UNI-T UDP3305S Series User Manual page 20

Programmable dc power supply
Hide thumbs Also See for UDP3305S Series:
Table of Contents

Advertisement

The interface of list mode is as follows:
To set the list mode parameters manually:
Press "Basic parameter" key to enter the setting interface. Users can edit the starting group number, the required
number of output groups (maximum 2048 groups), the number of cycles (maximum 99999 times) and the output
status after the timing time expires. Set parameters for each group: Set voltage, current, and time parameters for
each group based on the required number of output groups.
To set the list mode parameters using template:
Press "Template" to open the template editing menu.
1. Press "Type" to select the desired template;
2. Press "Edit object" to select "Voltage" or "Current";
3. Define waveform start group values and total points;
4. Press "More" to set the waveform parameters such as the voltage, maximum and minimum current, high and
low levels, pulse width, period and inversion;
5. Press "Construct" to complete the waveform output setting;
6. Press
several times until the instrument returns to the initial interface of the list mode. Press the "Enable"
key, the timing output function is set up.
Note: Press
to return to the previous menu.
UDP3305S Series Programmable DC Power Supply User Manual