Rev. 2021-07-01 OkuStim System
TD60K06 -
03
18 Information Regarding Electromagnetic
Compatibility
The OkuStim system can be used in a professional healthcare or home environment.
Warning:
Use of this equipment adjacent to or stacked with other equipment should be
avoided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that
they are operating normally.
Warning:
Use of accessories and cables other than those specified on page 31 or
provided by the manufacturer of this equipment could result in increased elec-
tromagnetic emissions or decreased electromagnetic immunity of this equip
ment and result in improper operation.
The OkuSpex cable has a length of 1.6 m.
Warning:
Portable RF communication devices (incl. accessories like antenna cables or
external antennas) should be at least 30 cm away from the OkuStim system
and all its parts, including cables specified by the manufacturer. Otherwise
the performance of the OkuStim system could be compromised.
See table in this chapter.
Warning:
Interference of the OkuStim is possible in the frequency range of 80 MHz to 86
MHz. In this case, the error message „Error Defect!", see chapter 13, will be
displayed.
32 36
Instructions for Use