Installation And Disassembly Process; Installation Location; Product Handling And Package Removal - Haier DW-86L338J Service Manual

Ult freezer, biomedical after-sales
Hide thumbs Also See for DW-86L338J:
Table of Contents

Advertisement

A
A -----------Condenser
C -----------One stage compressor
E ----------- Gas-liquid separator
G ----------- Pressure switch
I ----------- Two stage compressor
XI. Installation and disassembly process

11.1 Installation location:

11.1.1 Avoid exposure to direct sunlight
11.1.2 The surrounding air is well-ventilated
11.1.3 Avoid excessive dust
11.1.4 Avoid mechanical swing or vibration
11.1.5 The ambient temperature is between 10°C and 28°C; the maximum temperature is not more than 32°C. The
optimal temperature is between 18°C and 25°C, and the air conditioning system shall be used if necessary.
11.1.6 Height of equipment working position: lower than 2000 m
11.1.7 Working humidity: lower than 80% RH If the maximum working temperature is 32°C, the humidity should be
lower than 57% RH.
11.1.8 Input voltage: Within 220±10%.
*Note: Because the ULT Freezer is sensitive to the ambient temperature, if installed in an environment other than
the above, it will not operate normally. Please use the freezer after improving the environment.

11.2 Product handling and package removal

11.2.1 Handling: The product is very heavy, so forklift or pallet jack shall be used to avoid damage to personnel and
ULT Energy-saving Series
B
E
C
After-sales Service Manual
B
D
F
I
G
H
B -----------Dry filter
D -----------Capillary tube
F -----------Heat exchanger
H -----------Oil separator
J----------- Evaporator