Page 3
01934 641569 E-mail safelab@safelab.co.uk Website: http://www.safelab.co.uk NOTE: This Airone DS Semi-Mobile Ducted Fume Cupboard requires an annual service and inspection under the requirements of COSHH regulation 9 by a suitably qualified engineer. DOCUMENT NO: M27 ISSUE NO: 07 ISSUE DATE: 27/06/18...
PRINCIPLES OF OPERATION Air is drawn into the cupboard at the working aperture below the sliding sash at a velocity high enough to ensure entrainment of any fumes/odours given off by the processes carried out within the cupboard. The contaminants are entrained in the airstream, directed past the baffles within the fume cupboard and drawn into the extract ductwork.
Page 5
INSTALLATION AND ASSEMBLY INSTRUCTIONS Contents of Packaging (as standard): - Operational manual - Quality Pass - Service and Maintenance Letter - Mains power lead - Conformity certificate - Safety Log Book - Warranty form Remove outer packaging. Connect the duct spigot on the top of the unit to the extract duct system with the flexible duct (supplied).
The work tray is white polypropylene as standard. Sash Operation: The vertical sliding laminated glass sash of the AIRONE DS fume cupboard is suspended and counterweighted providing added advantages over fixed-sash or folding sash fume-hoods. E.g. the sliding laminated glass sash can be closed and this provides protection to operators.
SPECIFICATION (CONTINUED) Model 1000DS 1200DS 1500DS Dimensions (mm) Width Depth Height Width Depth Height Width Depth Height External 1000 1850* 1200 1850* 1500 1850* Internal working area Max. Sash opening in 1040 1340 operation Max. Sash opening for 1040 1340 loading Weight (Packed) 120kgs...
Page 8
OPERATING INSTRUCTIONS Position the Airone DS semi-mobile ducted fume cupboard where it is to be used and lock the front castors (see page 9 for a guide to positioning the fume cupboard). Connect the flexible duct to the extract duct system and clip wire tether to anchor point on the docking before inserting the mains plug into the 13A electric socket.
Page 9
OPERATING INSTRUCTIONS (CONTINUED) GUIDE TO POSITIONING LOCATIONS FOR THE AIRONE RS MOBILE FUME CUPBOARD DOCUMENT NO: M27 ISSUE NO: 07 ISSUE DATE: 27/06/18...
Page 11
CALIBRATION (SHOULD ONLY BE UNDERTAKEN BY A SUITABLY QUALIFIED PERSON) Switch off cupboard at mains socket. Press and hold the Mute Button on the Low Airflow Alarm control panel whilst getting an assistant to switch the unit on at the mains socket. When an audible beep is heard, release the Mute Button.
Page 12
There should be regular formal inspections including earth bonding and insulation testing. Your Airone DS Semi-Mobile Ducted Fume Cupboard should have an annual service by a suitably qualified person, to maintain its good condition and reduce the possibility of hazard to the operator.
Page 13
MAINTAINANCE AND PERFORMANCE MONITORING (CONTINUED): INSTRUCTIONS FOR FACE VELOCITY MEASUREMENTS (Record results on record form - page 14 ) PROCEDURE: Imagine the face of the fume cupboard divided into nine cells. Stand as far as practicable from the fume cupboard with the sensing head in the plane of the sash and take airflow readings at approximately the centres of each of the nine cells.
Page 14
AIRONE DS EXAMINATION RECORD Site:..................Unit location: ................Air flow meter used: ..............Date Average Drop Any deterioration or Initials Readings in each of the cells as ( m/s ) > filters Fume damage observed. above. Check PASS flow Fail if face velocity...
CLEANING The materials used to construct the Airone DS Semi-Mobile Ducted Fume Cupboard have been selected to give maximum durability and a long life. It is beneficial however to regularly clean and decontaminate the internal and external surfaces. It is recommended that the cabinet is switched on during any cleaning procedure and that suitable protective clothing (face-mask, gloves and safety glasses) is worn.