Summary of Contents for Flyability ELIOS 2 RAD

  • Page 1 FLYABILITY ELIOS 2 RAD USER MANUAL VERSION 1. 2 05.05.2022 FLYABILITY SA +41 21 311 55 00 — INFO@FLYABILITY.COM EPFL INNOVATION PARK BLDG C | 1015 LAUSANNE | SWITZERLAND...
  • Page 2: Table Of Contents

    Cockpit Interface Flight Display Out of range behavior Settings Menu Post flight analysis with Inspector 3 Calibration Setting up RiumConfig Calibrating the Elios 2 RAD sensor module 7.2.1 Connect the detector module to a windows PC 7.2.2 Irradiation type 7.2.3 Position 7.2.4...
  • Page 3: Before Starting

    1 Before starting This manual focuses on the features of the dosimetry payload of Elios 2 RAD. It complements the following documents, which the Elios 2 RAD user must also read and understand fully: – ELIOS 2 Original Instructions – certification compliance, safety guidelines and checklists –...
  • Page 4: Elios 2 Rad

    3 Elios 2 RAD Elios 2 RAD is a specially modified Elios 2 designed to carry a dosimetry detector module. Apart from the mounting bracket for the detector module, and the modified casing with a connector, all user changeable parts are compatible with a standard Elios 2. There is no significant change in flight behavior and collision resilience, though flight time is reduced slightly due to the added weight of the sensor module.
  • Page 5: Detector Specifications

    The LOW dose detector can also be user recalibrated to work with lower dose rate (down to 3.5uSv/h). The range and linearity shall be assessed by the user. ELIOS 2 RAD USER MANUAL Page 5 of 16...
  • Page 6: Cockpit Interface

    5.2 Out of range behavior The Elios 2 RAD detectors will provide accurate readings within their specified range. Below the lower limit of the sensor, the dose rate value will be displayed in orange, and the message “Too low” will appear.
  • Page 7 ELIOS 2 RAD USER MANUAL Page 7 of 16...
  • Page 8: Settings Menu

    Session Accumulated Dose, the total dose absorbed by the drone since the battery is connected Lifetime Accumulated Dose, the total dose absorbed by the drone over its lifetime before this session Connected sensor, this indicates the range of the detector currently connected. ELIOS 2 RAD USER MANUAL Page 8 of 16...
  • Page 9: Post Flight Analysis With Inspector

    The lower limit of the detector is indicated in orange and the upper (saturation) limit in red. If a sensor disconnection was detected, the value will be equal to -1 in the graph ELIOS 2 RAD USER MANUAL Page 9 of 16...
  • Page 10 ELIOS 2 RAD USER MANUAL Page 10 of 16...
  • Page 11: Calibration

    When launching the program, you will be given the option to create a password (mot de passe means password). This is not needed for the Elios 2 RAD sensor module, but we recommend you do it anyways, because otherwise it will just keep bothering you with it. You may encounter parts of the program that are not translated, to help you we show you the main translated parts here below.
  • Page 12: Calibrating The Elios 2 Rad Sensor Module

    Low level = X If the Elios 2 RAD sensor module is not connected to your PC at this point, then you will get this message This message means: “No sensor, verify the connection and try again”. This can be due to a faulty connection, or due to the drivers not being installed correctly.
  • Page 13: Irradiation Type

    fit with the center on the sensor. (see pictures) ● The position of the center of detection is indicated by a gray dot on the casing, as shown in red on the image below. ELIOS 2 RAD USER MANUAL Page 13 of 16...
  • Page 14: Measurements

    Once the calibration factor is adjusted, we suggest validating an uncertainty below 10% from measurements at 20%, 50% and 80% of the global dose rate range. Irradiation duration must be adjusted in function of the dose rate to reach a sufficient counting statistic. ELIOS 2 RAD USER MANUAL Page 14 of 16...
  • Page 15: Updated Technical Specifications

    Detector contains lead. Applications exempted from the restriction in Article 4(1) COMPLIANT specific to medical devices and monitoring and control instruments. DIRECTIVE 2011/65/EU OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL ELIOS 2 RAD USER MANUAL Page 15 of 16...