Initial Inspection - HP 11793A Operating And Service Manual

Micr0wave converter
Table of Contents

Advertisement

Model
11793A
Microwave Converter
Instrument Type
Measuring Instrument
Sensor Module
7.
Recommended Test Equipment
Minimum Specification
Suggested Model
Contains specific circuitry to control the HP 11793A.
Must have a frequency range of 26.5 GHz and be compatible
with the measuring instrument.
HP 8901 B or HP 8902A'
HP 11792A'
Table 4 lists the test equipment recommended to check, adjust, and troubleshoot the Microwave
Converter. The Minimum Specifications column describes the essential requirements for each piece of test
equipment. If substitute equipment is used, it must meet or exceed the minimum specifications.
Microwave LO Source
Microwave Signal
Source
Table 4. Recommended Test Equipment
See
LO
Amplitude Range in Table 1,
LO
Requirements. Note
that the required output power depends on the LO amplifier
option in the HP 11 793A.
Frequency Range:
2
to 26.5GHz
Output Level:
-1
to +1 dBm
HP 86738"
HP 86738
I
To Troubleshoot the Microwave Converter
I
[
Instrument Type
I
Minimum Specification
I
Multimeter
DC Volts Range: 1OOV
DC Volts Accuracy: 2%
AC Volts Range: lOOV
AC Volts Accuracy: 10% at 50 Hz
RF Test Equipment
Troubleshooting of the microwave components of the Microwave Converter requires
measurements to 26.5GHz of gain or insertion loss, SWR, power, conversion loss, and
frequency. Instruments which are required to perform these tasks include: signal genera-
tors and sweepers, network analyzers, spectrum analyzers, and power meters with ac-
companying couplers, pads, line stretchers, etc. Select the appropriate equipment and
use standard test procedures to make these measurements.
8.
I N S T A L L A T I O N
9.
Initial inspection
Inspect the shipping container for damage. If the shipping container or cushioning material is damaged,
the shipping materials should be kept until the contents
of
the shipment have been checked for
completeness and the instrument has been checked mechanically and electrically. If there is mechanical
damage or if the instrument does not pass the performance tests, notify the nearest Hewlett -Packard
office. Keep the damaged shipping materials (if any) for inspection by the carrier and a Hewlett-Packard
representative.