Download Print this page

Agilent Technologies 8753ES Datasheet page 26

Network analyzers, 30 khz to 3 or 6 ghz
Hide thumbs Also See for 8753ES:

Advertisement

Accessories
Verification Kits
Measuring known devices other than the standards
used in calibration is an easy way to verify the
proper operation of an 8753ES measurement sys-
tem. Agilent offers verification kits which include
devices, with data, for verifying the error-corrected
measurements of an 8753ES and 50-ohm test sets.
85029B 7-mm verification kit
Contains a set of precision 7-mm devices, with
data traceable to NIST* for verifying the calibrated
performance of an 8753E measurement system.
The 85031B 7-mm calibration kit is required for
complete verification.
Test-Port Return Cables
Agilent offers high-quality RF cables to connect
the 8753ET/ES and test sets to devices under test.
These cables offer excellent RF shielding for high-
dynamic-range measurements.
11851B 50-ohm type-N RF cable kit
Recommended for use with the 11850C/D three-way
power splitters. Kit includes three phase-matched
610-mm (24-in.) cables and one 860-mm (34-in.) cable.
Return loss
Phase tracking
8120-5639 50-ohm type-N RF cable
A 50-ohm type-N test-port cable
(both connectors are type-N male).
Return loss
11857B 75-ohm type-N test-port return cables
A pair of 610-mm (24-in.) test-port return cables
for use with the 8753ES or 85046B 75-ohm
S-parameter test set.
Return loss
Phase tracking
11857D 7-mm test-port return cables
A pair of 610-mm (24-in.) test-port return cables
for use with the 8753ES with the 85046A and
85047A S-parameter test sets. These cables can
be used with connector types other than 7-mm
with the appropriate precision adapters.
Return loss
Phase tracking
11857F type-F test-port return cables
Type-F test-port cable set (75-ohm). Includes one
cable with type-N(male) to type-F(male) connectors
and one cable with type-N(male) to type-F(female)
connectors.
26
>24 dB to 3 GHz
±4° at 1.3 GHz
>30dB to 6 GHz
>24 dB to 2 GHz
±2° at 1.3 GHz
>24 dB to 3 GHz
>20 dB to 6 GHz
±2° at 1.3 GHz
11850C/D three-way power splitters
11850C
Impedance
50 ohms
Frequency range
DC to 3 GHz
Tracking
±25 dB, ±3°
Equivalent source
30 dB at 1.3 GHz
match
(ratio or leveling) 20 dB at 3 GHz
Nominal
9.5 dB + 1 dB/GHz 7.8 dB
insertion loss
Input port match
DC to 1.3 GHz
20 dB
1.3 GHz to F max 10 dB
Maximum operating +20 dB
level
Damage level
+30 dB
RF connectors
RF input
50-ohm type-N (f) 50-ohm type-N(f)
All others
50-ohm type-N (f) 75-ohm type-N(f)
Includes
Three 11852B 50- to 75-ohm
minimum-loss pads
Recommended
11851B RF cable kit
accessories
11667A 50-ohm power splitter
Frequency range
Typical insertion loss
Equivalent source match
DC to 4 GHz
4 GHz to 8 GHz
8 GHz to 18 GHz
Tracking (between output arms)
DC to 4 GHz
4 GHz to 8 GHz
8 GHz to 18 GHz
Maximum operating level
Connectors
Option 001
RF input
Outputs
Option 002
RF input
Outputs
Dimensions
Recommended accessories
11852B 50-ohm to 75-ohm minimum loss pad
Frequency range
Nominal insertion loss
Return loss
300 kHz to 2 GHz
2 GHz to 3 GHz
Maximum input power
11850D
75 ohms
DC to 2 GHz
±2 dB, ±2.5°
30 dB at 1.3 GHz
20 dB at 2 GHz
20 dB
10 dB
+20 dB
+30 dB
DC to 18 GHz
6 dB
26 dB
21 dB
17 dB
±15 dB
±2 dB
±25 dB
±27 dBm
50-ohm type-N (f)
type-N (m)
type-N (f)
type-N (f) on RF
precision 7-mm
46 mmH x 52 mm W x19 mm D
(1.8 x 2.0 x 0.7 in)
11851B RF cable kit
DC to 3.0 GHz
5.7 dB
32 dB
27 dB
250 mW (+24 dBm)

Hide quick links:

Advertisement

loading

This manual is also suitable for:

8753et