Page 1
Required Material and Equipment Sci-Print MP2 or MP2+ 2mm allen wrench 3mm allen wrench Intended Audience Certified Scinomix Technician Prerequisites Procedure 1. Power off instrument 2. Turn air supply to instrument off 3. Using a 2mm allen wrench remove the four screws for the swing arm tower housing, and set housing aside.
Page 2
TS-1078 Tamp Swing Arm Document Owner: Effective Date: Page 2 of 5 Replacement Service Team AUG-24-2022 (MP2/MP2+) Figure 1. 4. Mark or take pictures of the air lines and flow controls, to record the order and placement for the install.
Page 3
TS-1078 Tamp Swing Arm Document Owner: Effective Date: Page 3 of 5 Replacement Service Team AUG-24-2022 (MP2/MP2+) 6. Unplug the flex cable from the board. Figure 3. 7. Using a 2.5mm allen, remove the two screws for the flex cable/airline bracket.
Page 4
TS-1078 Tamp Swing Arm Document Owner: Effective Date: Page 4 of 5 Replacement Service Team AUG-24-2022 (MP2/MP2+) 8. Using a 3mm allen, remove the top shoulder screw and guard. Figure 5. 9. Using a 2.5mm allen, remove the middle shoulder screw holding the arm to the tower.
Page 5
TS-1078 Tamp Swing Arm Document Owner: Effective Date: Page 5 of 5 Replacement Service Team AUG-24-2022 (MP2/MP2+) Figure 7. Figure 8. 12. Replace the middle shoulder screw to secure the arm to the tower. 13. Replace top shoulder screw and guard.