§4 SUBTROUBLESHOOTING ............. 4-1 4.1 Power On Indicator does not Light Up..............4-1 4.2 Cooling Device does not Start...................4-2 4.3 MC-300 does not Start....................4-3 4.4 An Interlock (INTLK) Number is Displayed..............4-4 4.4.1 INTLK 1.6 is displayed..................4-5 4.4.2 INTLK 2.7 is displayed.
Page 4
4.5.20 ERR 59 is displayed..................4-30 4.5.21 ERR 80 is displayed..................4-31 4.5.22 ERR 81 is displayed..................4-32 4.5.23 ERR 82 is displayed..................4-32 4.5.24 ERR 83 is displayed..................4-32 4.5.25 ERR 90 is displayed..................4-32 4.5.26 ERR 91 is displayed..................4-32 4.5.27 ERR 93 is displayed.
Page 5
6.18 Shutter ASSY. (Dual) .....................6-10 6.19 Coagulation Shutter ASSY..................6-10 6.20 Sensor and Solenoid of Coagulation Shutter ASSY..........6-11 6.21 Interface Board ..................... 6-11 6.22 Dual Shutter ASSY....................6-11 6.23 Cooling Device ......................6-12 6.24 Master Board ......................6-13 6.25 Laser Unit......................6-13 6.26 Water Filter ......................6-15 6.26.1 New type of water filter ................6-15 6.26.2 Old type of water filter ..................6-16...
• For correct service, thorough understanding of the contents in this manual is required prior to the service. • Refer to the operator’s manual and parts list for the MC-300. • Specifications are subject to change without notice for improvement. Refer to the Technical Bulletin (T.
• Never emit laser beam (direct beam) toward a human body or object. <Adjustment precautions> • After turning the main switch off, let the MC-300 sit for 5 minutes or more before turning it on again. • When the MC-300 is moved, let the system sit for 30 minutes or more before turning the main switch on.
Are the indications on the remote control normal 4.4 An Interlock Number is Displayed. after countdown? 4.5 An Error Number is Displayed. Does the MC-300 have the ERR or INTLK 4.4 An Interlock Number is Displayed. history? 4.5 An Error Number is Displayed.
Page 12
XMC20*RDA001B/E Does pressing the emergency off switch turn off 4.9 Emergency Off Switch does not Work. the laser, remote control, and compressor? Turn the key switch to the OFF position. Turn the master switch off. 3 - 2...
XMC20*RDA001B/E § 4 SUBTROUBLESHOOTING 4.1 Power On Indicator does not Light Up. Are the fuses blown? Replace the fuses (see 6.12). Is the voltage between the P1 3rd and 5th pins on Check the wirings of the master switch ASSY. the switching power supply (17350-E004) inside (17350-EA32) and transformer ASSY.
Replace the fuses on the relay board (see 6.3). blown? Reset the circuit breaker. Is the circuit breaker of the built-in cooling device If the breaker trips again, return the MC-300 to the tripped? factory. Is the voltage between the P312 (J312) 1st and Replace the relay board (see 6.3).
Do the controller and switches have any problem? Replace the cooling device (17350-E001) (see 6.23). 4.3 MC-300 does not Start. Is there continuity between the P304 1st and 2nd, and 4th and 5th pins on the key switch ASSY. Check the soldered spot. When in doubt, replace (17350-EA35) when the key switch is turned to the key switch ASSY.
XMC20*RDA001B/E 4.4 An Interlock (INTLK) Number is Displayed. Is INTLK 1.6 (overheat of the LD power supply) 4.4.1 INTLK 1.6 is displayed. displayed? Is INTLK 2.7 (excessive output) displayed? 4.4.2 INTLK 2.7 is displayed. Is INTLK 2.8 (excessive output - output motor/ 4.4.3 INTLK 2.8 is displayed.
(17350-EA39), repair or replace it (see 6.15). When there is a break in LD driver fan ASSY. 2 Do Fans 7 and 8 in the MC-300 underside rotate? (17350-EA40), repair or replace it (see 6.16). Turn the key switch to the OFF position and...
