10 and Figure 11
.
When the Bluetooth is " ON '
if there is
no
.
transmitted , the Bluelooih will turn off automatically after
a
few seconds
Note
under data transmitting
state
by Bluetooth
,
the Bluetooth
can
not be turned
off
.
:
Turn
On
BT ?
Turn
Off
81
?
Q
E 3
Yes
Yi
' h
Figure 10 Bluetooth " ON " interface
Figure
11 Bluetooth " OFF " interface
.
5.3
6 Shutdown
Under the main
menu
interface
short press the button
to
select " Power off
1
'
then long
press the button
,
the device
.
turn
off
5.3
.
7 E
il
main menu
\
press the button
then long press the button to exit the
Under
main
menu ,
short
to
select
"
Exit "
main
and
the
measurement
interface
.
return to
5.4
Data upload
.
5.4
1
Wired transmission
Connect
the
device
to
computer by
the USB cable
upload the
data
after connecting
with the PC
,
.
properly ,
refer to
"
Software
operating instruction " for
details
.
Note : Ihe PC software can be downloaded from our official website
.
5.4
2 Bluetooth wired transmission
ou the device
Bluetooth
and
the PC
software
upload
data
refer
Software operating instruction
Turn
to
,
to
"
for d
etails
.
5.5
Charging
Power
adapter
can
be selected to charge for Ihe device
.
It
indicates that the device is charging when the indicator is orange , the charging is
finished
indicator turns
green
to
.
5.6
Reset
Use
a pointed
a paper clip
press
and hard
object
( for
example
,
) to
Ihe reset button inside of the RESET
to reset the device
.
6.1
Cleaning
and Disinfecting
The device must
be turned
off before cleaning
and it should
not be
immersed into liquid
.
Please take
,
.
internal battery before cleaning , do not immerse it into liquid
Use
75 % alcohol to wipe the device enclosure , and
use
liquid
soap or
isopropanol to wipe the watchband for
.
disinfection , nature dry
or
clean it with clean and soft cloth
Do
noJ
spray any
liquid
on
the device directly
and avoid
penetrating into the device
.
liquid
6.2
IVIainlcnancc
.
A
Check the main unit and all accessories periodically to make sure that there is no visible damage that may
affect
user
'
s
safety and monitoring performance
.
It
is
recommended lhai
the device
should
be inspected
.
.
weekly
at
least
When
there
is obvious damage , stop using
it
B
.
clean and disinfect the
fore
after
User Manual
6.1 )
.
Please
device be
/
using it according to the
(
C
.
Plcasc charge the battery
in
lime when low battery appears
.
D
.
Recharge the battery soon after over
-
discharge
The device should be recharged every three months when
.
it
is not used for some time
.
It can extend the battery life following this guidance
.
.
.
E
The device
need not to be
calibrated during
maintenance
6.3 Transportation and Storage
.
A
The packed
device
can
be
transported by ordinary conveyance or according
io
transport contract
During traiLsporlation
, avoid
strong shock
, vibration and
splashing
with rain or snow , and it can not be
transported
mixed with toxic , harmful ,
material
.
.
.
B
The
packed
device should be stored in room with no corrosive gases and
good
vcnlilation
Temperature
-
40 ° C
~
60 ° C ; Humidity
:
<
95
Possible Reason
Trouble
Solution
.
The finger is not properly
1
1
.
Please insert the finger properly and
.
positioned
.
measure
again
.
2
The finger
is
shaking
or
the user
The values
can
not be
2
.
Lcl the user keep calm
is moving
.
.
displayed normally
or
3
Please
use the device in
normal
.
3
The device
is not
used in
stably
environment required by the
-
4
.
Please
contact ihe after
sales
.
.
manual
4
.
.
Thc
device works
abnormally
. Low
1
battery or the battery is
.
The device can not be
1
Please charge the battery
.
drained away
.
-
turned on
2
Please contact
the
after
sales
.
.
.
.
2
The device works abnormally
.
1
The device enters into the
-
I
.
Please
he
after
sales
.
conlacl
(
display disappears
energy
saving mode
.
The
.
2
Please charge the
battery
.
.
.
suddenly
2
Low
battery
-
.
.
3
Please
conlacl
(
lie after
sales
3
.
The device works abnormally
.
The device can not
be
1
.
The battery is not charged
fully
.
1
.
Please charge the battery
.
used for full time after
-
.
.
.
.
.
