11.3. RESULTS ANALYSIS

Calculate the `p-values' for each reading as
follows:
(working in kilograms)
meter reading
p =
--------------------- x 100%
16.48
If both 'p-values' are in the range 97% to 103%
the Milk Meter is considered acceptable.
If only one of the 'p-values' is outside the range
97% to 103%, perform the test a third time. The
meter is then considered acceptable if no single
'p-value' is outside the range 95% to 105% and
the mean of all three values is within the range
97% to 103%.
Any Milk Meter failing this test should be
withdrawn from service and submitted to a
certified
service
agent
recalibration.
Note: The periodic test procedure should be
regularly followed to
give confidence in Milk
Meter results.
Mk5 Milk Meter
Technical Manual
PERIODIC TEST PROCEDURE
METHODE DE TEST
PERIODIQUE
for
repair
and
Compteur a Lait Mark V
Manuel Technique

11.3. ANALYSE DES RESULTATS

Calculer les « valeurs p » pour chaque lecture
comme suit :
(résultats en kilogrammes)
lecture du compteur
p = ------------------------
x 100%
16,48
Lorsque les deux « valeurs p » sont comprises
entre 97% et 103%, le compteur est accepté.
Si l'une des deux « valeurs p » n'est pas
comprise entre 97% et 103%, refaire le test une
troisième fois. Le compteur est alors accepté si
aucune des « valeurs p » n'est située en dehors
de l'intervalle 95% à 105% et si la moyenne des
trois valeurs est comprise entre 97% et 103%.
Tous compteurs ne pouvant passer ce test
doivent être retirés et remis à un réparateur
agréé pour un nouvel étalonnage.
Note : La méthode de test périodique doit être
suivie régulièrement afin d'augmenter la fiabilité
des compteurs.