Thermocouple thermometers

Advertisement

Activate or deactivate the keys beep
The device is on and displays the measurement.
Press "Selection" until "BEEP" blinks on screen. Press "OK".
"OFF" or "ON" blinks on screen.
Press "Selection" to activate the keys beep "ON" or deactivate it "OFF".
Press "OK" to confirm the selection.
"BEEP" blinks on screen.
Press "On/Off/Esc" to return to the measurement display.
Set the alarm
This function allows to activate a temperature alarm on channel 1 when the device reaches or exceeds a previously
defined threshold. This threshold is between -200 and +1760 °C according to the thermocouple type of the probe.
The device is on and displays the measurement.
Press "Selection" until "AL 1" blinks on screen.
Press "OK".
The first digit blinks.
Press "Selection" to select the value of the first digit according to the thermocouple type of the probe: -, 0, 1 or OFF.
If "Off" is selected, the alarm is deactivated. Press "OK" to confirm, "AL 1" blinks on screen.
If -, 0 or 1 is selected, press "OK". The next digit blinks, press "Selection" to select its value then press "OK" to confirm.
Perform the same procedure to select the value of the following digits (press "On/Off/Esc" to return to the previous
digit). Press "OK" when the last digit is set, "AL 1" blinks on screen.
Press "On/Off/Esc" to return to the measurement display.
Activate the backlight
The device is on.
Press "Backlight" to activate the device backlight.
Press again this key to deactivate the backlight.
Turn off the device
To turn off the device:
Carry out a long press on "On/Off/Esc".
Replace the batteries
Remove the front part at the back of the device.
Change the old batteries by AAA LR03 1.5V batteries.
Replace the front
Safety instructions
In order to avoid any risk of a voltage on the surface of the element on which we want to perform
a temperature measurement, the thermocouple(s) MUST NOT be under a voltage higher than 30 V
alternating or 60 V direct voltage relating to earth.
In order to avoid any damage on the device, the voltage between the two thermocouples MUST NOT
be higher than 0.5 V.