KDT-2106 Infrared Thermometer Forehead, Ear and Object Multi Mode Instruction Manual...
Page 2
-2106 Model: Introduction Thank you for purchasing the KONQUEST KTD-2106 Forehead Ear and Object Multi Mode Infrared Thermometer. Please read the User Manual carefully to make sure safe and proper use of this thermometer. Please read and fully understand the Safety Precautions before use.
Unpacking Check Please open the package carefully before use, check whether all accessories are available or not and whether any component is damaged during transportation, and perform installation and operation following this user manual. In case of any damage or operation problem, please contact the dealer or contact us directly.
Safety Precautions Read the following precautions carefully before using the thermometer. Attention Take care of the temperature probe lens, which is fragile. Dispose used batteries with care. To protect the environment, you are recommended to send the used batteries to a designated collection point.
Page 6
Process" chapter and carefully operate the thermometer when measuring children's temperature. Do not immerse the thermometer in water or other liquid, as it is not waterproof. Clean and disinfect the thermometer as described in the "Cleaning and Disinfection" chapter. ...
Warning Warning Do not force the temperature probe of the thermometer into an ear canal. Otherwise, the ear canal may get injured. Keep the thermometer out of the reach of children. The result may be inaccurate if you use the overdue thermometer. ...
Do not use the thermometer in an environment where flammable anesthetic mixture with air or with oxygen or nitrous oxide is available. Symbols Symbol Description Type BF applied part. Attention must be paid. The action is prohibited. Information about the manufacturer. Date of manufacture.
Page 9
Symbol Description Inaccurate reading or damage to the thermometer may Attention occur if the thermometer is not used correctly.
Body Temperature Basics Generally, you can take a body temperature on the forehead, in the ear canal, under the armpit, in the mouth, or in the anus. The temperature measured at different parts of the body may differ slightly. Body Part Normal Temperature Range Forehead 36.1°C–37.5°C / 97.0°F–99.5°F...
Intended use The KDT-2106 Dual Mode Digital Infrared Thermometer is intended for the measurement of human body temperatures. The forehead mode is indicated for use by people of all ages and the eardrum mode is indicated...
Contraindications Do not use the thermometer if the ear is infected with otitis or suppuration. Features Good safety ● Passive infrared receiving technology. Easy operation ● Ergonomic design ● One button measurement Quick measurement ● 1-second measurement High accuracy ● Advanced infrared temperature sensor, with high sensitivity ●...
Sounds and backlight instructions Range Sounds Backlight Forehead temperature 35.0℃-37.5℃/95.0℉-99.5℉ A long beep Green 37.6℃-42.2℃/99.6℉-108.0℉ 3 short double beeps Ear temperature 32.0℃-37.5℃/89.6℉-99.5℉ A long beep Green 37.6℃-42.2℃/99.6℉-108.0℉ 3 short double beeps Object temperature 0℃-100℃/32.0℉-212℉ A long beep White Note: When the forehead temperature is between 35.0°C/95.0℉ and 37.5°C/99.5℉, and the ear temperature is between 32.0°C/89.6℉...
Display and Operating Instructions Operating Instructions Sound and Screen Display backlight Displayed State Measuring Ear temperature Take off the probe cover, press and release the Power button for 1 second to power on the thermometer. Press the Mode See the table in the button, the thermometer enters the “Sounds and Ear mode.
Page 16
Operating Instructions Sound and Screen Display backlight Displayed State Measuring Object temperature Press and release the Power button for 1 second to power on the thermometer. Then press the Mode See the table in the button. The thermometer enters the “Sounds and Object mode.
Page 17
Operating Instructions Sound and Screen Display backlight Displayed State Recall 20 memories...
Page 18
Operating Instructions Sound and Screen Display backlight Displayed State In a power-on state, press the Memory button enter the memory mode. When the Memory button is released, 1 will be shown, followed by the recorded reading. Press the Memory button again for the next recorded data.
Page 19
Operating Instructions Sound and Screen Display backlight Displayed State Switching between object temperature and body temperature Press the Mode button to switch between object temperature and body temperature. Silent Body temperature contains the Forehead temperature and Ear temperature. Switching between ºF/ºC In a power-on state, press and hold the Unit switch button for more than 4 seconds to enter the...
Page 20
Operating Instructions Sound and Screen Display backlight Displayed State Switching between mute and un-mute Toggle the Sound switch to switch between mute and un-mute Silent symbol is displayed in Mute mode and disappears in Un-mute mode. Error information & low battery The ambient temperature is higher than 40.0ºC (104.0ºF) or lower than 3 short beeps...
Operating Instructions Sound and Screen Display backlight Displayed State When the battery voltage is lower than 2.5V ± 0.1V, the low battery Silent symbol will appear on the display. Please replace the batteries. Measuring Ear Temperature 1. When using the thermometer for the first time, move the battery’s insulating piece away.
