2.

Assembly

2.1.

Constituent parts

Fig. 1: SV-Adaptor Assembly
2.2.

Installing instructions

First connect the two wires with the electronic device. Then mount the senor adaptor to the EX II 2GD electronic
and tighten it with 32 Nm. Secure the assembly by a high temperature adhesive.
As soon as this connection is solid the electronics can be assembled in the sensor hole or thread socket. Make
sure that the sensor gets assembled to correct sensor connections which are stated in the type code of the
sensor adaptor.
Screw in the adaptor gently by hand until you feel that it touches the ground. Then turn it back for a quarter turn.
Secure the position with the counternut and make sure that the wedge lock washer is in between. This washer
is necessary to make sure that the counternut can not get loose while the system is under vibration for example.
The necessary torque for the counternut is 32 Nm.
Orientate the electronic device in the way you need it and secure the position by fastening the setscrews, which
are freely accessible, with a torque of 2.9 Nm.