TP17 voltage become as below? GRN:750, YEL/RED:500 GRN: 6.2 V, YEL/RED: 4.3 V Replace IC24 on the master Replace the LD power Return the MC-300 to the factory. board (17350-BA01) and the supply (17350-E005) master board (see 6.24). (see 6.5).
TP17 voltage become as below? GRN: 750, YEL/RED: 500 GRN: 6.2V, YEL/RED: 4.3V Replace the LD power Replace IC24 on the master Return the MC-300 to the factory. supply (17350-E005) board (17350-BA01) and the (see 6.5). master board (see 6.24).
Replace the master board (see 6.24). 4.4.5 INTLK 3.5 is displayed. Check the compressor error ASSY. (17350-EA52). Turn the master switch off. In five minutes, turn it on. If the symptom occurs, return the MC-300 to the factory. 4 - 8...
Does the water leak? Repair the leakage spot. Reset the leakage breaker. Is the leakage breaker of the cooling device If the breaker trips again, return the MC-300 to the (17350-E001) tripped? factory. Are the operating noises of the cooling device Replace the cooling device (see 6.23).
XMC20*RDA001B/E 4.4.7 INTLK 5.1 is displayed. Does the deflecting plate drop off? Mount the half-wave plate (17350-G008). Does the treatment beam leak even though the Adjust the coagulation shutter position. coagulation shutter closes? Is the TP20 voltage on the master board (17350- BA01) 2 V or less when the setting is GRN and output of 2000 mW? I s t h e a i m i n g b e a m o u t p u t w i t h i n t h e...
Disconnect and connect IC1 (master CPU) on the master board. Replace the master board (see 6.24). Does INTLK 1.6 occur when the MC-300 starts? 4.4.9 INTLK 9.9 is displayed. Disconnect and connect IC1 (master CPU) and IC23 (sub CPU) on the master board (17350- BA01).
XMC20*RDA001B/E 4.5 An Error (ERR) Number is Displayed. Is ERR 1 (coagulation shutter malfunction) 4.5.1 ERR 1 is displayed. displayed? 4.5.2 ERR 2 is displayed. Is ERR 2 (protective filter malfunction) displayed? 4.5.3 ERR 3 is displayed. Is ERR 3 (safety shutter malfunction) displayed? Is ERR 4 (EEPROM data output/time failure) 4.5.4 ERR 4 is displayed.
Page 25
XMC20*RDA001B/E Is ERR 18 (current limit failure - The current is 4.5.11 ERR 18 is displayed. higher than the limit.) displayed? Is ERR 25 (M1 initial error - LH color changeover) 4.5.12 ERR 25 is displayed. displayed? Is ERR 26 (M2 initial error - coagulation output 4.5.13 ERR 26 is displayed.
Page 26
XMC20*RDA001B/E Is ERR 80 (DPRAM time error) displayed? 4.5.21 ERR 80 is displayed. Is ERR 81 (DPRAM time error) displayed? 4.5.22 ERR 81 is displayed. Is ERR 82 (DPRAM check error) displayed? 4.5.23 ERR 82 is displayed. Is ERR 83 (DPRAM check error) displayed? 4.5.24 ERR 83 is displayed.
XMC20*RDA001B/E 4.5.1 ERR 1 is displayed. Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Are there any breaks in the coagulation shutter Replace the coagulation shutter ASSY. (see ASSY. (17350-EA13)? 6.20). Set DIP SW 2 on the control box to the OFF position.
XMC20*RDA001B/E 4.5.2 ERR 2 is displayed. Connect the delivery connector again. Is the symptom improved? Connect the filter connector again. Is the symptom improved? Set DIP SW 2 on the control box to the OFF Remove the filter unit. Check that the filter is not position.
XMC20*RDA001B/E 4.5.3 ERR 3 is displayed. Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Is there a break in the aiming shutter ASSY. Replace the aiming shutter ASSY. (17350-EA14)? Set DIP SW 2 on the control box to the OFF position.
