5. Cleaning and Disinfection
1. Please disinfect saline cup, spray nozzle, suction mask and so on after use. According to "GB / T 27949-2011
medical device disinfectant health requirements", "WS / T 367-2012 disinfection technical specifications",
recommended cleaning, disinfection specific methods are as follows:
1) Please cut off the power before cleaning. Remove the components from the device: cup module, saline cup cover,
mouthpiece, mask etc. and clean them with not more than 40° C warm pure water for about 5 minutes.
2) Use a clean medical gauze to suck up the water and keep them thoroughly dry after cleaning.
3) Clean the shell of device with a disinfectant wipe.
4) Keep the device in dry and clean place to avoid pollution
Note:
The body of the unit cannot be washed, please keep it away from water.
•
The disinfectant residues should be cleaned thoroughly for next safe use.
•
Do not use distilled water or purified water that has been filtered several times.
•
2.Disinfection
1)Put cup module (including Nozzle), liquid cup cover, mouthpiece, masks and other parts into 40° C maximum
water or medical disinfectants or distilled water for daily disinfection.
a.Put all parts into 40° C maximum water/medical disinfectant for 30 minutes and add non-toxic cleaners
to warm water at temperatures not exceeding 40°C and rinse the parts.
b. After rinsing, wipe the water from the device with a clean medical gauze.
c.Wipe the surface with 75% medical alcohol and remove all surfaces of mask and nozzle, then disinfect,
thoroughly dry and store.
11