4.8 Connections

4.8.1 Communication port connection

Librae Incapacitance Tester is provided with a D-sub (DA-15 Female) TTL I/O port (see Figure
3.8.2). This port can be used to synchronize some events with external instruments or
acquisition systems. TTL Output signal are electrically isolated to provide an electrical barrier
between Librae Incapacitance Tester and any other external device.
TTL signals are refered to Power Ground (pimn 14 and pin 15)
Refers to Figure 4.8.1.1 and Table 4.8.1.1 for connector pinout
Figure 4.8.1.1
Page: 30
< Librae instruction manual
Back to contents
4 Operation
DB-15 Pin#
Signal Name
Signal Type
1
Start/Stop
TTL OUT
2
Manual/Autostart
TTL OUT
3
Reserved
TTL OUT
4
Reserved
TTL OUT
5
Reserved
TTL OUT
6
Reserved
TTL OUT
7
Reserved
TTL OUT
8
Reserved
ANALOG OUT
9
Reserved
TTL IN
10
Reserved
TTL IN
11
Reserved
TTL IN
12
Reserved
TTL OUT
13
Reserved
TTL OUT
14
GND
POWER
15
GND
POWER
Table 4.8.1.1
NOTE: TTL OUT is designed for connection with scientific instruments! DO NOT CONNECT ANY
POWER DEVICE!
NOTE: DO NOT SINK a current more than 10mA from each TTL pin. DAMAGE WILL OCCUR
4 Operation
Back to contents
Back to contents
Description
Start -> TTL High Level
Stop -> TTL Low Level
Manual -> TTL High Level
Autostart -> TTL Low Level
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Power Ground
Power Ground