IRadimed MRidium 3860+ Reference Manual

IRadimed MRidium 3860+ Reference Manual

Mri infusion pump
Hide thumbs Also See for MRidium 3860+:

Advertisement

MRidium
MRidium
3860
3860
MRI Infusion Pump
MRI Infusion Pump
® ®
+ +
Reference Guide
Reference Guide
www.IRadimed.com
www.IRadimed.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the MRidium 3860+ and is the answer not in the manual?

Questions and answers

Summary of Contents for IRadimed MRidium 3860+

  • Page 1 MRidium MRidium 3860 3860 MRI Infusion Pump MRI Infusion Pump ® ® Reference Guide Reference Guide www.IRadimed.com www.IRadimed.com...
  • Page 2: Initial Setup

    Initial Setup Initial Setup 1.1| Selecting the Proper IV Set 1.2 | Proper Flow Clamp Positioning for Priming the IV Set a. Select the 1056 IV Set when infusing from a bag or bottle fluid source (18mL). b. Select the 1057 IV Set when infusing from a syringe fluid source (15mL). Pull Through Push In Ready to Prime...
  • Page 3 Initial Setup Initial Setup 1.3 | Power ON the Pump 1.5 | Loading the IV Set a. Insert the round Disk on the IV set into the “Disk Slot” on top of the pump. Press the ( I ) button as shown to power on the pump. b.
  • Page 4 Before Programming Channels A or B Programming Channel A 2.1 | Selecting Infusion Type: Drug Libraries A or B 2.2 | Selecting Medication: Drug Library A Use Library A for Channel A, and Library B for Channel B a. Press the UP or DOWN arrow keys to scroll through listed medications.
  • Page 5 Programming Channel B Programming Basic Infusions 2.3 | Selecting Medication: Drug Library B 3.1 | Rate Mode: Channel A a. Press the MENU button. a. Press the softkey by “New Patient- Rate Mode” on Main Menu Screen. b. Press Dose Rate Cal. B (Reference 2.1) c.
  • Page 6 Programming Basic Infusions Titrating Basic Infusions 3.2 | Rate Mode: Channel B 3.3 | Rate Mode: Channel A or B During Infusion In Progress a. Press the softkey by “New Patient- Rate Mode” on Main Menu Screen. a. Press the softkey by Channel A or B at the bottom of the screen. (as shown in step 2.1) b.
  • Page 7: Programming A Bolus

    Programming a Bolus KVO (Keep Vein Open) 4 | Proper DERS Bolus Programming 5 | Resuming Infusion During KVO a. To program a bolus, press Bolus Softkey A/B or a. Press Start/Stop A or B (depending on channel in use) to press Menu Button.
  • Page 8: Alarm Messages

    Alarm Messages Alarm Messages 6.2| PT Occlusion Alarm First, Review ALARM Messages, then Press the ALARM SILENCE ALARM SILENCE Alarm will sound if occlusion between the pump and patient occurs. button before proceeding. Check between the pump and patient to assure that All Clamps have been Opened and fluid flow is permitted.
  • Page 9: Alert Messages

    - During Bolus Setup: Check “Time” settings and adjust as necessary. Typically the “Time value needs to be reviewed and decreased if appropriate. _____________________________________________________ If ALARMS persist or other technical issues not mentioned in this guide, Remove Pump from use and contact your facility’s Biomedical Department and/or IRadimed Technical Support at: 1-866-677-8022 ___________________________________________________...
  • Page 10 Full product technical specifications are available upon request. Scan For Refer to MRidium1138 Operation Manual for further instruction. Information subject to change. Training Videos ©IRadimed Corporation 2020 - All rights reserved LB026-A 1025 Willa Springs Drive, Winter Springs, FL 32708 | (407) 677-8022...