4 - Procedure a) Pressure generator models CH1-125B and CH2-125B ................11 b) Pressure generator model CH1-1250B ......................12 c) Pressure generator models CH2-1250B and CH2-2000B and CH2-4000B and CH2-7000B and CO-800B and CO-1000B ..............................14 5 - Cleaning/Maintenance a) Cleaning ............................... 15 b) Maintenance ..............................
Page 4
RE. ENR.110 - Rév B – le 15/03/2014 page : 4...
Chap.2 - Security a) Device Device Identification indicates the model of comparator ex : CH2-2000B indicates the serial number of comparator indicates the date of commissioning of the unit Intended use of the device Before switching on the device, ensure that the technical characteristics (measuring range, precision, fluid etc...) correspond to your needs >...
Page 7
> wear gloves : protect the skin against irritation, allergies. AREMECA provides for oil dead weight, a bottle of colorless mineral oil. Safety data sheet of used oils are available on request. RE. ENR.110 - Rév B – le 15/03/2014...
- 1 manual Check that there isn't damage or breakages during transport. If there are missing items, please immediately notify AREMECA or your local distributor for what is missing. b) Installation Preferably, install your unit in a stable and controlled environment You will get better performance if : >...
- Operations on the previous page are identical then : - Set up the connectors of the pressure generator : • G1/2 provided with o-ring (aluminium, copper or two-component) for maximum operating pressure 1250 bars. • Connect directly to the output M16x150 for pressure > 1250 bar >...
4. Turn the capstan 1 in the direction of clockwise until reaching the end stop before running to purge air from the system. 5. Turn the capstan 1 dans le sens inverse des aiguilles d’une montre jusqu’à atteindre la butée de fin de course arrière.
Chap.4 - Procedure Procedure for pressure generator models CH1-125B and CH2-125B > Hydraulic diagram > Pressure mounted > Observe the maximum operating pressures of the pressure generator Maximum operating pressure : CH1-125B 125 bar and CH2-125B 125 bar a) Connect the instrument to be tested b) Check the valve A is open c) Check the capstan 1 the pressure generator is in end position stop.
Procedure for pressure generator models CH1-1250B > Hydraulique diagram The model CH1-1250B have 2 valves A and A’ for multiple round and return of the piston to fill the devices to be tested. The valve A’ shall be used only for filling, it must remain open. >...
Page 13
> Reduce pressure o) After the readings measures, turn the capstan 1 in the opposite direction clockwise until the end position of the stop and slowly open the valve A p) The instruments can be disassembled. RE. ENR.110 - Rév B – le 15/03/2014 page : 13...
Procedure for pressure generator models CH2-1250B and CH2-2000B, CH2-4000B and CH2-7000B and CO-800B and CO-1000B > Hydraulic diagram > Pressure mounted > Observe the maximum operating pressures of the pressure generator Maximum operating pressure : C0-800B 850 bar and CO-1000B 1050 bar CH2-1250B ...
Chap.5 - Cleaning/Maintenance a) Cleaning The generator cleaning and checking fluid levels are the only periodic maintenance required. No additional maintenance should be necessary if all instructions are followed. We recommend a complete overhaul of the unite every 2 years. This operation can only be performed by the manufacturer.
> Option : cleaning kit Ref. OP0025 you will aspirate oil durty present in the tank. Don't clean the tank with alcohol c) Inspection It's recommend to a service of your device every 2 years. The average response time is 2 weeks. Cleaning generator, replacement o-rings, testing pressure rise and stability are made.
Need help?
Do you have a question about the CO-800B and is the answer not in the manual?
Questions and answers