3. Technical Data
Master Control Unit
Model No.
Model Name
Size (inch) LxWxH
Weight
Cycle Time (min)
Min Operating Pressure
Max Operating Pressure
Rated Voltage
Max Current
Fuse Rating
Rated Frequency
Classification
Mode of Operation
Environment
(Temperature)
Environment (Humidity)
Standard
Mattress Replacement
Model No
Size (inch)
Weight (lbs)
Cells Number
Cells Material
Cover Material
Base Material
FC-MOX0027
Duet
12.2" x 6.7" x 13.2"
14.3lbs
3 - 95
8 +/- 5mmHg
33 +/- 5mmHg
AC 110-120V
5 Amp
T5AH 250V
60 Hz
Class I, Type BF
Not AP/APG type
Continuous
Operation: 15°C to 35°C (59°F to 95°F)
Storage:5°C to 60°C (41°F to 140°F)
Operation: 30% to 75% non-condensing
Storage: 30% to 90% non-condensing
IEC 60601-1,
CAN/CSA C22.2 No. 60601-1,
IEC 60601-1-2
FM-MOX0009(36")
FM-MOX0044(42")
FM-MOX0045(48")
36"(W)x80"(L)x10"(H)
42" (W)x80"(L)x10"(H)
48" (W)x80"(L)x10"(H)
39.6
18 cells
Nylon coated with PU
Nylon woven fabric w/ PU coating finish
Woven Polyester fabric w/ PVC backing
4