System Accuracy Specifications

The Agilent U2802A thermocouple input measurement accuracy with the
U2355A and U2356A is as shown in
N O T E
Table 4-1 U2802A measurement accuracy with U2355A or U2356A, at 23 °C ± 5 °C, with
Agilent U2802A 31-Channel Thermocouple Input User's Guide
• Assume a ±1 °C error in the CJ measurement due to sensor error and temperature
gradient error in the accuracy numbers in Table 4-1, Table 4-2, and Table 4-3 below.
• Table 4-1, Table 4-2, and Table 4-3 are derived from the U2802A and DAQ input accuracy
specifications without including the thermocouple error. Refer to
Accuracy"
on page 46 for calculation methodology.
ITS-90
Temperature
T/C
Range (°C)
Type
Low
High
B
0
1820
E
–270
1000
J
–210
1200
K
–270
1372
N
–270
1300
R
–50
1768
S
–50
1768
T
–270
400
different number of averaging points.
Table 4-
Thermocouple Measurements Accuracy
(U2355A, U2356A @ 23 °C ± 5°C)
Optimum
Measurement
Range(°C)
Low
High
1100
1820
400
1100
–150
1000
–200
–150
–150
1200
–210
–150
–100
1200
–200
–100
–100
1300
–200
–100
300
1760
–50
300
400
1760
–50
400
–100
400
–200
–100
Product Specifications
1,
Table 4-
2, and
"Calculating System
Without
50 points
averaging
averaging
(± °C)
(± °C)
1.9
1.2
4.4
2.5
1.7
1.6
2.4
2.3
1.6
1.5
2.7
2.6
1.5
1.4
2.7
2.6
1.5
1.3
3.0
2.7
2.0
1.4
5.0
3.1
2.1
1.6
4.5
2.8
1.5
1.4
2.7
2.5
4
Table 4-
3.
500 points
averaging
(± °C)
1.0
2.0
1.6
2.3
1.5
2.5
1.4
2.6
1.3
2.6
1.3
2.6
1.4
2.4
1.4
2.5
43