Percussionaire Phasitron Manual

Institutional and homecare cleaning

Advertisement

Quick Links

Institutional and Homecare
Cleaning Manual
Percussionaire® Phasitron® & Nebulizer

Advertisement

loading

Summary of Contents for Percussionaire Phasitron

  • Page 1 Institutional and Homecare Cleaning Manual Percussionaire® Phasitron® & Nebulizer...
  • Page 2 PERCUSSIONAIRE® CONTROLLER DEVICE Cold cleaning and disinfection (after basic cleaning of unit) of the surface of Percussionaire® controller & display device to be performed weekly, or as needed or if transferring between patients. PHASITRON® KIT & HARNESS ARE SINGLE PATIENT USE ONLY •...
  • Page 3 WARNING Read the label for the safe number of days that the Phasitron® kit may be used for your control unit, and ensure that you are using an appropriate Phasitron®. Not adhering to this may cause the Phasitron® to malfunction and cause ineffective therapy.
  • Page 4 NOTE Before cleaning or disinfection, the tubing harness must be removed from the controller device, Phasitron® and nebulizer. Remove tubing from the Percussionaire® controller device: 1. Disconnect white, yellow, red and green tubing connectors from device. Figure 1 2. Green and red tubing...
  • Page 5 (if applicable) quick- connector from the Figure 4a Phasitron® body. 4. If present, remove the antibacterial filter from the red tubing either side. Do Figure 5 not wash. • continued on page 6 Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 6 Phasitron® body. 7. Remove mouthpiece, mask, corrugated tubing or other patient interface from Figure 9 Phasitron®. NOTE All tubing should be disconnected from Percussionaire® controller device and Phasitron®. WARNING Tubing lines must not be connected during cleaning process. PERCUSSIONAIRE ®...
  • Page 7 Phasitron®: 1. Separate Nebulizer from Figure 10 Phasitron®. 2. Holding the ring on the nebulizer cap, twist the nebulizer bowl to remove Figure 11,11a bowl from cap. • continued on page 8 Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 8 3. Remove white cap on the rear of the Phasitron® by unscrewing in a counter- Figure clockwise motion. 12,12a 4. Remove sliding Venturi (and spring, when present) from Phasitron® body. Figure 13 PERCUSSIONAIRE ®...
  • Page 9 (no white dish soap or anti- bacterial dish soap may be used). 3. Hand wash all parts of the Phasitron® kit and accessories in the warm soapy water. 4. Use a soft bristle, nylon brush to remove stubborn soil.
  • Page 10 2. Pour bleach solution into an airtight container large enough to hold 2 liters or 64 ounces. 3. Completely immerse all the Phasitron® parts and tubing harness (including mouthpiece, mask, and other accessories) into bleach solution. Do not disinfect antibacterial filter.
  • Page 11 3. Lightly coat the outer edge of the sliding Venturi (Figure 15) diaphragm and the yellow ring on the aerosol bowl with the provided Percussionaire® lubricant lubetube. Figure 15a WARNING Only use Percussionaire® approved lubricant. Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 12 Reassembly of the Phasitron® and tubing harness: 1. Holding the ring of the nebulizer cap twist nebulizer bowl into top Figure 17,17a until it locks. • continued on page 13 NOTE Ensure that the inner stem on the diaphragm is pressed firmly into Figure 18 the venturi.
  • Page 13 2. Visually inspect O-ring on sliding Venturi for cracks or damage. Insert Venturi and spring (if applicable) Figure into Phasitron® body. 19,19a • continued on page 14 Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 14 3. Screw white cap onto rear of the Phasitron® body in Figure clockwise motion. 20,20a 4. Add mouthpiece, mask, corrugated tubing or other patient interface to the Figure 21 Phasitron®. PERCUSSIONAIRE ®...
  • Page 15 Figure 23 nebulizer. 3. Turn on device and run for two minutes at a time to remove moisture. NOTE Not all tubing harnesses have (require) the check valve. Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 16 Reattach the tubing harness to Phasitron®: 1. Connect yellow tubing quick connector to Figure 24 nebulizer bowl. 2. If there is a check valve, it is important to ensure that the direction of arrow on check valve points towards Figure 25 the nebulizer.
  • Page 17 5. Connect the green tubing (if applicable) quick connector to the top of the Figure 29 nebulizer. 6. Connect the white tubing quick connector to the rear of the Phasitron® body. Figure 30 Institutional & Homecare Cleaning Manual Phasitron® & Nebulizer...
  • Page 18 Reattach the tubing harness to Percussionaire® Controller device: 1. Connect red tubing quick connector to red on device Figure 31 (if applicable). 2. Connect white tubing quick connector to white/silver Figure 31 on device. 3. Connect green tubing quick connector to green on device (if applicable).
  • Page 19 1. Use either an approved wipe or spray a suitable agent onto a damp, lint free cloth and wipe the surface of the device (do not spray any type of cleaning liquid onto the Percussionaire® controller device). Do not submerge any Percussionaire® controller device.
  • Page 20 Percussionaire® Corporation 130 McGhee Rd., Ste. 109, Sandpoint, Idaho 83864 MDSS GmbH, Schiffgraben 41, 30175 Hannover, Germany   percussionaire.com +1.208.263.2549 0123 LIT-MAN-CLEAN Rev A...