Power Transformer And Power Distribution; Power Transformer And Power Distribution Cabling - Toshiba Aquilion TSX-101A/I Series Site Planning Manual

Table of Contents

Advertisement

8.1.2

Power transformer and power distribution

8.1.2.1

Power transformer and power distribution cabling

Works for the power transformer and power distribution cabling must be performed to meet
the electrical specification shown in subsection 8.1.1.
Design and installation should be made considering the maximum input currents listed below
if a power distribution cable is very long.
See figures 8.1-1 and 8.1-2 "Power supply system".
X-ray system current (gantry rotation
section)
Console power supply
Power for the gantry stationary section
* Power is calculated assuming a power factor of 1.0.
*1: When high-power X-ray generator kit (option) is incorporated
CAUTION:
Note: Install the special transformer for this system in the hospital power supply.
Output systems
An independent power supply must be provided for this system at the cubicle
output stage. Otherwise, this system may adversely affect other facilities or
equipment.
Maximum input current
Approximately 240 A or 275 A
and at 200 V
Approximately 30 A in single phase and at 200 V
Approximately 30 A in three-phase and at 200 V
(average)
34
No. 2A201-519EN*A
*1
in three-phase