XMC20*RDA001B/E 4.5.4 ERR 4 is displayed. Set DIP SW 2 to the OFF position, turn the key switch to the ON position, change the power output, and set DIP SW 2 to the ON position. Is the symptom improved? Disconnect and connect IC1 (CPU) and IC8 on the master board (17350-BA01).
RED: 0 to 1 V DC +4 to 5 V Replace the laser head or return the MC-300 to the factory (see 6.25). Disconnect and connect C23 (ROM) on the master board. Replace the motor driver (17350-E160) on the Is the symptom improved? master board.
XMC20*RDA001B/E 4.5.6 ERR 7 is displayed. Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Does the voltage between TP9 and TP10 on the Does the solenoid activate when CH1 and CH2 master board (17350-BA02) switch between DC + are switched? 15 V and -15 V each time the channel switch is pressed?
Replace the master board (see 6.24). Is ERR 10 always displayed in the READY state? Return the MC-300 to the factory. Set DIP SW8 to the ON position and disconnect the delivery connector. At this time, is the d2 p Return the MC-300 to the factory.
Adjust VR8 so that the voltage between TP10 and master board (17350-BA01) 0.1 V or less when A_GND on the master board becomes 0 V ±0.1 V. the aiming is off? Return the MC-300 to the factory. 4 - 22...
Set DIP SW 8 on the control box to the OFF position. Disconnect the delivery connector. Is the d2 p Return the MC-300 to the factory. value the same as the output setting? Adjust GND: VR14, YEL: VR15, and RED: VR16 Disconnect the delivery connector.
Replace the LH signal cable ASSY. ASSY. (17350-EA53)? Is the operating noises of the LH motor audible Replace the laser head or return the MC-300 to when the DIP SW 1 and 2 are set to the OFF the factory (see 6.25).
Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Are there any breaks in the coagulation motor Return the MC-300 to the factory. ASSY. (17350-EA11)? Are there any breaks in the C-motor driver ASSY. Replace the C-motor driver ASSY. (see 6.6.1).
Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Are there any breaks in the aiming motor ASSY. Return the MC-300 to the factory. (17350-EA12)? Are there any breaks in the A-motor driver ASSY. Replace the A-motor driver ASSY. (see 6.6.2).
Supply the cooling water (distilled water or pure limit of the cooling device water meter? water). Reset the leakage breaker. Is the leakage breaker of the cooling device If the breaker trips again, return the MC-300 to the tripped? factory. Replace the cooling device (17350-E001) (see 6.23).
XMC20*RDA001B/E 4.5.17 ERR 53 is displayed. A r e t h e t w o s w i t c h e s a c t i v a t e d a l m o s t Adjust the activation timing for the switches at the simultaneously when the foot switch is pressed? bottom of the foot switch (see 7.29).
XMC20*RDA001B/E 4.5.18 ERR 54 is displayed. Disconnect the fiber optic cable of CH1 and connect it again. Is the symptom improved? Are there any breaks in the interface cable ASSY. Replace the interface cable ASSY. (17350-EA57)? Select CH1. Is the voltage between TP13 and GND, and TP14 and GND on the interface board (17350-BA02) Replace the fiber optic cable switch ASSY.
Are there any breaks in the Peltier driver ASSY. Replace the Peltier driver ASSY. (17350-EA56)? Are there any breaks in the LH signal cable Replace the LH signal cable ASSY. ASSY. (17350-EA53)? Return the MC-300 to the factory. 4 - 30...
XMC20*RDA001B/E 4.5.21 ERR 80 is displayed. Disconnect and connect IC1 and IC23 (CPU) on the master board (17350-BA01). Replace the master board Is the symptom improved? (see 6.24). 4 - 31...
XMC20*RDA001B/E 4.5.22 ERR 81 is displayed. See “4.5.21 ERR 80 is displayed”. 4.5.23 ERR 82 is displayed. See “4.5.21 ERR 80 is displayed”. 4.5.24 ERR 83 is displayed. See “4.5.21 ERR 80 is displayed”. 4.5.25 ERR 90 is displayed. See “4.5.21 ERR 80 is displayed”. 4.5.26 ERR 91 is displayed.
