Maintenance And Storage - l-com SRWQ100-YL105-9001 User Manual

Table of Contents

Advertisement

Sr.
Function
No.
Write device
13
01 06 00 19 00 02 D9 CC
address
Write baud
14
01 06 00 1A 00 00 A8 0D
rate
Write restore
15
01 06 00 1B 00 FF B9 8D
default
Write manual
16
temperature
Write zero
17
01 06 00 3E 00 FF A8 46
calibration
Write slope
18
calibration
Read
19
measuring AD

Maintenance And Storage

The electrode should be cleaned regularly.
Be careful while disassembling and rinsing the chlorine membrane,
Do not wipe chlorine membrane on the electrode with filter paper or sandpaper.
If the membrane head is fouled and clogged, the electrolyte is dry, lack or is
contaminated. Stop using the membrane and its head should be removed and cleaned.
After cleaning the electrode, replace the membrane head, add electrolyte. After
long-term storage, it needs to be polarized and calibrated before use.
The cable connector must be kept clean, dry, free from moisture, water, acids, alkalis or
salts, etc.
If water supply is cut off or electrode is not used on site, the electrode should be taken
out, cleaned and covered with protective sleeve containing water for preservation.
If the electrode is not used and stored for a long time, remove the electrode, disconnect
the cable, drain the electrolyte, thoroughly clean the anode and the cathode with
deionized water at 30 °C-40 °C, dry it and put on the protective cover. Place it in a dry
place for storage.
Send command
01 10 00 3A 00 02 04 00
00 41 A0 40 EC
01 06 00 3F 00 FF F9 86
01 03 00 66 00 01 64 15
Return command
01 06 00 19 00 02
D9 CC
01 06 00 1A 00 00
A8 0D
01 06 00 1B 00 FF
B9 8D
01 10 00 3A 00 02
61 C5
01 06 00 3E 00 FF
A8 46
01 06 00 3F 00 FF
F9 86
01 03 02 2E E0 A4
6C
11
Remarks
Set to 2
Set to 2400
Factory default values
are restored after sent
Set to 20.0
Confirm to calibration
zero
Confirm to calibration
slope
2EE0 turns to integer
12000