PROJECT 25 TRUNKING OPERATING INSTRUCTIONS
When testing a radio, the analyzer generates a control channel signal. Take care to prevent this
signal from unintentionally capturing other radios in the area. Observe the following precautions:
Do not use an antenna on the analyzer for over-the-air testing.
Use double-shielded cables on the analyzer to carry signals to and from the radio.
Locate the analyzer at least 35 feet from the antenna of a unit that is working in the
same system that the analyzer is testing.
Adjust the squelch to where the LED indicator for squelch just turns off or is
closed. When the signal from the radio is present, the squelch LED will illuminate
indicating that squelch has been detected and there is a signal present.
24-1 INTRODUCTION
The Project 25 Trunking Option is an
enhancement to the R2670 and R2625
Digital Communications System Analyzers.
Refer to sections 1, 2, and 3 under the
General Operation tab in this manual for
general
installation
description of the control functions, and
general
operational
following sections of this manual contain
information on screen field definitions and
test setup.
Error/Warning Messages
Refer to Appendix M-3 for a listing and
description of error and warning messages
related to the Project 25 trunking test mode.
Messages common to all test modes are
described in paragraph 2-4 under the General
Operation tab of this manual.
CAUTION
information,
a
information.
The
Section 24
24-2 SOFTARE VERSION SCREEN
To view the software version of the R2670
or R2625 Analyzer:
1. Turn power on and wait for the display
to appear on the screen.
2. Press the SPF hard key, and move the
cursor to "VERSION".
3. Select the display table softkey.
will configure the analyzer to generate a
screen that displays the Project 25
software version (figure 24-1).
4. Move the cursor to the Project 25
(PRJ25) position and select the view
options softkey.
figure 24-2 will be displayed.
screen indicates the analyzer options
installed.
5. Select the return softkey to return to the
power up screen.
292
This
A screen similar to
This