in output state
Press 〔FUN〕key , select resistance or thermal
3.
resistance (RTD) function, and display „‟ or „C‟
and thermal resistance „Pt100‟ graduation no.;
4.
In
thermal
resistance
〔RANG〕key, select corresponding graduation no.;
Press 〔〕/〔〕key, select output set bit;
5.
Press 〔〕/〔〕key , change the value of set bit,
6.
and the value can carry or abdicate automatically,
and hold the key, the value will change constantly
after one second.
7.
Press
[ZERO] key, the output will be set as
000.0C
Press [C / ℉ ] key , select Centigrade or
8.
Fahrenheit unit.
Section Seven-Performance Index
Accuracy is specified for a period of one year after calibration, at 23±5ºC, with relative humidity to 75%.
Accuracy specifications are given as: ± ([% of reading] + [number of least significant digits])
to the number of increments or decrements of the least significant digit).
Output
Range
100mV
-10.00~110.00mV
DCV
1000mV
-100.0~1100.0mV
400Ω
OMH
4000Ω
이 자료는 저작권법에 따라 보호받는 저작물이므로 무단배포 및 무단복제를 금하며, 내용을 이용하려면, 한국 VictorRuoshui 社
의 서면동의를 받아야 합니다.
www.victorruoshui.co.kr
(RTD)
function
,press
Output range
Resolution
0.01mV
0.1mV
0. 1Ω
0.0~400.0Ω
1Ω
0~4000Ω
Section Six- Function Setting
The following operation can change the automatic
power off function of the meter:
1.
When the meter is in power-off state, press [power]
key and the LCD displays in full , loose [power] key
and press [RANG] key, the meter enters into
maintenance state and the LCD displays „AP –XX‟
2.
Press [] key and the LCD displays symbol
„AP-OF‟, the meter stops automatically power-off
function ; The LCD displays symbol „AP-ON‟, the
meter restores automatically power-off function, and
the meter exit from maintenance state if cutting off
the power again.
Accuracy
0.05%+ 30uV
0.05%+0.3mV
0.05%+0.2Ω
0.05%+2Ω
refers
("Counts"
Illustration
Max. output current ±2mA
±0.5~±3mA Note 1, Note 2
±0.05~±0.3 mA Note1, Note
2