XMC20*RDA001B/E 4.5.27 ERR 93 is displayed. See “4.5.21 ERR 80 is displayed”. 4.5.28 ERR 94 is displayed. See “4.5.21 ERR 80 is displayed”. 4.6 Each Setting of Control Box can not be Changed. Are there any breaks in the flexible flat cable Replace the flexible flat cable.
Is S.U. displayed on the TIME display of the Connect the delivery unit to the main body. control box? Press the STATUS switch to put the MC-300 in Is S.S. displayed on the TIME display of the control box? READY.
XMC20*RDA001B/E 4.8 Actual Output is out of Standards. Disconnect the delivery connector. Are the d2 p 4.5.7 Err 10 is displayed. and d3 p values the same as each output setting? Is the end face of fiber optic cable soiled? Clean its end face.
XMC20*RDA001B/E 4.9 Emergency Off Switch does not Work. Are there any breaks in the emergency cable Replace the emergency cable ASSY. ASSY. (17350-EA34)? Is the continuity status of the P401 3rd and 4th, and 5th and 6th pins on the emergency switch ASSY.
XMC20*RDA001B/E § 5 REMOVING PROCEDURES 5.1 Covers 5.1.1 Left cover 1. Remove the screw caps [32906-M091 (n = 6)]. 2. Unscrew PC3 × 15 (n = 6) to remove the left cover (17350-M405). 3. Reassemble the parts in reverse order. 5.1.2 Right cover 1.
XMC20*RDA001B/E 5.1.3 Rear cover 1. Loosen SB3 × 6 (n = 3), pull out the top of the rear cover (17350-M403), and lift the cover to remove it. 2. Reassemble the parts in reverse order. 5.1.4 Fiber optic cable cover 1.
XMC20*RDA001B/E 5.1.5 Top cover 1. Remove the fiber optic cable cover (17350- 4100) (see 5.1.4). 2. Remove the rear cover (17350-M403) (see 5.1.3). 3. Unscrew PC3 × 5 (n = 4) to remove the top cover ASSY. (17350-4300). 4. Reassemble the parts in reverse order. 5.1.6 Front cover 1.
3. Disconnect P308 (J308) from the relay board (17350-BA03). 4. Unscrew SB3 × 4 (n = 8) to replace the two compressor fan assemblies (17350-EA37) on the right and left sides of the MC-300. 5. Reassemble the parts in reverse order. 6 - 1...
XMC20*RDA001B/E 6.3 Relay Board Replacement part: 17350-BA03 1. Remove the covers (see 5.1). 2. Remove the duct (see 5.2). 3. Disconnect all connectors from the relay board (17350-BA03). 4. Unscrew PC3 × 4 (n = 4) to replace the relay board.
XMC20*RDA001B/E 6.5 LD Power Supply Replacement part: 17350-E005 1. Remove the covers (see 5.1). 2. Remove the duct (see 5.2). 3. Disconnect all connectors from the LD power supply (17350-E005). 4. Unscrew PC3 × 4 (n = 4) to replace the LD power supply.
XMC20*RDA001B/E 6.7 Key Switch ASSY. Replacement part: 17350-EA35 1. Remove the front cover (see 5.1.6). 2. Remove the nut to replace the key switch ASSY. (17350-EA35). 3. Reassemble the parts in reverse order. 6.8 Line Indicator ASSY. Replacement part number: 17350-EA41 1.
XMC20*RDA001B/E 6.9 Front Panel ASSY. Replacement part: 17350-EA49 6.9.1 Channel switch 1. Remove the front cover (see 5.1.6). 2. Remove the crimp pin from P40 on the front panel ASSY. (17350-EA49). 3. Replace the two channel switches of the front panel ASSY.
XMC20*RDA001B/E 6.10 Delivery Unit ASSY. Replacement part: 17350-EA54 1. Remove the front cover (see 5.1.6). 2. Remove the ground wire from the delivery unit ASSY. (17350-EA54). 3. Unscrew FK2 × 4 (n = 4) to remove the delivery unit ASSY. 4.
