TABLE OF CONTENTS Section Page Introduction Product Description Tests Performed Specifications Model 535-H Target Diagram Dead Zone Vertical Measurement Calibration Horizontal Measurement Calibration Axial and Lateral Resolution Focal Zone Sensitivity (Maximum Depth of Penetration) Functional Resolution, Definition and Fill-in Care and Handling of Rubber Based Phantoms...
The acoustic properties of all biologic and non-biologic materials are affected by temperature variations. Most diagnostic imaging systems and tissue-mimicking phantoms are calibrated at room temperature, commonly referred to as 23°C. To ensure measurement accuracy ATS incorporates a thermometer strip affixed to the outside surface of the phantom housing.
DEAD ZONE Description and Reason For Testing The dead zone is the distance from the front face of the transducer to the first identifiable echo at the phantom/patient interface. In the region of the dead zone no clinical data can be collected. The dead zone occurs because an imaging system cannot send and receive data at the same time.
Testing Procedure Note: The Model 535-H Basic QA phantom provides two scanning surfaces used to evaluate horizontal measurement calibration. Linear array scanning systems should use #1 scanning surfaces. Sector scanning systems should use #3 scanning surfaces.
Testing Procedure In the Model 535-H scan surface #1 and #3 are used to evaluate axial resolution at depths of 5.5 and 11 cm. To evaluate lateral resolution, scan surfaces #2 and #4 are used, at depths of 4.5 and 16.5 cm.
(geometry) of the reflecting object, and electromagnetic interference from the local surroundings. Testing Procedure The Model 535-H is provided with two sets of anechoic target structures, at two depths from scan surface #1. The ability to use all four scanning surfaces allows measurements to be obtained at 6 depths.
Testing Procedure The Model 535-H is provided with two sets of anechoic target structures, at two depths from scan surface #1. The ability to use all four scanning surfaces allows measurements to be obtained at 6 depths. 1. Place the phantom on a clean, flat surface with scanning surface #1 positioned for use.
Phantom is free from functional defects in materials and workmanship. The lifetime of this phantom is estimated to be 10 years from the date of manufacturing. If ATS Laboratories, Incorporated, deems the phantom to be defective, at its sole option, the Phantom will be repaired or replaced free of charge, in a reasonable amount of time.