(8)
External dimensions and mass
Unit name
Gantry (with cover)
Gantry (without cover but with
median-line metal plate)
Patient couch (CBTB-019A)
Patient couch (CBTB-019B)
Console (CKCN-012B/2A)
Computer system (CPU BOX)
(REC BOX)
Speaker
System transformer
*1
: The dimensions of the console are those when a recommended table is used.
(9)
Maximum cable length
See subsection 8.6 "Cable Connections Between the Units".
(10)
Installation altitude
The installation altitude should be no more than 1000 m above sea level. Prior consultation is
required for installations higher than 1000 m.
(11)
Transport and storage conditions
Observe the following conditions during transport and storage after shipment.
Ambient temperature : -10°C to 50°C
Relative humidity
Atmospheric pressure : 700 hPa to 1060 hPa
Vibration
Width (mm)
2,330
2,290
630
630
*1
2,470
450
870
140
800
: 40% to 80% (No condensation)
2
: 9.8 m/s
(1G) or less (during storage)
2
: 19.6 m/s
(2G) or less (during transport)
External dimensions
Depth (mm)
960
875
2,690
2,390
815
815
650
135
770
11
No. 2A201-519EN*A
Mass
(kg)
Height (mm)
1,950
1,750
1,935
1,650
450
480
450
450
1,470
435
700
140
1,470
260
200
2
980
550