Self-Tests - HP 16532A Service Manual

1 gsa/s digitizing oscilloscope
Table of Contents

Advertisement

HP 16532A - Service

6-7. Self-Tests

The self-tests for the HP 16532A oscilloscope module will identify the improper operation of major func-
tional areas in the module. They are not intended for component level diagnostics.
If there are multiple oscilloscope modules, the user must select the specific oscilloscope module to be
tested at the main test system menu. Multiple HP 16532A modules that are connected as one unit are rec-
ognized as a single unit in the test menu. The external trigger cables connected to the ECL IN and ECL
OUT do not need to be removed in order to do the self-test for each board. All other cable connections
must be removed.
All self-tests can be run without access to the interior of the instrument. If a self-test fails, refer to
section 6-8, "Troubleshooting."
Self-Test Access Procedure:
1. Disconnect all oscilloscope inputs and turn on the power switch.
2. From the startup screen, shown in figure 6-2, touch the Configuration field. Then, touch Test in the
pop-up menu.
6-4
Figure 6-2. Startup Screen

Advertisement

Table of Contents
loading

Table of Contents