Safelab Airone 1000RS/E Instruction Manual

Safelab Airone 1000RS/E Instruction Manual

Mobile filtration fume cupboard

Advertisement

Quick Links

INSTRUCTION MANUAL
Mobile Airone 1000RS/E
Filtration Fume Cupboard
SAFELAB SYSTEMS
Airone Building ● 8 Beaufighter Road ● Weston-Super-Mare ● BS24 8EE
Telephone: 01934 421340  Fax: 01934 641569  E-mail: Safelab@safelab.co.uk
WWW.SAFELAB.CO.UK
DOCUMENT NO: M37 ISSUE NO: 02 ISSUE DATE: 01/11/2015
1

Advertisement

Table of Contents
loading

Summary of Contents for Safelab Airone 1000RS/E

  • Page 1 INSTRUCTION MANUAL Mobile Airone 1000RS/E Filtration Fume Cupboard SAFELAB SYSTEMS Airone Building ● 8 Beaufighter Road ● Weston-Super-Mare ● BS24 8EE Telephone: 01934 421340  Fax: 01934 641569  E-mail: Safelab@safelab.co.uk WWW.SAFELAB.CO.UK DOCUMENT NO: M37 ISSUE NO: 02 ISSUE DATE: 01/11/2015...
  • Page 2 Contents Page 2 Contents. Page 3 Foreword. Page 4 Principles of Operation. Page 5 Installation and Assembly Instructions. Page 6 & 7 Specification. Page 8 Chemical List. Page 9 &10 Operating Instructions. Page 11 Wiring Diagram. Page 12 Calibration. Page 13,14,15 & 16 Maintenance and Performance Monitoring.
  • Page 3 01934 641569 E-mail safelab@safelab.co.uk Website: http://www.safelab.co.uk NOTE: This Mobile Airone 1000RS/E Filtration Fume Cupboard requires annual service and inspection under the requirements of COSHH regulation 9 by a suitably qualified engineer. DOCUMENT NO: M37 ISSUE NO: 02 ISSUE DATE: 01/11/2015...
  • Page 4: Principles Of Operation

    PRINCIPLES OF OPERATION Air is drawn into the cupboard at the working aperture A below the sliding sash at a velocity high enough to ensure entrainment of any fumes/odours given off by the processes carried out within the cupboard. The fumes from within the cupboard are carried by the moving body of air through the cupboard B to the pre-filter C and main filter D to absorb vapours and other noxious fumes/odours before being expelled by the exhaust fan E as clean air F.
  • Page 5 INSTALLATION AND ASSEMBLY INSTRUCTIONS Contents of Packaging (as standard): - Filtrete pre-filter. - Operational manual - Main filter. - Quality Pass. - Allen key - Service and Maintenance Letter. - Mains power lead - Conformity certificates - Safety Log Book. - Warranty form.
  • Page 6: Specification

    The work tray is white polypropylene as standard. Sash Operation: The vertical sliding laminated glass sash of the AIRONE 1000RS/E fume cupboard is suspended and counterweighted providing added advantages over fixed-sash or folding sash fume-hoods. E.g. the sliding laminated glass sash can be closed and this provides protection to operators.
  • Page 7: Specification Diagram

    SPECIFICATION (CONTINUED) Pre-filter: One piece cardboard framed blended synthetic fibre with polypropylene spun- bonded scrim – Safelab code: 101297 pack of 12 (the pre-filter will require replacing a minimum of every three months). Main filter. CMS schools impregnated/activated carbon filter - Safelab code: 101094, Weight: 22 kg subject to humidity.
  • Page 8 CHEMICALS WHICH ARE SUITABLE FOR USE WITH THE SAFELAB CMS (EDUCATION) CARBON FILTER: Inorganic Aluminium Chloride and Bromide Ammonia Ammonia Chloride Bromine Chlorine Chromium (VI) Dichloride Dioxide (Chromyl Chloride) Hydrochloric Acid vapour Hydrogen Chloride Hydrogen Sulphide Iodine Iodine Chlorides Lead fumes...
  • Page 9 OPERATING INSTRUCTIONS Position the Airone RS/E mobile fume cupboard where it is to be used and lock the front castors (see page 10 for a guide to positioning the fume cupboard). Clip wire tether to anchor point before inserting mains plug into a13A electric socket Connect flexible gas, water and waste pipes to the plumbed in fittings or docking station ensuring that the gas and water fittings are securely pushed home and that the waste coupling is screwed on tightly.
  • Page 10 Services on LH side Services on RH side OPERATING INSTRUCTIONS (CONTINUED) GUIDE TO POSITIONING LOCATIONS FOR THE AIRONE RS MOBILE FUME CUPBOARD DOCUMENT NO: M37 ISSUE NO: 02 ISSUE DATE: 01/11/2015...
  • Page 11: Wiring Diagram

    WIRING DIAGRAM DOCUMENT NO: M37 ISSUE NO: 02 ISSUE DATE: 01/11/2015...
  • Page 12 CALIBRATION (SHOULD ONLY BE UNDERTAKEN BY A SUITABLY QUALIFIED PERSON) Digital Control Panel - Switch on the unit using the green rocker switch and allow it to run for a minimum of fifteen minutes. Switch off the unit and switch it back on again straight away while holding in the “CAL” button on the digital control panel membrane.
  • Page 13 Filtration Fume Cupboards. Please phone our Service Department for details. Tel. No. 01934 421342 Safelab Systems recommend that a simple air sampling test be carried out once a week at eye and mouth level and/or at the outflow port of the AIRONE Fume Cupboard utilising an appropriate GASTEC Chemical Detector Tube (available from Safelab Systems).
  • Page 14 MAINTAINANCE AND PERFORMANCE MONITORING (CONTINUED): INSTRUCTIONS FOR FACE VELOCITY MEASUREMENTS (Record results on record form - page 15 ) PROCEDURE: Imagine the face of the fume cupboard divided into nine cells. Stand as far as practicable from the fume cupboard with the sensing head in the plane of the sash and take airflow readings at approximately the centres of each of the nine cells.
  • Page 15 MAINTAINANCE AND PERFORMANCE MONITORING (CONTINUED): AIRONE 1000RS EXAMINATION RECORD: Site:............................Location of Airone 1000RS: ....................…………………………………………… Air flow meter: ..........used Date Average Drop Any deterioration or Initials Readings in each of the cells as ( m/s ) > filters Fume damage observed.
  • Page 16 Operational Safety Log-Book. CLEANING The materials used to construct the Airone 1000RS/E Mobile Fume Cupboard have been selected to give maximum durability and a long life. It is beneficial however to regularly clean and decontaminate the internal and external surfaces.
  • Page 17: Pre-Filter Replacement

    Remove the pre-filter from the top of the main filter, and unscrew the filter clamping knobs on each side of the filter. Check that the new or replacement Safelab carbon filter is correctly selected (consult Safelab Ltd by phone, fax or e-mail for assistance).
  • Page 18: Spare Parts

    WARNING: If a spill exceeds the capabilities of the Airone 1000RS/E, it is likely to exceed the capabilities of carbon filtration face masks. This means that only an air or oxygen cylinder based breathing apparatus could provide adequate protection.