MAINTENANCE AND CARE
Always turn off the power switch and allow the unit to cool before cleaning it.
Be careful not to scratch yourself on the unit's protrusions.
Supplying and draining water for evaporation
After each sterilization cycle, replenish the water for evaporation.
Be sure to drain away the left over water for evaporation at the
end of each day.
After draining the water, be sure to close the drain valve.
The water for evaporation is still hot immediately after finishing of
sterilization. Wait for a while until it becomes cool water before
draining it. If a hose is connected to the drain opening, this hose
should be made of heat resistant material.
Replacing the water inside the exhaust tank )
^Before using the unit, check the water level of the exhaust tank.
If it has reached the HIGH level mark, discard this water and add
new water up to the LOW level mark. If the inside of the tank
becomes dirty, pour in a mixture of neutral detergent and warm
water, and shake well to remove the dirt. Then rinse out the tank
thoroughly.
If the main unit or the chamber interior become dirty
•To remove dirt from the main unit, wipe it with a soft cloth mois-
tened with water or neutral detergent and wrung out. Never use
solvents such as paint thinner or benzine as they may mar the
paint's finish.
>Wipe the interior of the chamber to clean using a soft cloth mois-
tened with water. If the chamber is very dirty, it should be
cleaned using a soft cloth moistened with neutral detergent, and
then rinsed with water.
Clean the bottom of the chamber thoroughly using a brush with a
handle or similar implement.
Be careful not to rub too forcefully to avoid damaging the sensor
element of the heat sensor mounted on the heater.
Drain off the water used to rinse the chamber through the drain
opening.
( Cleaning of the door gasket and chamber opening )
Should the door gasket or chamber opening become dirty, steam
leakage can result. They should therefore be wiped carefully with
a moistened cloth each day before using the unit.
Drain valve
Drain opening
Heat resistant
material
Heater
Sensor element
of the heat
sensor
(MLS-3020U)
Chamber opening
Door packing
28