User Manual I. Disclaimer & Warranty • The intellectual property rights for this manual belongs to Holland Green Science Corporation (hereinafter referred to as “the Company”). • Products of the company are under the patent protection of China and other countries and regions (inclusive of obtained patents and patents currently under application).
Page 4
User Manual the discovery of any defect, which notice shall include the product model and serial number (if applicable) and details of the warranty claim; and (b) after the Company's review, the Company will provide Buyer with service data and/or a Return Material Authorization ("RMA"), which may include biohazard decontamination procedures and other product-specific handling instructions, then, if applicable, Buyer may return the defective Products to the Company with all costs prepaid by Buyer.
Page 5
User Manual THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT. EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT, THE COMPANY DISCLAIMS ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, ORAL OR WRITTEN, WITH RESPECT TO THE PRODUCTS, INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE OR USE.
User Manual LIMITATION OF LIABILITY NOTWITHSTANDING ANYTHING TO THE CONTRARY CONTAINED HEREIN, THE LIABILITY OF THE COMPANY UNDER THESE TERMS AND CONDITIONS (WHETHER BY REASON OF BREACH OF CONTRACT, TORT, INDEMNIFICATION, OR OTHERWISE, BUT EXCLUDING LIABILITY OF THE COMPANY FOR BREACH OF WARRANTY (THE SOLE REMEDY FOR WHICH SHALL BE AS PROVIDED UNDER THIS SECTION) SHALL NOT EXCEED AN...
User Manual II. Safety IMPORTANT Safety Items to be Observed. Safety Warnings ! DANGER (may cause serious damage to property and or casualties) • Please carefully read this User Manual prior to operating the instrument and observe the instructions on safe operation. •...
Page 8
User Manual regulations. ! ATTENTION (may affect operational performance or service life) • The instrument is deenergised only when the power cable is disconnected from the power source. • Ensure the instrument and its accessories are protected external vibration. • The Company reserves the right to modify the design and technical data of the pump without notice.
User Manual After Sales Support If problems are encountered or technical support is required when installing or using the instrument, please contact serviceusa@hollandgreenscience.com The company may provide technical assistance and information regarding the instrument or equipment or service without charge at its sole discretion. Buyer assumes sole responsibility for any reliance on or use of such assistance and information, and the company makes no warranty thereon.
III. Anemos Oil-free Scroll Vacuum Pump 3.1 Introduction The Anemos oil-free scroll vacuum pump uses oil-free scroll compression technology. It has a low leakage rate, and a high compression ratio. The vacuum pump offers a steady pumping speed over a wide pressure range. Because the change of the compression volume is continuous, the driving torque and power are almost constant.
User Manual Precautions • The vacuum pump should not be used to remove acid or alkaline solutions or corrosive gases. Such substances will cause the pump's fixed scroll, orbiting scroll, and other parts to be corroded and shorten the lifespan of the pump. The vacuum pump can be used in applications where organic solvents need to be removed.
User Manual IV. Technical Specifications 4.1 Specifications Model 11111008 Product Name Anemos Nominal pumping speed L/min mbar 0.008 Ultimate Vacuum mbar·l/s <1×10-6 Leak rate pa·m 3 /s <1×10 -7 Motor voltage (V) Single-phase 100-240 Motor power (W) Motor speed (rpm)
There should be adequate ventilation for the pump. 5.2 Electrical Connection The Anemos vacuum pump uses a single-phase power. Wiring method is shown below. Connect the power line to the power connector on the pump. Turn on the power switch (“1” is pressed and “0”...
User Manual 5.3 Operation • Before starting be certain that all sure seals are tightly secured. • If the pump is used for the first time or used after a long period of inactivity, open the inlet and outlet ports and run the pump for 5 minutes to remove any water vapor that might get into the compression chamber.
User Manual VI. Interface There is a 15-pin interface on the base of the pump for pump control and performance data monitoring. 6.1 Pin definition 15-pin male serial port (operation interface) No Signal Function Remark 5V power 5V power ground Serial communication 485+ 485+ Pressure signal 1-5V/0-10V...
User Manual VII. Inspection Unpack the equipment carefully and check for any damage which may have arisen during transport. In the event of identified damage, please contact serviceusa@hollandgreenscience.com The package includes the following items Item Description Quantity Main Unit Quick Setup Card Plastic cover Power cable NW25 Net filter (Including O-ring)
User Manual VIII. Maintenance When the pump is used for a specified period, some inspection and maintenance work is required. Otherwise, it will affect the performance and lifespan of the pump. Maintenance frequency Remark Inspection One year or 8000 Two years or 16000 part hours hours...
User Manual IX. Troubleshooting Fault Possible Reason Remedy Power supply protection Check if the protection component is component malfunction normal. Loose or broken wires Check for bad contact and repair, replace the problematic wire. Pump Not Motor overload or open Return to the service department for Starting circuit...