Advertisement

Quick Links

Configura CushionAir System
Instruction Manual
(Fits Configura Comfort, Lite 2 and Advance)
IFU-CA1-001EN REV 05
JAN 2021

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Configura CushionAir and is the answer not in the manual?

Questions and answers

Summary of Contents for Accora Configura CushionAir

  • Page 1 Configura CushionAir System Instruction Manual (Fits Configura Comfort, Lite 2 and Advance) IFU-CA1-001EN REV 05 JAN 2021...
  • Page 2: Table Of Contents

    Swords, Co. Dublin, Ireland, K67 E0A2 T: +353 (0)1 695 0614 Design Policy and Copyright ® And ™ are trademarks belonging to Accora Ltd unless otherwise stated. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.
  • Page 3: General Safety Instructions

    General Safety Instructions SPECIFIC WARNINGS 1. The Configura CushionAir system is a Medical Do not use the cushion or pump in the Device product. presence of flammable objects, chemicals or oxygen-rich environments as this may cause 2. The pump serial number of this product should be an explosion.
  • Page 4: Intended Use

    Do not use where aerosol spray products are being used. Clinical Applications Placing extra layers between the patient and Intended Use - The Configura CushionAir System is the cushion potentially reduces the benefits suitable for use in the following areas: provided by the cushion and should be •...
  • Page 5: Model Numbers And Compatibility

    (mains powered) (battery powered) moving or alternating services, clinical advice must be sought before use. General - The Configura CushionAir system must only Maximum patient 160kg / 25st / 353lbs be used with Configura chairs as recommended by Accora. See Section 4 – Model Numbers and Compatibility for part numbers and compatibility.
  • Page 6: Before Use

    WARNING • Check that the tube set does not stretch or The Configura CushionAir system must not be snag when the chair is operated through a full used until it has been set-up and configured cycle.
  • Page 7 2. Unzip the overlay cover at the back of the overlay and fold back the cover to expose the foam cushion (Configura Comfort chair shown in Figure 7.2 below): Figure 7.2 3. The foam cushion is attached to the overlay by a Figure 7.6b single dotted cut line at the front of the cushion only.
  • Page 8 8. Ensure that the tube set is correctly routed out of Configura Advance – (Figure 7.8c below) The tube the back of the chair: set must be routed alongside the backrest fitting tube (2) and above tube (3) and then upwards Configura Comfort –...
  • Page 9: Using The Battery Powered Pump

    MEDIUM or LOW. Mute Key – press the mute key (5) to The Configura CushionAir system must not be disable/enable the audible alarm. used until it has been set-up and configured by suitably trained and qualified personnel...
  • Page 10 3. Inflation nozzles - the inflation nozzles (Figure 8.3) are at the top left of the pump. While in use, both the tubes in the tube set of the cushion must be connected to the inflation nozzles. 4.5 Secure the battery cover with the screw: Figure 8.3 4.
  • Page 11 Press the Power Switch to turn the pump ON. Check if it works. • If the pump still does not work, contact your local dealer, or Accora. 6.3 In case of alarm: • Check if the tube set is firmly connected to the inflation nozzles.
  • Page 12: Using The Mains Powered Pump

    Installation and set-up/configuration must be carried out by a competent person. The Configura CushionAir system must not be used until it has been set-up/configured by suitably trained and qualified personnel after a clinical assessment and risk assessment has been carried out.
  • Page 13: Maintenance

    Configure CushionAir System, contact the clinician or clinical prescriber. 1. Maintenance for all Configure CushionAir Systems: We recommend the Configura CushionAir System is checked weekly. If any damage is detected, inform the clinician or healthcare professional and take the cushion out of service.
  • Page 14: Cleaning

    14. Guarantee 2. Cleaning the cushion cells: The warranty period for the Configura CushionAir • Brush off or wipe down all surfaces of the cover system is 24 months, covering parts and labour only. with soap and water before wetting with any We require a purchase order for any parts liquid disinfectant.
  • Page 15: Table Of Symbols

    Do not dispose of in domestic refuse Caution: risk of electrical shock Type BF applied part Class II double insulated Medical Device in accordance with EU Medical Device Regulation 2017/745 EC Representative 16. Contact details Accora Ltd. Charter House Barrington Road Address Orwell Cambridge SG8 5QP...
  • Page 16 Accora Barrington Road Orwell Cambridge SG8 5QP United Kingdom T: +44 (0)1223 206100 info@accora.care www.accora.care IFU-CA1-001EN REV 05 JAN 2021...

Table of Contents