PDK PCS-P100 User Manual

Portable pneumatic pressure calibrator
Hide thumbs Also See for PCS-P100:

Advertisement

Quick Links

User Manual
MODEL : PCS-P100
Portable Pneumatic Pressure Calibrator
PDK CO., LTD.
www.pdk.co.kr

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PCS-P100 and is the answer not in the manual?

Questions and answers

Summary of Contents for PDK PCS-P100

  • Page 1 User Manual MODEL : PCS-P100 Portable Pneumatic Pressure Calibrator PDK CO., LTD. www.pdk.co.kr...
  • Page 2: Table Of Contents

    PCS-P100 User Manual Table of Contents 1. General Introduction ....................... 2 1.1 Introduction ..........................2 1.2 Features ............................2 1.3 Use ............................... 3 1.4 Calibrator Interface ........................4 1.5 Specifications ..........................5 1.6 Ordering Information ........................5 1.7 Accessories ........................... 6 1.8 Optional items...
  • Page 3: General Introduction

    Full-graphic touchscreen display with intuitive menu selection makes it high visibility and operability. Upper side of PCS-P100 can equipped quick connectors and adaptors for quick and easy to connect UUT. 1.2 Features Dual-Stage lever type hand pump to generate and control up to 10 MPa 2) Dual-Stage lever type hand pump to generate and control up to –97 kPa for vacuum...
  • Page 4: Use

    PCS-P100 User Manual 1.3 Use 1) Potable pressure calibration 2) Simple calibration for common calibration lab 3) Calibrate pressure gauge for pressure measuring instruments retail business 4) Pressure generating and controlling for pressure test lab 5) Pressure calibration for laboratory 6) Pneumatic high pressure calibration 7) Easy &...
  • Page 5: Calibrator Interface

    PCS-P100 User Manual 1.4 Calibrator Interface ① Test port ⑩ Vent / Release valve ② On / Off ⑨ Electrical Connections ③ Full color touch screen display ⑧ Hand strap ⑦ Lever type hand pump ④ Fine pressure ⑥ Pressure / Vacuum...
  • Page 6: Specifications

    PCS-P100 User Manual 1.5 General Specifications Pressure range -0.097 to 10 MPa Accuracy ± 0.02% F.S ±15 V.dc, ±24 mA.dc, Electrical measurement and source Accuracy ±(0.01% of Reading + 1digit), Loop Power Full color touch screen LCD, 110 mm(4.3") diagonal, Display 480 x 272 pixels., 6 digit...
  • Page 7: Accessories

    PCS-P100 User Manual 1.7 Accessories ① ② ③ ④ Item Quantity ① Carrying case ② Rechargeable battery ③ Battery charger ④ Test lead set 1.8 Optional Items ① ② ③ ④ ⑤ Item Description ① Dirt / Moisture Trap Up to 10 MPa CA-P Series ②...
  • Page 8: Main Menu

    PCS-P100 User Manual 2. Main Menu 2.1 Power on / off 2.1.1 Power on ① Press and hold down Power button until ② screen shows. It will changes ② screen to ③ screen. 2.1.2 Power off ① Press and hold down Power button until in 0% to 100% on ② screen.
  • Page 9: Unit / Resolution / Zero Cal

    PCS-P100 User Manual 2.2 Unit / Resolution / Zero Cal. 2.2.1 Unit The current pressure value is displayed, and the full scale and the green bar increases proportionally. This is a function that prevents internal sensor damage due to overpressure.
  • Page 10 PCS-P100 User Manual 2.2.3 Zero cal. If pressure does not indicate zero when fully vented condition, touch to zeroing. Zeroing function only works when 0 ± 5% F.S. range. www.pdk.co.kr...
  • Page 11: Slide

    PCS-P100 User Manual 2.3 Slide Touch and slide the screen to move page as below figure for Electrical Signal measurement. > > > > > www.pdk.co.kr...
  • Page 12: Ma 2Wire, 24V Out / Ma 3Wire , 24V Out

    PCS-P100 User Manual 2.4 mA 2wire , 24V out / mA 3wire , 24V out 2.4.1 mA 2wire , 24V out Touch Help to see the reference connection. 부분을 확인할 수 있습니다. Touch Exit to change the screen back. 왼쪽 화면으로 전환됩니다.
  • Page 13: Vdc 3Wire, 24V Out / Screen Switching

    PCS-P100 User Manual 2.5 Vdc 3wire , 24V out / Screen switching 2.5.1 Vdc 3wire , 24V out Touch Help to see the reference connection. Touch Exit to change the screen back. Measure up to ±15V, Supply up to 24 VDC.
  • Page 14: Task Menu