2
The device works abnormally
2
Please
contact
the
after
sales
charge
The battery
can not
be
full charged even after
The battery works abnormally
.
Please contact the after
-
sales
.
10
hours charging
.
lime
1 .
1 .
The device is
not
operated
Please operate the device according to
can not
The data
be
according
the manual
.
the manual
.
lo
.
stored
2 .
.
2 .
.
The device works abnormally
Please contact the after
-
sales
Symbols
Meaning
Symbols
Meaning
©
Caution
consult
,
PRbpm
Pulse rate
bpm
)
(
accompanying
documents
Pulse oxygen saturation
(
i
Fully charged
Manufacturer
.
-
Serial mimhei
Use
by date
and
Recycling garbage
USB
enter its
WEEE
(
2012 / 19 / EU )
data to be
Battery
anode
Battery
It means
this pulse
is protected
IP 22
against harmful effects of
RST
RESET hole
dripping
water when tilled
15
°
.
at
Temperature limitation
Humidity limitation
II
Atmospheric pressure
0
This way up
limitation
to
r
menu
Close the sound
software
O
Pause Ihe sound
prompt
Open the sound
'
-
d x
0
Menu / Power button
Close ihe
when the
Recyclable
Open
,
hole
Bluetooth icon
Manufaclure Dale
Bluetooth
(
wired device )
out
the
Sensor
Off
The probe is disconnected
.
Finger Out
The
Sensor
R *
Recording
Probe failure
,
Fault
code
P
/
N
Material
LOT
Batch No
.
finger clip falls off
1
The
.
2
Probe
error
Alarm inhibit
3
.
Signal inadequacy
indicator
.
Note
Your device may
not
contain all the following
symbols
.
sPo
|
see note 1
]
2
-
Display range
0 %
99
%
:
Measured range
0
%
~
100
%
-
70 %
100 %
:
2 %
;
Accuracy
|
see
note
2 ]
-
0
%
69
%
unspecified
:
.
Resolution
1
%
.
PR
range
Display
30 bpm 250 bpra
~
Measured
range
30 bpm
250 bpm
± 2
bpm during the pulse rate
range of 30
Accuracy
see note
31
|
± 2
% during the pulse rate range
Resolution
Low
perfusion 0.4 % :
Accuracy
under low
perfusion [ see
note
41
PR
:
± 2 bpm during the pulse rate range
and
2 %
during the
pulse
rate
range of 100 bpm
der normal and ambient light conditions , the SpO
Light interference
%
-
Continuous
bar
graph display , the higher display indicates the
Pulse Intensity
.
stronger pulse
Upper
lower
of
and
limit
measured values
SpO
0
%
100
%
-
PR
0 bpm
254 bpm
sensor
see note
51
Optical
|
Red light
Wavelength : about 660
nm
, optical output
.
Infrared light
Wavelength
:
about 905
nm
optical
Up
to
99 group
of
data under auto mode ,
not
.
Memory
exceed 72 hours
-
Up to
24
liour
data under
manual mode
% )
,
Safety class
Internally powered equipment
type BF applied
International Protection
22
1 P
-
Working voltage
DC 3.6
V
4.2
V
Working
current
< 100
mA
A rechargeable lithium battery
Power supply
battery denotes anode
,
the black wire
cathode
life
Battery
Charge and discharge
Output voltage
Adapter specification
Output current
Dimensions and Weight
cathode
Dimension
61 mm
Weight
About 60 g
>
Note 1 : the claims of SpC
accuracy shall be
supported by clinical study measurements taken
2
range
.
get the stable oxygen level
By artificial inducing
,
Sp
values collected
by
the secondary standard pulse oximeter
()
time
,
to form paired data
,
which
used for
equipped
.
.
There
are 12 healthy
volunteers
(
male
:
6
female
:
.
data in the clinical report
Note 2 : because
pulse oximeter equipment measurements
pulse oximeter equipment measurements can be expected
-
.
CO
OXIMETER
Note
3
:
Patient simulator has been used
to
verily the
difference between the PR
measurement value and the value
prompt
Note 4
percentage modulation of infrared signal
as
the indication of
:
simulator has been used to
verify
ils accuracy under conditions
prompt
different due
to
low signal conditions
compare Ihcm with the known SpO
,
-
Note
5
: optical sensors as the light
emitting components
wavelength
range
The
information may be useful for the
.
PR
sound
example , photodynamic
therapy
operated
by
clinician
the PR sound
State
Prompt condition delay
Low vo
I
tag cprompt
Is
SpO
:
prompt
330 ms
finger
is not inserted
.