XMC20*RDA001B/E 6.12 Fuses Replacement parts: 80402-02060 (100 V area) 80402-02043 (200 V area) 1. Turn the master switch off. Insert the precision screwdriver 2. Disconnect the power cord from the main body (flatblade) here and pry it. Then inlet and wall outlet. the fuse carrier will pop out.
XMC20*RDA001B/E 6.14 Emergency Switch ASSY. Replacement part: 17350-EA20 1. Remove the control box from the main body (17350-2000). 2. Unscrew PT3 × 8 (n = 4) to remove the lower box (17310-M203). 3. Disconnect the connectors of the emergency switch ASSY. (17350-EA20) from the control box cable ASSY.
XMC20*RDA001B/E 6.16 LD Driver Fan ASSY. 2 Replacement part: 17350-EA40 1. Remove the covers (see 5.1). 2. Remove the duct (see 5.2). 3. Disconnect P309 (J309) from the master board (17350-BA01). 4. Unscrew SB3 × 18 (n = 8) to replace LD driver fan ASSY.
XMC20*RDA001B/E 6.23 Cooling Device Replacement part: 17350-E001 1. Remove the covers (see 5.1). 2. Open the cock of the cooling device (17350- E001) to drain the cooling water. * Prepare a bucket and cloth. 3. Remove the hose under the relay block (17350-M321) on the right side when viewed from the front.
XMC20*RDA001B/E 6.24 Master Board Replacement part: 17350-BA01 1. Remove the covers (see 5.1). 2. Disconnect all connectors from the master board (17350-BA01). 3. Unscrew PC3 × 6 (n = 4) to remove the master board. 4. Remove the EEPROM (IC8) on the master board.
Page 68
XMC20*RDA001B/E 8. Remove 3 ch temperature control board (17350-E006) and disconnect all connectors on the Peltier driver (17350-E900). 9. Unscrew SB3 × 4 (n = 4) to remove the Peltier driver (17350-EA64). 10. Remove the master board (17350-BA01) (see 6.24). 11.
* Mark about 22 mm position from the hose end with a felt pen in advance. 8. Supply the cooling water (see 7.7). 9. Prepare to start the MC-300, turn the main switch on, and check the water does not leak from the hose connection.
XMC20*RDA001B/E 6.26.2 Old type of water filter Replacement par: 17350-9902 17350-9904 1. Remove the covers (see 5.1). 2. Unscrew SB3 × 10 (n = 2) to remove the old type of water filter ASSY. (17350-9902), filter attachment plate (17350-M372), and filter fixing plate (17350-M373) from the frame of the main body.
Page 71
SB3 × 10 (n = 2). 11. Supply the cooling water (see 7.7). 12. Prepare to start the MC-300, turn the main switch on, and check the water does not leak from the hose connection.
XMC20*RDA001B/E 6.27 Controller of Cooling Device Replacement part: 17350-9903 1.Remove the cooling device (17350-E001) (see 6.23). 2.Remove the top cover of the cooling device. 3. Loosen two outside screws, one inside screw, and one side screw to remove the controller along with the white sheet metal component on which the controller is attached.
Page 73
XMC20*RDA001B/E 4. Remove the wire rod which is wired in the controller. 5. Remove the plastic cover fixing the controller while releasing the clasps of the cover with a small flatblade screwdriver. * The plastic cover is divided into two. 6.
XMC20*RDA001B/E § 7 ADJUSTMENT 7.1 Master Board Check 1. Measure the resistance value with a digital Red probe Black probe Resistance (Ω) IC8 (EEPROM) t e s t e r ( r e s i s t a n c e r a n g e ) . P e r f o r m BA01 D.GND Not short : 8th pin...
XMC20*RDA001B/E 7.4 EEPROM Initial Setting 1. Make a record of all parameter settings. 2. Set DIP SW 1 to 6 on the control box (17350-2000) to the OFF position. 3. Turn the main switch on. 4. Turn the key switch to the ON position. 5.