    PCS-P100 User Manual 3. TASK Menu 3.1 TASK menu Touch to go to task set up menu. Max / Min test, Leak test, Switch test and Setting can be selectable. Leak Test Max/Min/Avg Setting Switch Test www.pdk.co.kr...
  • Page 15: Max/Min/Avg

    PCS-P100 User Manual 3.2 Max / Min / Avg This function only can use in pressure measure mode. Indicates pressure maximum, minimum and average (Mean). 1) Touch then select Max & Min 2) Touch to start measure 3) Touch to reset the values 3.3 Leak Test...
  • Page 16: Switch Test

    PCS-P100 User Manual 3.4 Switch Test This function only can use in pressure measure mode. Indicates opened and closed values when PCS sensed a change. The pressure difference between the change-of-states (i.e. OPEN- CLOSE or CLOSE-OPEN) value shows in Hysteresis.
  • Page 17 PCS-P100 User Manual 3.5.1 Time User can set time in Time page. 3.5.2 Cal.Pressure User can calibrate the pressure values in Cal.Pressure page. www.pdk.co.kr...
  • Page 18 PCS-P100 User Manual 3.5.3 About PCS User can see the current PCS information in About PCS page. www.pdk.co.kr...
  • Page 19: Operation

    PCS-P100 User Manual 4. Operation 4.1 Basic Structure ③ Vent / release ② Test port ① ④ Lever type hand pump ① On / Off ⑦ Pressure / Vacuum switching valve ⑤ Charging port ⑥ Fine pressure adjustment Description On / Off ①...
  • Page 20: Pressurize

    PCS-P100 User Manual 4.2 Pressurize ③ Vent / release ② Test port ③ ④ Lever type hand pump ① On / Off ⑦ Pressure / Vacuum switching valve ⑤ Charging port ⑥ Fine pressure adjustment 1) ② Connect the UUT to test port. (1/8" PF)
  • Page 21 PCS-P100 User Manual 3) ③ Turn Vent / Release valve to CW 2) ⑦ Turn Pressure / Vacuum switching to close the Vent / Release valve. valve to CCW to Switching to ‘ P ’. Normal pressure generation Low pressure generation 4) ④...
  • Page 22 3. Low pressure gauges can very easily be over pressured if not careful. Please take caution when applying pressures. 4. Store the PCS-P100 in a dry non-corrosive environment. 5. PDK is not responsible for problems or damage resulting from improper use or operation of the equipment. www.pdk.co.kr...
  • Page 23: Vacuum

    PCS-P100 User Manual 4.3 Vacuum ③ Vent / release ② Test port ④ Lever type hand pump ① On / Off ⑦ Pressure / Vacuum switching valve ⑤ Charging port ⑥ Fine pressure adjustment 1) ② Connect the UUT to test port. (1/8" PF)
  • Page 24 PCS-P100 User Manual 2) ⑦ Turn Pressure/Vacuum switching 3) ③ Turn Vent/Release valve to CW valve to CW to Switching to ‘ V ’. to close the Vent/Release valve. 4) ④ Generate the target pressure with Hand pump Pressing the pump up once without volume will apply high vacuum.
  • Page 25 3. Low pressure gauges can very easily be over pressured if not careful. Please take caution when applying pressures. 4. Store the PCS-P100 in a dry non-corrosive environment. 5. PDK is not responsible for problems or damage resulting from improper use or operation of the equipment. www.pdk.co.kr...
  • Page 26: Troubleshooting

    PCS-P100 operation. This troubleshooting guide is intended as an aid in identifying the reason for PCS-P100 behavior and determining whether the behavior is due to normal operation or an internal or external problem.
  • Page 27 PCS-P100 User Manual Symptom Possible cause Solution Test port leak Verify Test port is leak tight. While using measure UUT leak Verify UUT is leak tight. mode, pressure is leaking contact the manufacturer or Internal leak authorized agent. Pressure supply too low Verify pressure supply.
  • Page 28 PCS-P100 User Manual PDK CO., LTD SERVICE CENTER Company Address Contacts (Postal Code. 34122) Tel. 042-862-6880 PDK CO., LTD 10-6, Expo-ro 339beon-gil, Fax. 042-862-6881 Head Office Yuseong-gu, Daejeon, Korea E-mail. pdk@pdk.co.kr (Postal Code. 08506) Tel. 02-815-7950 PDK CO., LTD 1-610, IT Castle, 98, Gasan digital Fax.