Pulse rate prompt
330
ms
Probe
error
16
ms
prompt
.
EMC
Tabic
1
:
Guidance
and
manufacturer
'
s declaration
The Pulse Oximeter is intended for
use
in the electromagnetic environment specified below
user
of the device should
assure
that it is used in such environment
Emission test
Compliance
RF
emissions
C
1
SPR
11
Group
I
Table 2
:
Guidance and manufacturer ' s dcclaratiou
The Pulse Oximeter is intended for
use
in the electromagnetic environment specified below
user
of ihe Pulse Oximclcr should
assure
lhat it is used in such environment
test
IEC 60601 test
Immunity
Electrostatic discharge
ESD
±
8
kV contact
(
)
IEC 61000
4
2
15 kV air
-
-
50
60
Power frequency
(
/
Hz
)
bpm
99 bpm and
magnetic field
30 A /
m
-
.
of
100 bpm
250 bpm
IEC 61000
4
8
-
-
Table 3
:
Guidance and manufacturer '
s
declaration
-
The Pulse Oximeter is intended for use
in
the electromagnetic environment specified below
of
30
bpm
99 bpm
.
user of the Pulse Oximeter should assure that
it
is used
250
bpm
Immunity test
IEC 60601
test level
deviation
?
-
Radiated
RF
10
V
/
m
80 MHz
2.7 GHz
-
IEC 61000
-
4
3
NOTE
80
MHz and
800
MHz , the higher
frequency range applies
1
Al
NOTE 2 These guidelines may
not apply
in all situations
absorption
and
reflection from structures
,
objects
and people
a
from fixed
as
Field strengths
transmitters , such
base
mobile radios
amateur radio ,
AM and FM radio broadcast and TV broadcast
,
.
with accuracy
To assess the electromagnetic em ironment due
power
< 6.65 mW
.
survey should be considered
If
the
measured field
strength in the location in
power :
<
6.75 mW
oulput
exceeds the applicable
RF
compliance level
above ,
the Pulse Oximeter should be observed
total
duration docs
.
operation
If abnormal performance is observed
,
additional
relocating the Pulse Oximeter
.
.
h Over the frequency range 150 KHz
80 MHz
field
to
,
pari
V
on
( 3.7
)
( The red
the
wire
on
the ba
ncry
denotes
Table 4 :
.
)
Guidance
and
: no
than
less
500 times
The [ Code SI ] is intended for use in the electromagnetic environment specified below
:
DC 5 V
user of the Pulse Oximeter should assure
:
1000 mA
Test
Band
Freque
a
: 56
)
L
W
x
24 mm
(
H
(
)
>
mni
(
)
)
ncy
(
MHz
( MHz
)
(
including a lithium battery
)
380
over
the full
385
.
390
range of 70
compare the
to the
% to
100
%
SpO
equipment
and the tested equipment
the
al
.
he accuracy analysis
( li
is
applicable for the probes
(
430
450
-
-
470
,
,
6 ; age
:
18
50
;
skin color
:
black
: 2
light
:
8
while
: 2
)
Radiated
RF
710
-
are
statistically distributed , only about
thirds of
hvo
IEC 610
745
704
to
fall within
Amis
of the value measured by a
-
-
-
00
4
3
787
780
Test
(
,
-
-
pulse
rate accuracy
it is stated
as
the root
mean
square
sped flea
810
.
set by simulator
tions for
870
ENCLO
pulsating signal strength
,
patieut
800
SURE
of
low perfusion
.
SpO
and
PR
values
are
?
PORT
.
930
:
and
PR
values of input signal
1 MMUN
will
affect other medical devices applied the
,
ITYto
clinicians
who
carry
out
the optical treatment
For
.
RF
1720
.
wireless
1845
comm
u n
1700
ications
equipme
-
1990
Prompt signal generation delay
1970
nt )
20 ms
20 ms
20
ms
20
ms
2400
2450
2570
-
5240
5500
5100
-
electromagnetic emission
5800
-
5785
.
The purchaser
or
Ihe
.
If
necessary to achieve the IMMUNITY TEST LEVEL
NOTE
and the
ME
EQUIPMENT
antenna
.
pcnnitlcd by
IEC
61000
-
4
-
3
a
)
For some services
, only tlic uplink frequencies
b
)
The carrier shall be modulated
c
an alternative
FM modulation
)
As
to
-
clcctromagnctic immunity
not rq
rcscnt actual modulation
^
.
