2 - Verification and Performance Tests
Table 2-1. Test Equipment Required for Verification and Performance Tests
Type
Current Monitor Resistor
DC Power Supply
Digital Voltmeter
Electronic Load
GPIB Controller
Resistors
(Load resistors may
substitute for electronic
load if load is too noisy
for CC PARD test)
Oscilloscope
RMS Voltmeter
Variable-Voltage
Transformer
Measurement Techniques
Test Setup
All tests are performed at the rear terminals of the supply as shown in Figure 2-1. Measure the dc voltage directly at
the +S and -S terminals. Set the Remote/Local switch to Remote and connect the output for remote sensing. Use
adequate wire gauge for the load leads.
12
Specifications
15 A (0.1 ohm) 0.04%
Minimum 5 A output current rating
Resolution: 10 nV @ 1V
Readout: 8 1/2 digits
Accuracy: 20 ppm
100V, 5 A minimum, with transient capability
Controller with full GPIB capabilities
400 ohm, 5W
1 ohm, 100 W (or 2 ohm adjustable)
0.6 ohm, 100W (6611C)
9 ohm, 100W (6612C)
49 ohm, 100W (6613C)
99 ohm, 100W (6614C)
or an appropriate 150W Rheostat
Sensitivity: 1 mV
Bandwidth Limit: 20 MHz
Probe: 1:1 with RF tip
True RMS
Bandwidth: 20 MHz
Sensitivity: 100 V
Adjustable to highest rated input voltage
range.
Power: 500 VA
Recommended Model
Guildline 9230/15
Agilent 6632B
Agilent 3458A or equivalent
Agilent 6060B (60V max.), 6063B
(240V) or equivalent
HP Series 300 or equivalent
Agilent p/n 0811-1857
Ohmite D12K2R0 (2 ohm adjustable)
Agilent 54504A or equivalent
Agilent 3400B or equivalent