MinXray HF100H High Frequency Portable Medical X-ray unit Installation and Operating Instructions Version 2.0 CAUTION; Federal law restricts this device to sale and use by or on the order of a licensed physician. MinXray, Inc. 3611 Commercial Avenue Northbrook, Illinois 60062-1822 U.S.A.
It consists of an x-ray tubehead/control attached to which is a continuously adjustable light beam collimator. A trunnion enables the HF100H to be attached to a suitable stand, such as the MinXray XGS series of portable mobile stands. The HF100H is supplied with a detachable AC power cord, and a detachable exposure cord with a 2-stage exposure switch.
MinXray, Inc., its agents and representatives, do not accept any responsibility for: Injury or danger to patient or other personnel from x-ray exposure. Overexposure due to poor operating techniques or procedures.
HF100H MAIN PARTS OF MinXray Knob to adjust x-ray field size (horizontal) Main power switch Tape measure Collimator light on button Knob to adjust x-ray field size (vertical) Fuse box Connector for exposure switch Connector for power cord Exposure switch and cord...
Warning Statements, Figures and Symbols WARNING THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT & OPERATOR UNLESS SAFE EXPOSURE FACTORS & OPERATING INSTRUCTIONS ARE OBSERVED DANGER : HIGH VOLTAGE. NO USER SERVICEABLE COMPONENTS INSIDE. REFER SERVICE TO QUALIFIED SERVICE TECHNICIAN. FOR CONTINUED PROTECTION AGAINST HAZARDOUS CONDITION, REPLACE FUSE WITH SAME TYPE AND RATING ONLY.
HF100H is properly attached to the model XGS stand, it can remain attached for use and transport. 2. Connect the AC power cord to the socket on the back of the HF100H. Note the rated voltage of the HF100H before connecting it to an AC power source.
1-3 Rejection limit: ±10%+1 msec. If the timer needs adjustment: (1) remove the 4 screws on the side of the top cover of the HF100H, and lift the top off of the body of the unit; (2) adjust the potentiometer "VR6" on the printed circuit board M3108AH on the inside of the top cover (see Fig.1) so that radiation exposure time equals the...
Page 11
If the peak tube potential needs adjustment: (1) remove the 4 screws on the side of the top cover of the HF100H, and lift the top off of the body of the unit; (2) adjust the potentiometer "VR4" on the printed circuit board M9112 located on top of the lower portion of the unit (see Fig.
Page 12
2.5 V 3. Open the top cover of the HF100H by removing the 4 screws and lifting the cover off of the unit. Connect ch 1 of oscilloscope to Ep pin of PC board M9112 and connect ch 2 of oscilloscope to Ip pin of PC board M9112.
INSTALLATION CHECK LIST INSTALLATION CHECK HF100H LIST Fill out this checklist SERIAL NO. when installation is DATE complete. MANUFACTURED EQUIPMENT LOCATION TEST DESCRIPTION ACCEPTANCE RESULTS CHECK LIMITS Line Voltage 105~135 V or ……….V 200~260 V NO GO ≤ 4.8% Line Voltage ..……..%...
4. Set the exposure time for the view to be taken, and the imaging medium used. Press the arrows below the sec. display to choose the exposure time. 5. Use the retractable tape measure on the collimator to position the HF100H at the proper source-image distance (SID) for the image to be taken.
Page 15
8. There are two types of warning indicators on the HF100H: ERROR and WARM-UP. If either one of this is lighted, x-ray cannot be generated. a. WARM-UP indicator: The internal circuits have a warm-up time of approximately 10 seconds. The WARM-UP light will turn on after each exposure.
Please keep this equipment away from likely sources of RF radiation. Maintenance Clean the HF100H as needed with a damp sponge or cloth and mild soap. The following must be checked before each use: Tubehead 1.
Page 17
4. Check the timer accuracy (refer to page 8 - Exposure Time). Have this unit check by qualified x-ray service personnel immediately if any problems occur with the HF100H. MinXray can supply detailed technical service and parts information to qualified service personnel upon request.
HF100H MinXray SPECIFICATIONS 100 kV ±10% Maximum peak tube potential : Leakage technique factors : 100 kV, 0.65 mA 0.65 mA is maximum rated continuous current for 20 mA with duty cycle 1:30 Filtration (without collimator) : 2.2 mm Al equivalent at 100 kV Total Filtration : 4.2 mm Al equivalent at 100 kV...