Technical Indicators
Measuring Range
Accuracy
Temperature Range
Response Time
Pressure Range
Shell Material
Output
Power Supply
Pipe Thread
Cable Length
Protection Grade
Instructions Before Use
The membrane head is fragile, it cannot be repaired in case of damage.
Before using the electrode, gently remove the protective cover. Put electrode in the
solution to test for polarization for more than 6 hours. For more details, refer Polarization
Section.
If no electrolyte is in the membrane, the measurement will be inaccurate or fluctuating.
After adding membrane head with electrolyte, do not store the electrode in air for more
than 30 mins.
Clean and calibrate the membrane head timely. Presence of dirt, adhesive or encrust on
the membrane head can cause inaccuracy in measurement.
Make sure no bubbles are present in membrane head for accurate results.
Extension, cutting, joint and forcefully pulling signal line may cause inaccurate results.
Electrode Wiring
Please follow the instructions carefully, wrong wiring will damage the product completely.
Examine all the wiring carefully in the system. Make sure wiring is complete before
switching ON the power.
It is strictly forbidden that RS485A line and RS485B line should not touch the power line.
It will permanently damage the communication of the electrode.
0.00
20.00 mg/L (ppm)
2% or ± 10 ppb HOCI
0
60.0 ˚C
90% less than 90 seconds
0
1 bar
ABS, 316
RS485, 4
20 mA
9
30 VDC (Recommend 12V)
M 39*1.5
5m or customized
IP68
2