6/ Repair Procedures
5. Remove the Tabletop from the Modulus SE. See Modulus SE Service Manual
section 5.3.
6. Disconnect the four color coded tubing connectors from the SIB Assembly
tubing.
7. Carefully cut and remove the cable ties that secure the electrical harness to the
pan.
8. Disconnect connector J1 from P1 on the SIB Assembly.
9. Open the Vanity panel on the rear of the Anesthesia Machine.
10. While supporting the Interface Assembly, remove the two M4 keps nuts (item
5, Figure 6-37) using a 7 mm nutdriver.
11. Carefully route the harness and tubing through the pan to remove the
assembly.
2
1
1.
Label, patient interface, 1001-3464-000
2.
Harness SA, SIB, 1503-3081-000
3.
Panel, sensor interface, Modulus SE, 1004-3933-000
4.
Spacer, sensor interface panel, 1004-3934-000 (2)
5.
Nut, keps M4, 0144-3717-314 (2)
Figure AA.85.083
Patient Interface/Harness Assembly (shown mounted on the Modulus SE left
side)
1503-0151-000
3
5/26/0
5
4
6-45