The
purchaser or
Ihe
The MANUFACTURER should consider reducing the minimum separation distance , based
.
RISK MANAGEMENT ,
and
reduced
separation distance
level
Compliance level
LEVELS
shall
be ailculated using the following equation
±
8
kV
contact
15 k
V
air
Where
P
is the maximum power in W , d is Ihe minimum
IMMUNITY TEST LEVEL in V /
30 A /
m
arning
'
near
active HFSURGICAL EQUIPMENT
Don
t
electromagnetic irmnunity
for magnetic
resonance
.
The customer the
Use of this
equipment adjacent
could result
in
improper operation
in
such environment
.
equipment
should be observed to
Compliance level
Use of accessories
,
transducers and cables other than those
manufacturer of this equipment
-
10
V
/
m
80 MHz
2.7 GHz
decreased electromagnetic immunity
Portable RF communications equipment
external
antennas
)
should
.
including cables specified by
.
Electromagnetic propagation is affected by
this equipment could result
.
Active medical devices
used
in
accordance
vilh these guidelines
stations for
radio ( cellular/cordless
)
telephones and land
A
cannot be predicted
theoretically
Note
to
fixed RF
transmitters
electromagnetic
site
, ail
When the device is disturbed
which
the Pulse Oximeter is used
another environment to
to
verify
normal
Bluetooth Specification
measures may
be necessary , such
as
or
reorienting
frequency : 2402 MHz
Working
Modulation mode
Transmitting
strengths should
be
less
than
3
V /
m
.
Receiving sensitivity :
-
manufacturer
'
s
declaration
electromagnetic
Immunity
.
The customer
or the
that i
(
is used
in
such
an
environment
I MM
UN
I
Modul
TY
Modulation
Distance
a
Service
)
alien b
)
TEST
b )
( m )
)
(
W
)
LEVEL
V /
m
)
(
Pulse
modulation
TETRA 400
K
8
0
,
3
27
b )
18 Hz
FMc
)
GMRS
,
5 kHz
460
±
2
0
,
3
28
FRS 460
deviation
1
kHz
sine
Pulse
LTE
Band
modulation
02
0
,
3
9
13
,
17
b )
217 Hz
GSM
800 / 900
,
Pulse
TETRA
modulation
2
0
3
28
800
,
,
b )
iDEN 820
,
18
Hz
CDMA
850
,
LTE
Band
5
GSM 1800
;
CDMA
1900
Pulse
;
GSM 1900 ;
modulation
2
0
3
28
,
DECT
;
b
)
.
LTE Band
217 Hz
1
3
,
4
,
25
;
UMTS
Bluetooth ,
WLAN ,
Pulse
802.11
modulation
2
0
,
3
28
b
/ g
n ,
b
/
)
RFID 2450
,
217 Hz
LTE
Band 7
Pulse
WLAN
modulation
-
802.11
0
2
0
3
9
,
b )
a
/
n
217 Hz
the distance between the
,
transmitting
or
ME
SYSTEM
may be reduced
to I
The I
m test
distance is
are
included
.
.
using
a
50
%
duly
cycle
square wave signal
50
modulation at 18 Hz may
used because while it does
,
% pulse
be
, it
would be worst case
.
on
using
higher IMMUNITY TEST LEVELS that
are
appropriate for the
.
Minimum separation distances for higher IMMUNITY TEST
M
:
=
E
d
separation
distance in ra , and
E
is the
m
.
KF shielded room of an
ME SYSTEM
and the
.
,
imaging
where the intensity of EM
DISTURBANCES
is high
to or
stacked with other equipment should be avoided because it
.
,
If such
use
is necessary
this equipment and the other
are
.
verify
that they
operating
normally
specified
or
provided by
the
could result
in increased electromagnetic emissions
or
.
of
this equipment and result
in
improper operation
"
(
including peripherals such
as antenna
cables and
be used
no
closer
than
30 cm
(
12
inches
) lo
any part of the device
.
manufacturer
Otherwise
degradation of Ihe performance of
Che
,
.
are
subject to special EMC precautions and Ihey
must
be installed and
.
,
(
he data measured
may
fluctuate
please measure repeatedly or in
ensure
accuracy .
its
2480 MHz
:
GFSK
power
:-6
dBm , + 4 dBm
-
93 dBm

Advertisement

loading
Need help?

Need help?

Do you have a question about the CMS50F and is the answer not in the manual?