the systems. For the machine that its system has already been opened, the pipe welding 
shall be finished at once.   
 
As  the  machines  have  strict  requirement  for  system  moisture  content,  so,  when  a 
machines pipe is opened, the opening of pipe shall be sealed with tape at once to prevent 
air  from  entering  in  system,  and  pipe  welding  of  opening  shall  be  finished  within  20 
minutes.   
 
As  the  system  of  DW‐86L386  series  of  ultra‐low  temperature  cabinet  products  has  very 
strict  requirement  for  vacuum  degree,  the  system  will  occur  bad  refrigeration  or  non‐ 
refrigeration due to insufficient vacuum. Therefore, it is required that maintenance shop 
shall  spend  more  time  in  vacuumizing  this  series  of  products  than  normal  refrigerators; 
vacuum pump above two liters needs more than 6 hours, vacuum pump below two liters 
needs more than 8 hours. To ensure vacuum degree, when one system is being repaired, it 
is required that two vacuum pumps (high and low pressure) shall vacuumize separately at 
the same time, pressure gage and gage valve shall be installed on high and low pressure 
infusion tubes. 
 
When  evacuation  time  has  been  reached,  firstly,  stop  high  pressure  evacuation,  and  the 
opening of high temperature infusion tube shall be sealed and welded immediately(as the 
pressure  of  freezing  medium  is  too  high,  when  a  machine  is  refrigerating,  high  pressure 
can not seal the opening), low pressure evacuation shall continue for an hour. During this 
period, prepare freezing medium needed for infusion and relevant equipment. After one 
hour  low  pressure  evacuation,  turn  off  pressure  gage  valve;  now,  it  is  required  that  the 
machine  shall  keep  the  state  of  negative  pressure  for  more  than  one  hour  to  check  if 
machine  system  is  in  vacuum  state  or  not.  During  this  period,  check  if  pressure  gage 
appears  the  condition  of  rising,  otherwise,  re‐vacuumize  for  more  than  one  hour,  then 
keep  the  state  of  negative  pressure  for  more  than  one  hour,  carry  out  such  evacuation 
operation until pressure doesn't rise, then infusion can be carried out. 
 
 