4. Press the Mode button, the thermometer enters the Ear mode. The “Ear” symbol is displayed on the screen. 5. Insert the temperature probe into the ear canal. 6. Press and release the Measure button. The ear temperature reading will be display on the screen instantly. Note: Children under 1 year: Pull the ear straight back.
Press the Mode button, the thermometer enters the Forehead mode. the “Head” symbol is displayed on the screen. Point the thermometer probe to the temple of the forehead, about ½” to 2” (1-5cm) away from the skin surface. Press and release the Measure button for 1 second. The temperature reading will be displayed on the screen instantly.
Page 24
After a measurement (1) After each measurement, you can enter the recall mode and query earlier temperature readings. For more details, see "Recall 20 memories" in the preceding table. (2) After each measurement, clean the temperature probe with a soft cloth, and put the thermometer in a dry and well-ventilated place.
Replacing Batteries 1、Slide the battery cover off along the marked direction and take it off. 2、Insert the two AAA batteries into the compartment according to the stated polarities. Make sure that the batteries are installed correctly. Otherwise, the thermometer may be damaged. If the low-battery symbol is displayed on the screen, replace the batteries.
Cleaning and Disinfection Cleaning Recommended detergents: * Medical detergents; * Home use mild detergents; Cleaning steps: (1) Take the batteries out before cleaning. (2) Clean the temperature probe with a soft cloth. Clean the lens of the temperature probe with a cotton swab. (3) Wipe the thermometer body with a slightly damp soft cloth.
* Medicinal alcohol (concentration: 75%) * Sodium hypochlorite solution (concentration: 3%) Disinfecting steps: 1) Wet the clean soft cloth with a small quantity of disinfectant, wipe the thermometer and quickly dry it. 2) Disinfect the thermometer body and the area around the temperature probe with a cloth slightly moistened with 75% medical alcohol.
that the storage and transportation environments meet the requirements. 4) Check regularly whether safety risks exist. 5) Remove the batteries if the thermometer will not be used for more than two months. Troubleshooting Problem Possible Cause Solution Low battery Change the batteries. Polarities of the Make sure that the batteries thermometer...
Specifications Product Name Infrared Thermometer Product Model KDT-2106 Power Supply Mode Internal power supply Operating Voltage DC 3V Battery Model AAA x 2 Battery Life Alkaline dry battery for around 20,000 measurements Operating Mode...
Weight (g) Thermometer (with batteries): 86.9 g Temperature: 10°C~ 40°C (50°F–104°F) Operating Humidity: 15%–95% RH, non-condensing Environment Atmospheric pressure: 86–106 kPa The infrared thermometer has been tested and conforms to the standard ASTM E1965-98. ASTM laboratory accuracy requirements in the display range of 96.8°F to102.2°F (36°C-39°C) for ear canal IR thermometers is ±0.4°F (±0.2°C).
DOCUMENTS. ● Portable and mobile RF communications equipment can affect Infrared Thermometer KDT-2106. ● The Infrared Thermometer KDT-2106 should not be used adjacent to or stacked with other equipment. Guidance and manufacturer´s declaration – Electromagnetic emission –for all equipment and systems Guidance and manufacturer´s declaration –...
Page 32
Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer KDT-2106 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer KDT-2106 should assure that it is used in such an environment. Immunity test Compliance...
Page 33
Guidance and manufacturer´s declaration – Electromagnetic immunity The Infrared Thermometer KDT-2106 is intended for use in the electromagnetic environment specified below. The customer or the user of the Infrared Thermometer KDT-2106 should assure that it is used in such an environment. Immunity test Compliance...
Page 34
calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance where p is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in metres (m). Field strengths from fixed transmitters, determined by an...
Page 35
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the KDT-2106 is used exceeds the applicable RF compliance level above, the KDT-2106 should be observed to verify normal operation.
Page 36
Infrared Thermometer KDT-2106 as recommended below, according to the maximum output power of the communications equipment. Rated Separation distance according to frequency of transmitter maximum output 80 MHz to 800 MHz 800 MHz to 2,5 GHz power transmitter 0.01 0.12 0.23...
1. Failure resulting from unauthorized disassembly and modification. 2. Failure resulting from an unexpected dropping during application or transportation. 3. Failure resulting from not following the instructions in the user’s manual. Manufactured for Konquest Group Inc, Email: info@konquestusa.com www.konquestusa.com Telephone: +1-833-KONQUEST (+1-833-566-7837)
Page 38
Manufactured for Konquest Group Inc by FDA 510(k) Holder #K131243 Konquest Group Inc acts as Initial Importer of Device Made in China...