Exchanging the Power Supply Fuses

Exchanging the Power Supply Fuses
The fuse holders are located on the rear panel of the thermostat.
When
If wrong fuses are installed.
Tools required
Description
Flat head screwdriver
Parts required
p/n
2110-0015
Damaged electronics
C A U T I O N
Disconnecting or reconnecting the autosampler/ fraction collector to thermostat cable
when the power cords are connected to either of the two modules will damage the
electronics of the modules. In such a case, mainboards of both instruments must be
exchanged, otherwise they can damage the other instrument.
58
Description
Fuses T2.5 A/250V (CSA, UL listed)
Make sure the power cords are unplugged before disconnecting or reconnecting the
autosampler/ fraction collector to thermostat cable.
1290 Infinity Thermostat User Manual