Magnamed OxyMag Operation Manual page 56

Transport and emergency ventilator
Hide thumbs Also See for OxyMag:
Table of Contents

Advertisement

1 and 2 Represent spontaneous cycles with pressure support at ZERO.
3
, 4 and 5
Represent spontaneous breathing cycles of the patient with pressure support different from zero. TRISE TIME of
pressure support may be adjusted so that initial flow is smoothed.
6 If the patient enters in apnea, after TAPNEA (s) the ventilator will show this condition with alarm in its message
screen and will initiate backup ventilation selected as settings and parameters programmed.
Apnea alarm should be set to a safe value for the patient. However, apnea alarm can be RESET, in this
condition, there will be no information or alarm for apnea condition and no backup ventilation in action. The
equipment operator must be aware of the Apnea Alarm condition DEACTIVATED (OFF INDICATING ON
DISPLAY).
If backup ventilation selected is NO BACK-UP, the equipment operator must be aware of this situation
(INDICATING ON DISPLAY).
Default values are only for initial reference. Reset the ventilation parameters as needed by the patient.
Note
Pressure support (ΔPS) is a value above PEEP and can be adjusted between + 5 cmH2O and PMAX - PEEP.
To obtain CPAP mode with backup ventilation, select option CPAP/PSV, set ΔPS=OFF and select backup
ventilation.
56
Manual Oxymag_Rev22
area
and alarms on