Wahl DST400 User Manual page 19

Dstcal software
Table of Contents

Advertisement

Wahl DSTCAL Software User Manual
Build Date: Identifies date firmware was compiled.
Build Time: Identifies time firmware was compiled.
Hardware Revision: Identifies Hardware revision level. This parameter is hard coded
into firmware.
Serial number: Identifies serial number of PCB, programmable on DSTCAL-Pro only.
LCD Refresh: Sets update period for the LCD. 1 period = 250 mS, thus, LCD Refresh:
2 = 500 mS or 0.5 second update rate for LCD. Default value is 8.
Low Battery Check: Sets update period for checking of low battery. 1 period = 250
mS, thus Low Battery Check: 1200 = 300,000 mS or 5 minute interval for battery
checking.
Thermocouple Type – Allows selection of J, K, E, T or S thermocouples. Changing of
the thermocouple type requires the recalibration of the Cold Junction Compensation
using the newly selected thermocouple type cable and input source. See section XXXX
for instructions. Selection will display the following Caution:
Calibration Information – Thermocouple Models
Module Calibration Data – Thermocouple Models
mV Cal. Slope – Slope correction value for millivolt measurement as calculated by
measurements during the Millivolt Calibration process. Displays active values used for
the measurement.
mV Cal. Offset – Offset correction value for millivolt measurement as calculated by
measurements during the Millivolt Calibration process. Displays active values used for
the measurement.
Select Calibration to Perform: