Nidek Medical ICE-1000 Operator's Manual
Nidek Medical ICE-1000 Operator's Manual

Nidek Medical ICE-1000 Operator's Manual

Intelligent blocker

Advertisement

INTELLIGENT BLOCKER
ICE-1000
Model
OPERATOR'S MANUAL

Advertisement

Table of Contents
loading

Summary of Contents for Nidek Medical ICE-1000

  • Page 1 INTELLIGENT BLOCKER ICE-1000 Model OPERATOR’S MANUAL...
  • Page 2 NIDEK CO., LTD. : 34-14, Maehama, Hiroishi-cho, Gamagori, Aichi 443-0038, Japan (Manufacturer) Telephone: (81-533) 67-6611 Facsimile: (81-533) 67-6610 NIDEK CO., LTD : 3F Sumitomo Fudosan Hongo Bldg., 3-22-5, Hongo, (Tokyo Office) Bunkyo-Ku, Tokyo 113-0033, Japan Telephone: (81-3) 5844-2641 Facsimile: (81-3) 5844-2642 NIDEK INCORPORATED : 47651 Westinghouse Drive, Fremont, California 94539, U.
  • Page 3 BEFORE USE, READ THIS MANUAL. This operator's manual includes operating procedures, safety precautions, and specifications for the ICE-1000. IEC standards are applied in this manual. For correct use, this manual is needed. Especially, the safety precautions and operating procedures must be thoroughly understood before using the instrument.
  • Page 4 Use precautions Before Use CAUTION • Do not use the instrument for other than the intended purpose. NIDEK is not responsible for accidents or malfunctions caused by careless use. • Never disassemble nor touch the inside of the instrument. This may result in electric shock or malfunction. •...
  • Page 5 CAUTION • Use the special packing materials when transporting the instrument to protect from impact of dropping. Excessive vibration or impact to the instrument may cause malfunction. During Use • Connect the cable to the interface connector securely, maintaining the correct CAUTION orientation of the cable connector.
  • Page 6 After Use • If the instrument will not be used, turn the power off and close the tracing unit CAUTION cover. Settled dust may affect the measurement accuracy. • Do not yank the power cord to disconnect it from a wall outlet but hold the plug. This can damage the metal core of the cord and may result in electric shock, short cir- cuit, or fire.
  • Page 7 [Operating area and maintenance work area] Maintenance work area About 500 mm About 500 mm Operating area About 1000 mm Disposal • Follow local governing ordinances and recycling plans regarding disposal or CAUTION recycling of the components. It is recommended to commission the disposal to a designated industrial waste dis- posal contractor.
  • Page 9: Table Of Contents

    ..........27 2.3.1 Operating the ICE-1000 in the Mini LAB system ......27 2.4 Tracing .
  • Page 10 Setting the ICE-1000 ........
  • Page 11: Before Use

    BEFORE USE Outline of the Instrument NIDEK Intelligent blocker, ICE-1000 blocks the lens with a lens cup. The lens is blocked with a lens cup properly at a detected position (the optical center, cylinder axis angle, segment position of bifocal lenses, and marking).
  • Page 12: Configuration

    BEFORE USE: Configuration Configuration Front view Tracer LCD touch screen Lens clamp button Block button Blocking arm Screen switch button Lens table Cup holder Lens clamp Barcode scanner USB port Pattern setting unit stand...
  • Page 13 BEFORE USE: Configuration LCD touch screen This 800 × 600 pixel color LCD touch screen displays various items that have been entered. Lens clamp button Lowers the lens clamp to hold the lens. Press this button to lower the lens clamp and hold the lens temporary. Align the lens and press the button again to fully tighten the lens for blocking.
  • Page 14 BEFORE USE: Configuration Tracing unit Rim clips Upper slider Pattern tracing pin Stylus Lower slider Pattern setting unit sup- port Right-eye tracing button Both-eye tracing button Left-eye tracing button Upper slider Lower slider Frames are fastened between these sliders. Rim clips Hold the rims of frames.
  • Page 15 . Use the LAN port when connecting three or more LEX-1000s in the Mini LAB system configuration. Ethernet: Standards on wirings, and access to the wirings to allow mutual accessing of various information processing terminals to the LAN. The ICE-1000 transfers and receives data through the 10BASE-T.
  • Page 16 BEFORE USE: Configuration EDGER 2 port Used to connect the second lens edger. The system configuration for the EDGER 2 port can be set by the corresponding parameter. BARCODE port Used to connect an optional barcode scanner.
  • Page 17: Display

    BEFORE USE: Display Display 1.3.1 Layout screen This is the screen to enter the each data of lens layout and processing conditions. Screen change tab Information bar Menu button Right/Left indication Frame pupillary distance (FPD) Pupillary distance Height of optical center Cylinder axis angle (AXIS) Size Display button...
  • Page 18 BEFORE USE: Display Frame pupillary distance (FPD) This shows the distance between the right and left frame centers. The frame center is calculated by the boxing system. Range: [30.00 to 99.50 mm (in increments of 0.01 mm)] The width between the nasal ends of the left and right frames (DBL) can be entered for the FPD value.
  • Page 19 BEFORE USE: Display Load button Loads the JOB data (or PTN data) saved in the USB flash drive to display it on the LCD touch screen. Tracer button Displays data measured by the tracer on the LCD touch screen. Optical center mark Frame center mark Circumference Layout mode...
  • Page 20 BEFORE USE: Display Processing mode button Selects the processing conditions. Guide/Auto Bevel guided processing mode ON/OFF Metal, Plastic Hole/Flat Drilling ON/OFF TwoPoint Groove/Flat Grooving ON/OFF Nylor Polishing button The selectable options depend on the connected lens edger. Set them with the Polish/S.B Setting parameter. When the Polish/S.B Setting parameter is set to LEX: The options are with polishing (Polish: polishing is on) and without polishing (None: polishing is off).
  • Page 21: Blocking Screen

    BEFORE USE: Display 1.3.2 Blocking screen This is a screen to block lenses. JOB#/PTN# Layout data Layout mode Lens type Lens detection mode CYL mode Image brightness Lens measurement value Magnification JOB#/PTN# Displays JOB# specified in the Layout screen. Layout data Displays layout data specified in the Layout screen.
  • Page 22 BEFORE USE: Display Magnification Displays magnification of the displayed image and pressing here switches the magnification between ×1 (actual size) and ×2. Right/Left indication Cup holder Cup mark Optical center mark Lens clamp pin area Lens clamp indication Narrow/Normal Frame change Right/Left indication This shows the left-eye or right-eye lens (Right/Left).
  • Page 23 BEFORE USE: Display Lens clamp pin area Displays the area where not touched to the lens clamp pin. Lens clamp indication Indicates the lens clamp condition. Open The lens is not clamped. Temp. The lens is temporary tightened. Clamp The lens is clamped. Frame change Pressed when blocking the lens to be changed the frame.
  • Page 24: Hole Edit Screen

    BEFORE USE: Display 1.3.3 Hole Edit screen This is a screen to edit hole position, detect demo lens and hole position. Hole position coordinate indication Mirror button L button Hole angle button R button Add hole buttons Shape outline/ hole position button Hole...
  • Page 25 BEFORE USE: Display Mirror button Turns the mirror function on or off. Turn the mirror function on to drill the holes symmetrically in the right and left lenses. L button Specifies the left lens. Add hole buttons Specify the type of the hole to be added. Select the desired hole type.
  • Page 26 BEFORE USE: Display Diameter indication Displays the set hole diameter. Press the numeric button to display the numeric keypad and change the hole diameter. Hole type Numeric button Simple hole, slotted hole, pair hole, notched hole, Depth counterbored hole, rectangular hole T screw, Jewel hole1 to Jewel hole 3 Not displayed * Each hole diameter of Jewel hole 1 to Jewel hole 3 depends on the connected ME-1000 and LEX-1000 settings.
  • Page 27: Shape Editor Screen

    BEFORE USE: Display 1.3.4 Shape Editor screen This screen is to change the imported lens shape data into the desired shape. Shape dimension Step Fix area button Clear button Undo/redo buttons Shape dimension Displays the value of the shape horizontal diameter and vertical diameter. Pressing the coordinate displayed part displays the +/- input button (or numeric keypad) and specify the value to change the hole position.
  • Page 28: Labels

    BEFORE USE: Labels Labels Cautionary labels are provided on the ICE-1000. If labels are curling up or characters fading and becoming barely legible, contact NIDEK or your authorized distributor. Indicates that caution must be taken. Refer to the operator's manual before use.
  • Page 29: Checking Contents

    BEFORE USE: Checking Contents Checking Contents Unpack the contents from the shipping carton and check them. The following are included in the standard configuration. • Main body • Pattern setting unit • Standard frame • Standard pattern • Frame change holder •...
  • Page 30: Before First Use

    BEFORE USE: Before First Use Before First Use Place the blocker on a stable table and connect the power cord. Place the main body on a stable table. Open the cover of the tracing unit and remove the cushion, tape, and fixing fitting retaining the inside of the tracing unit.
  • Page 31 Lock the cover. 3) Remove the lens clamp set screw. Keep the removed screws in storage because they are used for repacking the ICE-1000. Set screw 4) Close the top cover and fix with the screw. Remove the blocking arm set screw from the hole of right side cover.
  • Page 32 Electric shock or fire may occur in the event of malfunction or power leakage. Turn the power switch ON ( The initial screen is displayed on the LCD touch screen and the ICE-1000 is initialized. Check that the layout screen is displayed after the ICE-1000 is initialized.
  • Page 33: Tracing Unit Calibration

    In addition, when a measured circumference involves an error, perform the tracing unit calibration. Perform calibration with the standard frame. 1) Turn on the ICE-1000 and input a desired JOB# on a layout display. 2) Set the supplied standard frame ( 45) to Inscribed numbers the tracing unit.
  • Page 34 BEFORE USE: Before First Use Perform calibration with the standard pattern. 1) Set the supplied standard pattern into the tracing unit. Pattern setting unit Fit the standard pattern with the surface with the inscribed “A” facing the Pattern setting unit. See Steps 1 to 2 of 2.4.2 Pattern tracing for the method of setting a pattern.
  • Page 35: Operating Procedures

    OPERATING PROCEDURES Operation Flow Turning ON the ICE-1000 Entering JOB# 2.3.1 Operating the ICE-1000 in the Mini LAB system (Page 27) Tracing 2.4.1 Frame tracing (Page 30) Calling up lens shapes 2.3.1 Operating the ICE-1000 in the Mini LAB system (Page 27) Setting processing conditions 2.5.1 Setting processing conditions (Page 37)
  • Page 36: Start-Up

    A branch outlet that is concurrently used with other instruments may cause overheating and fire. Confirm that the cables from the connected instruments are securely connected to the ICE-1000, and the power for each instrument is turned ON. See “4.2 Connection” (Page 96) for the connecting method.
  • Page 37: Operating Procedures

    2.3.1 Operating the ICE-1000 in the Mini LAB system In the Mini LAB system, the ICE-1000 manages data as a data server. The ICE-1000 and the LEX-1000 comprise the system. Turn the power switch ON ( ) to start the ICE-1000.
  • Page 38 • When blocking a left-eye lens only, press the LEFT button to switch the screen to the layout display for the left-eye lens. • When the ICE-1000 is connected to the ME-1000, LEX-1000, or SE-9090 Express and AHM-1000 system, set the grooving and drilling specifications as necessary.
  • Page 39 JOB area and quickly press the (DATA SET) button. 2) The ICE-1000 transfers the data corresponding to the JOB# to the LEX-1000. 3) The data is displayed on the LEX-1000. • Import JOB data for processing lenses after blocking the lenses with the lens cups.
  • Page 40: Tracing

    OPERATING PROCEDURES: Tracing Tracing After entering a JOB# on a layout display, perform tracing. The following tracing methods are select- able. Select the most appropriate method of tracing in each case. Frame tracing (Both eyes) Traces the both eyes of general frames. Frame tracing (Single eye) Traces the left- or right-eye of a general frame.
  • Page 41 OPERATING PROCEDURES: Tracing 4) Move the frames left or right to bring them to the approximate center of the tracing unit. Start both-eye tracing. Press the Both-eye tracing button Tracing is completed when the frames are released. Pull the lower slider forward to remove the frames. Call up the traced data.
  • Page 42 OPERATING PROCEDURES: Tracing Semiauto tracing This is the procedure to fit the stylus into the groove by hand before tracing for the frame whose stylus is not automatically set in the groove. This is for frames, especially plastic ones, whose grooves are not in the middle of the edges. Set frames into the tracing unit.
  • Page 43 OPERATING PROCEDURES: Tracing Goggle type frame tracing When the frames are sharply warped, the stylus may come off the groove. In such a case, fasten only the one frame between the rim clips and perform single-eye tracing. This tracing is available in both left and right frames. The procedure for tracing the right eye frame is described here.
  • Page 44: Pattern Tracing

    OPERATING PROCEDURES: Tracing 2.4.2 Pattern tracing This is the procedure to trace the patterns of two-point frames etc. Set a pattern onto the pattern setting unit. 1) Set a pattern while pressing the white button on the pattern setting unit. Push the pattern as far as it will go.
  • Page 45: Demo Lens Tracing

    Block the lens so that the markings on the demo lens are aligned to the groove orientation of the lens cup. When blocking the lens with the ICE-1000, set the lens type to Demo.
  • Page 46: Stopping Tracing

    OPERATING PROCEDURES: Tracing Fit the lens cup into the pattern setting unit. Fit the lens cup while pressing the white button of the pattern setting unit. Start tracing as same as Step 2 to Step 4 of “2.4.2 Pattern tracing” (Page 34). Press theLeft-eye tracing button for the right eye or the Right-eye tracing button for the...
  • Page 47: Setting Processing Conditions And Entering Layout Data

    OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Setting Processing Conditions and Entering Layout Data Perform tracing or import traced data before setting processing conditions and entering layout data. Lens layout data and Processing conditions 2.5.1 Setting processing conditions As processing conditions, set the lens material, lens type, frame type, whether or not to polish lenses, whether or not to safety bevel lenses, and soft processing mode.
  • Page 48 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Lens material Select the lens material. Pressing the lens material button displays the pop-up menu. Select the button of a desired material. • Plastic (Generic plastic lens) • HiIndex (Plastic lens with high refractive index) •...
  • Page 49 Grooving is OFF • Setting this mode applies guided beveling automatically during lens processing. This prevents the lens from being processed in auto mode. • For guided beveling in the ICE-1000, the software version of the lens edger must be compatible. Polishing Select whether the lens edge is polished.
  • Page 50 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Each press of the Polish button switches polishing ON and OFF. • None (Polishing is OFF) • Polish Safety beveling Select whether the lens edge is safety-beveled. The selectable options depend on the connected lens edger. When the Polish/S.B Setting parameter is set to LEX: Each press of the S.F.B.
  • Page 51: Entering Layout Data For Single Vision Lenses

    Soft processing mode is ON Soft button displayed in gray Soft processing mode is OFF • For soft processing in the ICE-1000, the software version of the lens edger must be compatible. • Setting this mode applies the soft processing automatically during lens processing.
  • Page 52 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data The range of the FPD value is 30.00 to 99.50 mm in increments of 0.01 mm. Pressing the CE button clears the entered value. When no value is entered or the value is not changed, this button closes the numeric keypad.
  • Page 53 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Pressing the PD button displays the pop-up menu. Press the 1/2PD button to enter the monocular PD value The range of the monocular PD value is 15.00 to 49.50 mm in increments of 0.01 mm. 3) Optical center height Enter the optical center height for the left- and right-eye lenses respectively.
  • Page 54 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data The lens shape is enlarged or reduced laterally while Enlarged shape size maintaining a similar form. At this time, the frame cen- ter is regarded as the reference position. For example to the right: Size = +5 mm: Original shape size Original shape size + 5 mm •...
  • Page 55 Sets criteria of the single vision lens detection. The setting is performed with the single vision detection mode button. In this mode, the ICE-1000 blocks the lens by detecting the optical center and cylinder axis angle of the lens. In this mode, the ICE-1000 blocks the...
  • Page 56 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data • When the Cup Mode parameter is set to Pliable, Mini, or Nano, the displayed cup mark is constant in size. In addition, the settings of the Full eye Minimum Size, Grind Minimum Size (H), and Grind Minimum Size (Y) parameters become disabled See “4.1 Setting Parameters”...
  • Page 57: Entering Layout Data For Bifocal Lenses

    2.5.3 Entering layout data for bifocal lenses The ICE-1000 automatically blocks bifocal lenses by detecting the segment base point (top line center of segment). According to prescription, enter lens layout data (FPD, near PD, and height of the optical center).
  • Page 58 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Select either of the following. Enter the distance from the center point on the top line of the segment straight down to the point on the lens shape. Enter the height from the point level with the center of the segment on the top line straight down to the lowest point on the...
  • Page 59: Entering Layout Data For Progressive Power Lenses

    For the alignment method of the eye point for far vision, select either of the following three methods. The ICE-1000 automatically detects the printed eye point mark for far vision and hori- zontal line and blocks the lens in accordance with the detected marks.
  • Page 60 The range of the EP value is 0.0 to + 6.0mm in Laser mark Laser mark increments of 0.1mm. The ICE-1000 automatically detects the marked laser marks of the lens and blocks the lens. Entering lens layout data Set the lens type to Progressive (progressive power lens) Enter the FPD (or DBL).
  • Page 61 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data Select either of the following. Enter the distance between the eye point for far vision and straight down point of the lens shape. Enter the distance between the eye point for far vision and the lowest point of the lens shape.
  • Page 62: Entering Layout Data For Manual-Type Lenses

    OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data 2.5.5 Entering layout data for Manual-type lenses When the lens type is set to Manual (EX lenses or prism lenses etc.), mark the lens using a lensmeter in accordance with prescription. Then manually align the lens using the marking on the lens and block the lens.
  • Page 63 OPERATING PROCEDURES: Setting Processing Conditions and Entering Layout Data O Setting alignment scale The height of the horizontally long ellipse and the spacing between the horizontal lines of the alignment scale can be changed by the corresponding parameters. Manual Mode scale: (A in the figure) Select the spacing between horizontal lines: 0 mm, 1.0 mm, 2.0 mm, or 3.0 mm.
  • Page 64: Shape Change Function

    OPERATING PROCEDURES: Shape Change Function Shape Change Function This function changes the loaded lens shape data into the desired shape. There are two entering methods for changes. • Changes in shape by changing the horizontal and vertical numeric values. • Changes in shape by dragging the lens shape with the touch pen. There is also a method to determine the area that is not to be changed and other area that can be changed.
  • Page 65 OPERATING PROCEDURES: Shape Change Function As necessary, specify the area that is not to be changed in shape 1) Press the Fix area button to switch to fixed area specification mode. Numeric values disappear and a section of the lens shape is displayed in yellow. At the ends of the yellow line, handles (yellow circles) are displayed.
  • Page 66 OPERATING PROCEDURES: Shape Change Function Shape change with numeric values 1) Select a step to be changed from the pop-up menu displayed by pressing the Step but- ton. Select the desired step button from Tenkey, 0.10, 0.25, or 0.50. 2) Press a numeric value to be changed. The +/- buttons or tenkey are displayed.
  • Page 67 OPERATING PROCEDURES: Shape Change Function 3) Check the post-change shape on the screen. • The cross in solid lines represents the frame center of the original shape data. Once the shape is changed, it is updated but the frame center (cross) is not be updated. The frame center is updated after the changed shape data is saved.
  • Page 68: Grooving Settings

    OPERATING PROCEDURES: Grooving Settings Grooving Settings Grooving specifications can be set. The grooving data is transferred to the lens edger together with the normal processing data; it is unnecessary to set the grooving specifications in the lens edger. • The setting is possible only when the ICE mini is connected to equipment having the grooving function in the Mini LAB system (ME-1000 or SE-9090 Express/AHM-1000 system).
  • Page 69 OPERATING PROCEDURES: Grooving Settings • Curve 1) Press the Curve button. The pop-up menu is displayed. 2) Select the desired curve from the menu. Auto Computer-calculated curve Guide Sets guided grooving to be automatic when processing. Manually enter the desired curve. Curve Pressing the numeric button displays the numeric keypad.
  • Page 70 In the ICE-1000, the groove specifications are set without measuring the lens curve. Therefore, the set specifications may the groove to be off the lens edge.
  • Page 71: Drilling Settings

    OPERATING PROCEDURES: Drilling Settings Drilling Settings It is possible to enter the specifications related to the drilling from the ICE-1000. Lens outline and hole position can be read from the demo lens. The drilling data is transferred to the lens edger together with the normal processing data; it is unnec- essary to set the grooving specifications in the lens edger.
  • Page 72: Measuring Demo Lenses

    OPERATING PROCEDURES: Drilling Settings 2.8.2 Measuring demo lenses Press the Hole tab to display Hole Edit screen. The Hole tab is displayed only when the mode is passive. Hole/Shape button Press the Hole /Shape button to display the Lens shape screen. Clean the lens table and the demo lens edge with a blower brush.
  • Page 73 OPERATING PROCEDURES: Drilling Settings Specifying R or L starts measurement. It takes about 20 seconds for measurement. “Detecting...” is displayed during measurement. Wait for a while without touching the instrument. When the measurement is complete, the detected outline and frame center (cross) are displayed in white on the screen.
  • Page 74 OPERATING PROCEDURES: Drilling Settings Specify the precise hole position and hole diameter. Drag the hole indication to the hole on the demo lens image. For pair or notched holes, align also the distance between the holes by dragging. Match the hole position and diameter displayed in red with the image. 1) Drag (or click) the red circle with the touch pen to align it with the hole position of the demo lens...
  • Page 75 OPERATING PROCEDURES: Drilling Settings Hole type selection and position specification Select the hole type among the ten add hole buttons. Add hole button Setting and positioning of hole Setting of hole size Diameter Hole diameter Normal circular hole. Simple hole Specify the hole center position.
  • Page 76 OPERATING PROCEDURES: Drilling Settings Slotted hole Start point coordinates Start point coordinates Slotted hole width Slotted hole width (Hole diameter) (Hole diameter) End point coordinates End point coordinates At the setting of Edge At the setting of Center • The coordinate specification method of the start and end points is selectable between Center and Edge with the Slot coordinate mode parameter.
  • Page 77: Setting Input In The Hole Edit Screen

    OPERATING PROCEDURES: Drilling Settings 2.8.3 Setting input in the Hole Edit screen For the added hole, various settings other than position can be added. In the Hole Edit screen, hole addition and deletion can be performed to the displayed lens shape. Hole addition After selecting the hole type with the Add hole button, press the hole position with the touch...
  • Page 78 OPERATING PROCEDURES: Drilling Settings Hole angle Set the angle with the hole angle button Each press of the button toggles the setting in the following order. • Auto • Angle • x-y • X Auto Only Auto or Angle is selectable when the pair or notched holes are selected.
  • Page 79 OPERATING PROCEDURES: Drilling Settings Group Specify the group number to drill multiple parallel holes. When hole 1 and hole 3 are Hole 1 Press the Group numeric button to display the set to the same group, those numeric keypad. Enter the group No. and press holes are drilled in the lens the Ent button.
  • Page 80 OPERATING PROCEDURES: Drilling Settings Display method of hole position coordinates There are six types of coordinate displays available by selecting the horizontal and vertical reference positions. Select the horizontal reference position from among Center , B-Edge , and H-Edge reference position automatically selected between the temporal side (outside)
  • Page 81 OPERATING PROCEDURES: Drilling Settings Coordinate display methods When the vertical reference position is set to Center Center , Center B-Edge , Center H-Edge, Center When the vertical reference position is set to B-Edge Center , B-Edge B-Edge , B-Edge H-Edge, B-Edge...
  • Page 82: Blocking

    Insert the lens cup as far it will go orienting the top mark to the back (ICE-1000) side. When the displayed cup mark is oblate ellipse, be sure to use the lens cup for half-eye lens.
  • Page 83 OPERATING PROCEDURES: Blocking Press the screen switch button to switch to the Blocking screen. Or press the Block tab to switch to the Block- ing screen. Place the lens on the lens table with convex side up. While watching the lens projected on the screen, position the center of the lens at an approximate center of the lens table.
  • Page 84 OPERATING PROCEDURES: Blocking • Place the progressive power lenses on the lens table, keeping the amount of tilting of the horizontal reference line (or markings on the laser marks) within ±10°. If the amount of tilting exceeds ±10°, the eye point mark for far vision changes to red and blocking becomes prohibited.
  • Page 85 OPERATING PROCEDURES: Blocking Align the projected, printed eye point for far vision of the lens with the eye point mark for far vision. Confirm that the amount of tilting of the hori- zontal reference line is within ±10° during the alignment.
  • Page 86 OPERATING PROCEDURES: Blocking • For the JOB data containing both-eye data, the shape data is automatically switched to the one for the remaining eye as soon as one lens has been blocked. The following will be performed simultaneously with blocking: •The processing conditions and the lens layout data that were entered will be saved with the shape data.
  • Page 87: Blocking With Frame Change Holder

    OPERATING PROCEDURES: Blocking 2.9.1 Blocking with frame change holder Use the provided holder when the lens can not be held with the three pins on the lens table because the lens diameter is small as for when lenses are being reused in different frames. 1) Press the F.C.
  • Page 88: Blocking Demo Lenses

    OPERATING PROCEDURES: Blocking 2.9.2 Blocking demo lenses When blocking demo lenses to trace them, mark the lenses with a lensmeter in advance. See “2.4.3 Demo lens tracing” (Page 35) for details. When blocking demo lenses, it is unnecessary to call up lens shapes, enter processing conditions and lens layout data.
  • Page 89 OPERATING PROCEDURES: Blocking Error messages when of the blocking error The cause of the blocking error is displayed with the error message. Error message Action The blocking position is out of the blockable range. In the modes other than Manual, move the lens position or switch the Blocking position is out of range.
  • Page 90: 2.10 Data Control

    OPERATING PROCEDURES: Data control 2.10 Data control The data saved in the USB flash drive is deleted and changed the file name in the Data screen. The Data screen is displayed by pressing the Data button in the Layout screen. Shape change Hole edit Data edit...
  • Page 91 OPERATING PROCEDURES: Data control Displays the layout screen. When copying the JOB data, display Layout screen and after changing JOB#, save the Data read data. Cautions in changing a folder name and data name When a folder name is changed, the manufacturer name is also changed. When changing the data name, if "Maker name"...
  • Page 92 OPERATING PROCEDURES: Data control Data Edit screen The displayed button function us same as the screen displayed with the Layout tab. Pressing the Exit button displays “Do you save the data? Yes/No” and pressing Yes (or No) button returns to the Data screen. •...
  • Page 93: 2.11 Check After Use

    1) Is the power turned OFF? 2) Is the cover of the tracing unit closed? 3) Is the ICE-1000 dirty or damaged? 4) Are all the accessories accounted for and free from damege? Accessories; Pattern setting unit, Standard frame, Standard pattern...
  • Page 94: 2.12 Initial Layout Screen Save Function

    OPERATING PROCEDURES: Initial Layout Screen Save Function 2.12 Initial Layout Screen Save Function This function sets the items on the layout screen displayed when the power is turned on. When the factory settings on the layout display differ from those ordinally used, it is possible to change and save the items to be displayed on the initial layout screen upon system activation.
  • Page 95: Maintenance

    MAINTENANCE Troubleshooting In the event that the ICE-1000 does not work correctly, then troubleshoot with the following table before contacting NIDEK or your authorized distributor. Symptom Action • Confirm that the power cord is properly connected. • Confirm that the voltage applied to the outlet is within the range specified.
  • Page 96: Replacing Fuses

    MAINTENANCE: Replacing Fuses Replacing Fuses If the ICE-1000 is not started even though the power switch is turned on, the fuses may be blown. Replace the fuses with spare ones. • Be sure to turn the power off and disconnect the power cord from the outlet before CAUTION replacing the fuses.
  • Page 97: Setting Time And Date

    MAINTENANCE: Setting Time and Date Setting Time and Date Display the Parameter 1 screen. Press the Menu button on the layout screen and press the Parameter button from the dis- played pop-up menu. Press the Clock button twice to display the Clock screen.
  • Page 98: Adjusting The Cup Holder

    Plunger Cleaning When the covers or panels of the ICE-1000 becomes dirty, wipe with a soft cloth. For severe dirt, immerse the cloth in a neutral detergent, wring well, and wipe. Finally wipe with a dry and soft cloth. • Never use an organic solvent such as paint thinner.
  • Page 99: Connection And Settings

    CONNECTION AND SETTINGS Setting Parameters The parameters can be changed according to the needs of the operator. Display the Parameter 1 screen. Press the MENU button on the layout screen and press the parameter button from the pop- up menu. Press the button of the desired parameter to select and change the setting.
  • Page 100 Blocker/VCA should be selected for special models. 2) ID No.: 0 to 9999 Factory setting: 32 This parameter sets the ID No. for multiple ICE-1000 units in a network. Set any number from 0 to 9999 as an identification number. 3) Server No.: 0 to 9999 Factory setting: 32 This parameter sets the ID No.
  • Page 101 CONNECTION AND SETTINGS: Setting Parameters 5) Tracer: Inside, Outside Factory setting: Inside This parameter sets the tracer from which shape data is imported. Ordinarily, never change the parameter setting from Inside. Outside is only for the ICE- 1000NT (without the tracing unit equipped). Setting the parameter to Outside disables the operator from loading shape data from the built-in tracing unit.
  • Page 102 CONNECTION AND SETTINGS: Setting Parameters 9) Full eye minimum size: 25.0 to 35.0 mm Factory setting: 30.0 mm This parameter sets the minimum diameter of the lenses (when the standard cup is used) that can be processed by a lens edger. Set any value from 25.0 to 35.0 mm (in increments of 0.1 mm) according to the lens edger.
  • Page 103 Factory setting: Necessary This parameter sets whether the height of the optical center value should be fixed or not when the ICE-1000 imports traced data on condition that the ICE-1000 displays no layout data on the screen. When the parameter is set to Necessary, it is necessary to modify the height of the optical center value after traced data is imported from a tracer (the corresponding button is dis- played in yellow).
  • Page 104 CONNECTION AND SETTINGS: Setting Parameters When the parameter is set to Necessary, the data is shown by a numeric button on a yellow background except when JOB data is imported from built-in memory. Unless the EP value on the yellow background is modified or confirmed, the intelligent blocker prohibits the oper- ator from blocking or saving the JOB data.
  • Page 105 CONNECTION AND SETTINGS: Setting Parameters This parameter allows the operator who are holding the lens with hands during blocking to release hands from the lens at the lowest position by stopping the holder arm with the lens blocked with the lens cup. In the case of the setting of 0.0 to 3.0 seconds: The holder arm returns to the original position after it has stopped for the set period of time.
  • Page 106: Connection

    • Confirm that all instruments are turned off before connecting them to the ICE-1000. CAUTION Mini LAB system In this small- or middle-scale system, the ICE-1000 serves as a data server. Tracing is performed with the ICE-1000. LEX Drill LEX-1000 ICE-1000 The optional barcode scanner is also available.
  • Page 107 CONNECTION AND SETTINGS: Connection Method of connection to the ICE-1000: ICE-1000 LEX-1000 LEX Drill RS-232C Cable RS-232C Cable EDGER 1 COM1 Drill COM1...
  • Page 108: Setting The Ice-1000

    *2: Never change the factory setting of Mac Address that had been set uniquely for each individual ICE-1000 before shipment. *3: Set these parameters in accordance with the network to which the ICE-1000 is connected. *4: When the NIDEK-produced PC server software is used, set the host port No. to “55555”. When the...
  • Page 109: Specifications And Accessories

    SPECIFICATIONS AND ACCESSORIES Safety Features To ensure safe use, the ICE-1000 is provided with the following features. <Blocking pressure detection function> Even if fingers are caught between the cup holder and the lens, a blocking pressure higher than the specified pressure is never applied against the fingers to protect them from injury.
  • Page 110: Specifications

    SPECIFICATIONS AND ACCESSORIES: Specifications Specifications Lens Measuring Unit • Method Imaging of reflected image (Lens outline imaging) Imaging of projected image (ALM function) • Maximum lens diameter 80 mm diameter • Camera CMOS color camera (Lens outline imaging) CMOS black and white camera (ALM function) •...
  • Page 111 Dimensions and Weight etc. • Dimensions 325 (W) ×510 (D) × 345 (H) mm • Weight Approx. 23 kg (main body of ICE-1000 only) • Power source AC100 to 120V ± 10%, 200 to 240V ± 10%, 50/60 Hz • Power consumption...
  • Page 112 SPECIFICATIONS AND ACCESSORIES: Specifications Environmental Conditions (in use) • Temperature +5 to +45ºC • Humidity +5 to +31ºC : 30 to 80% +31 to +40ºC : 30% at the minimum and 80% (+31ºC) to 50% (+40ºC) changes linearly at the maximum •...
  • Page 113: Standard Configuration

    SPECIFICATIONS AND ACCESSORIES: Standard Configuration Standard Configuration 5.3.1 Standard accessories • Lens table for frame changing 1 unit • Pattern setting unit 1 unit • Standard frame 1 unit • Standard pattern 1 unit • Frame change holder 1 unit •...
  • Page 114: Optional Accessories

    SPECIFICATIONS AND ACCESSORIES: Standard Configuration 5.3.2 Optional accessories • Barcode scanner • RS-232C cable...
  • Page 115: Index

    INDEX Numerics 1/2PD ........42 LAN port .
  • Page 116 INDEX:...

Table of Contents