* The target temperature is printed on the label affixed to the laser head ASSY. (17350-1100) side. * When the MC-300 stops, turn the power on again. 7.9 Cooling Device Idling Time Measurement * Perform the measurement only when the relay board (17350-BA03) is replaced.
7. Minimize the aiming beam output with the AIMING switch. 8. Set the treatment beam output to 700 mW. 9. Press the STATUS switch to put the MC-300 in READY and press the foot switch to emit the treatment beam.
XMC20*RDA001B/E 7.11.2 Second and third monitors adjustment 7.11.2.1 Adjustment in high power mode 1. Disconnect the fiber optic and delivery unit cables. 2. Set DIP SW 1, 2, 4, 6, and 8 to the OFF position. 3. Set parameters for Address 178 to 00 and Address 69 to 00. 4.
XMC20*RDA001B/E 7.11.2.2 Adjustment in low power mode 1. Connect a digital tester between TP1 and A_GND, and TP2 and A_GND on the master board (17350-BA01). 1) TP1: Second monitor 2) TP2: Third monitor 2. Set the color to GRN and turn the aiming off. Adjust VR13 and VR18 so that the voltage between TP1 and TP2 becomes DC 0 V.
XMC20*RDA001B/E 7.13 Overpower and Underpower Adjustment 7.13.1 Overpower and underpower adjustment of second monitor 1. Set DIP SW 1, 2, and 6 to the OFF position. 2. Set parameter for Address 1 to 01. 3. Set the color and output as shown in the right Monitor TP VR Color Power (mW) table and measure the TP1 and TP3 voltages...
XMC20*RDA001B/E 7.13.2 Overpower and underpower adjustment of third monitor 1. Set DIP SW 1, 2, and 6 to the OFF position. 2. Set parameter for Address 1 to 01. 3. Set the color and output as shown in the right Monitor TP VR Color Power (mW) table and measure the TP1, TP2, TP3, and...
XMC20*RDA001B/E 7.14 Aiming Monitor Origin Adjustment 1. Disconnect the fiber optic and delivery connector cables. 2. Install a power meter (LASER MATE Q) in the fiber optic cable connector. 3. Set DIP SW 1, 2, and 6 to the OFF position. 4.
XMC20*RDA001B/E 7.16 Fiber Optic Cable Alignment 1. Set the MC-300 as follows: Laser color: GRN, aiming beam output: minimum, treatment beam output: 200 mW 2. Emit the treatment beam on a wall at a distance of 0.5 to 1.0 m away from the end of the fiber optic cable.
XMC20*RDA001B/E 7.18 Aiming Beam Output Adjustment 1. Connect the slit lamp delivery unit to DELIVERY 1. 2. Install a power meter (LASER MATE Q) to the laser aperture of the delivery unit. At this time, the treatment beam should be out of focus. * Change the wavelength of the power meter depending on the laser color.
XMC20*RDA001B/E 7.20 Coagulation Time Setting 1. Connect any delivery unit to the MC-300. 2. Install an analyzing recorder (YOKOGAWA Model: 3656) to the laser aperture of the delivery unit. At this time, the treatment beam should be out of focus.
STANDBY. 7.23 Slit Lamp Delivery Unit Spot Size Adjustment 1. Connect the slit lamp delivery unit to the MC-300. 2. Check alignment of the spot size indications (50, 200, 500, and 990 µm) on the control box and markings on the slit lamp delivery unit.
11. Press the STATUS switch to put the MC-300 in STANDBY. 12. Set parameters for Address 69 to 01 and Address 178 to 00. 13. Minimize the aiming beam output, put the MC-300 in READY, and check that INTLK 2.8 does not occur.
10. Set parameter for Address 1 to 01 and set DIP SW 3 to the ON position. 11. Press the STATUS switch to put the MC-300 in STANDBY. 12. Emit the treatment beam and check that ERR 10 does not occur.
