Page 1
DA15K e015b DA15KE Data Acquisition Systems User Guide SIAP+MICROS S.r.l. tel +39 0438 491411 – fax +39 0438 401573 Via del Lavoro,1 31020 – Castello Roganzuolo di San Fior (TV) - ITALY email info@siapmicros.com di San Fior (TV) www.siapmicros.com...
Substitution of components and intervention inside should be performed only by qualified and instructed service personnel, after disconnecting the main power supply. Pay attention to any warning label on order to avoid potentially dangerous procedures. User guide (v.02) pag.1 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Substitution of components and intervention inside should be performed only by qualified and instructed service personnel, after disconnecting the main power supply. Pay attention to any warning label on order to avoid potentially dangerous procedures. User guide (v.02) pag.2 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
6. Turn on the external power supply. 1.6 During operation During operation, please avoid to alter the electrical connections related to analog and digital inputs; also avoid to alter the connections of the power supplies. User guide (v.02) pag.3 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
All line protections, both for short and for surges or lightning are self-restoring (PTC, self restoring fuses, ESD protections and so on). The product does not have removable/replaceable line protection. 1.9 Description of case and connections 1.9.1 Box of data processing electronics Installation and Dimensions User guide (v.02) pag.4 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 7
Three pins for the CAN – BUS port, for connecting devices with CAN – BUS. Three pins for the SDI – 12 port for connecting devices with SDI – 12 protocol User guide (v.02) pag.5 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 8
TX and RX COM3 status of the lines of transmission and reception of serial 3 TX and RX COM2 status of the lines of transmission and reception of serial 2 User guide (v.02) pag.6 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
DA15Kx 1.9.2 Box of data acquisition electronics Dimensions Installation Panel or wall mount with M3 screw Panel or wall mount with low profile DIN rail support User guide (v.02) pag.7 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 10
DIG IN B : Optoisolated digital input (see picture at the top page ) DIG IN C : Optoisolated digital input • Logic level • Sinusoidal Input (IF11-IF12) • Frequency • Sinusoidal Input (IF7-IF8) • Counter • Standard 12V sensor User guide (v.02) pag.8 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
1.11 Description of directories and files DA15K data logger is equipped with Windows CE 6.0 operating system and it is factory configured with the following structure of drives, directories and files: User guide (v.02) pag.9 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
All software contained in this directory or with a link in this directory are automatically started at the power on of the data logger. User guide (v.02) pag.10 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Voltage detected (1 flash empty, 5 flashes full). Few seconds later turning on the display, bootstrap mask with label “SIAP+MICROS” will appear. 30s later, Windows CE 6.0 O.S. will start to run and the applications placed in the directory “\NandFlash\Windows\StartUp”...
USB interface. At the end, save the changes using the "Save”. 2.2.3 Status Displays the status of the file archives with the occupied and free size for any file. User guide (v.02) pag.12 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
• alarms • The steps to restart the program • The error was detected in clock synchronization • The errors in connection with the communication devices. User guide (v.02) pag.13 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
"Hard Disk", copy the new program, navigate to the directory "\ NandFlash \ DA9000" and overwrite the file listed. After performing the copy, start the START.EXE software or restart the datalogger. User guide (v.02) pag.14 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
SD memory card can be removed in every moment, even if the software is running. In order to avoid truncation of the stored records, it’s highly recommended to close the program or turn off the DA15K data logger during replacement. User guide (v.02) pag.15 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
In detail the four types are: • Record track for Statistical Data • Record track for Instantaneous data • Record track for Alarm Data • Record track for Calibration Data User guide (v.02) pag.16 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
… Omissis” 4.2 EMC This equipment has been designed in accordance with the requirements of the Directives listed in the EC declaration attached to the product. User guide (v.02) pag.18 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
The manufacturer may update the product described in this document, or parts of it, at any time without notice. • It is not allowed reproduce or duplicate, entirely or part of this document without the permission of SIAP+MICROS srl. User guide (v.02) pag.19...
DA15Kx 6 Quick Guide Set clock without DAK 1. Close DA9000 program pressing X button; confirm typing YES and then OK User guide (v.02) pag.20 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 23
DA15Kx 2. Open folder 3. Open Nand Flash Folder 4. Open DA9000 Folder User guide (v.02) pag.21 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 24
DA15Kx 5. Start RTC program 6. Set clock and date and press WRITE 7. Switch OFF and switch ON the datalogger User guide (v.02) pag.22 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
DA15Kx 6.2 Changing configuration without DAK 1. Close DA9000 program. pressing X botton, confirm with YES and OK User guide (v.02) pag.23 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 26
3. Insert the Usb pen drive and wait few seconds for its loading. Then open “Hard disk” folder 4. Select "CNF" configuration file and, if present, "INI" settings file and Copy them. User guide (v.02) pag.24 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 27
DA15Kx 5. Open Nand Flash Folder 6. Open DA9000 Folder User guide (v.02) pag.25 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.
Page 28
DA15Kx 7. Paste new file in the DA9000 folder and confirm update with "Yes to All" 8. Remove USB pen drive and Switch OFF and switch ON the datalogger User guide (v.02) pag.26 Printed in e015-di DA15K [Eng].docx Siap+Micros S.r.l.