1. Getting started

Press ON/OFF key to turn on/off the device

2. Temperature selection

Press TEMP/MEM to activate the temperature selection function
Press MEASURE to select the required temperature,
from 0°C to 25°C in steps of 5°C
Validate your temperature setting by pressing ON/OFF Key

3. Measurement

Fill 50 mL of wine to measure into the test bottle
and close it with rubber septum and the aluminum cap
Note: to ensure the best repeatability, it is recommended
to use a graduated cylinder of 50 mL to fill the flask
Shake the bottle vigorously for 10 seconds
Open the valve of the syringe in order to expulse pressure
from the bottle when piercing the septum
Pierce the septum with the syringe
Close the valve as soon as the needle is positioned
in the headspace of the test bottle
Press MEASURE to launch measurement
Result appears on the display after 10 seconds
Note: Before the next measurement, evacuate all CO
particles
of the flask by filling it totally with water
Quick Guide
gas
2
1
2
3

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Wine Quality Solutions NOMASense CO2 P2000 and is the answer not in the manual?

Questions and answers

Summary of Contents for VINVENTIONS Wine Quality Solutions NOMASense CO2 P2000

  • Page 1 Quick Guide 1. Getting started • Press ON/OFF key to turn on/off the device 2. Temperature selection • Press TEMP/MEM to activate the temperature selection function • Press MEASURE to select the required temperature, from 0°C to 25°C in steps of 5°C •...
  • Page 2 • Calibration is required when a known content of CO is measured and value deviates by more than 0.1 - 0.2 g/L Please, contact Vinventions WQS for calibration service. 5. Battery charging • When the display shows LoBAT ; charge the battery •...

Table of Contents