XMC20*RDA001B/E 7.26 Cooling Water Flow Amount Check 1. Set DIP SW 1, 2, 6, and 8 to the OFF position. 2. Set parameter for Address 177 to 100. 3. If INTLK 4.5 does not appear in normal mode, press the reset switch and measure the flow amount.
XMC20*RDA001B/E 7.29 Foot Switch Adjustment 1. Unscrew FC4 × 8 (n = 2) to remove the foot switch ASSY.(10149-EA01) from the cover (17627-M001). 2. Remove the screws (n = 2) and the foot switch cover. Screws(n=2) 3. Lower the micro switch positioning screws with a 0.05 inch hexagonal wrench so that the screw tips are hidden as shown in the right picture.
XMC20*RDA001B/E 8.3.2 VR detail table Item Details LD aiming gain Turn VR1 counterclockwise fully. adjustment (Not used) Laser 1 monitor offset Adjust VR2 so that the voltage between TP15 and A_GND is 0 V when adjustment the laser is off. For the laser head and external power meter: monitor GRN output Adjust VR3 so that the actual treatment beam output at the fiber optic...
XMC20*RDA001B/E 8.4 Message Tables 8.4.1 Interlock Display Details In. 1.6 Overheat of the laser power supply In. 2.7 Excessive output of the laser beam (The current limit is reached.) In. 2.8 Excessive output of the laser beam (The current limit is not reached.) In.
XMC20*RDA001B/E 8.5.3 Control box COLOR switch EMISSION indicator INTERVAL display Emergency OFF switch STATUS indicator COLOR indicator SPOT SIZE display STATUS switch POWER knob AIMING level INTERVAL switch indicator TIME display AIMING switch Reset switch TIME switch COUNTER display POWER display 8 - 35...
XMC20*RDA001B/E 8.7 Parameter Tables 8.7.1 Ver. 1.10 Displayed Initial Description ress characters value Control of mode DIP SW 1 and 2: ON (Interlock/error ON) 0: Normal mode, 1: IDLE countdown DIP SW 1 and 2: OFF (Interlock/error cancel) 0: No INTLK and Err display, 1: IDLE countdown/INTLK and Err display, 2: SUB-Interlock/error cancel onbEEP Laser emission sound control 0: No emission sound, 1: With emission sound...
Page 132
XMC20*RDA001B/E Displayed Initial Description ress characters value rP1Endo RED ch1 endophotocoagulation delivery unit POWER gain rP2Endo RED ch2 endophotocoagulation delivery unit POWER gain rP1 bio RED ch1 binocular indirect ophthalmoscope POWER gain rP2 bio RED ch2 binocular indirect ophthalmoscope POWER gain rP1HAAG RED ch1 HAAG attachable delivery unit POWER gain spare rP1 yccb...
Page 133
XMC20*RDA001B/E Displayed Initial Description ress characters value oFSG3-H GRN 310 mW or more 3 monitor overpower offset oFSy3-H YEL 150 mW or more 3 monitor overpower offset oFSr3-H RED 150 mW or more 3 monitor overpower offset LocHnG Power monitor circuit changeover output (mW) (300: GRN = 300, RED/YEL = 150) oFSG2-L GRN 300 mW or less 2...
Page 134
XMC20*RDA001B/E Displayed Initial Description ress characters value rFi 50 RED 50 mW correction rFi 100 RED 100 mW correction rFi 200 RED 200 mW correction rFi 300 RED 300 mW correction rFi 400 RED 400 mW correction rFi 500 RED 500 mW correction rFi 600 RED 600 mW correction rFi 700...
XMC20*RDA001B/E Displayed Initial Description ress characters value Not used ch1 cnt Fixed to one channel, (no sensor detection) 1: 1 ch fixed, others: normal Not used Not used Not used upd EEPr Setting the parameter value to 50 and pressing the reset switch allows the parameters whose address is 200 or later to be input.
Error/Intlkc OFF for the output INTLK2.7/2.8, Err 10/15 system Not wait until the cooling It is possible to start the MC-300 within three minutes device stops idling after turning the master switch on. Parameter mode Press the INTERVAL switch to change the parameter setting.